1.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
2.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
3.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
4.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
5.Intelligent handheld ultrasound improving the ability of non-expert general practitioners in carotid examinations for community populations: a prospective and parallel controlled trial
Pei SUN ; Hong HAN ; Yi-Kang SUN ; Xi WANG ; Xiao-Chuan LIU ; Bo-Yang ZHOU ; Li-Fan WANG ; Ya-Qin ZHANG ; Zhi-Gang PAN ; Bei-Jian HUANG ; Hui-Xiong XU ; Chong-Ke ZHAO
Ultrasonography 2025;44(2):112-123
Purpose:
The aim of this study was to investigate the feasibility of an intelligent handheld ultrasound (US) device for assisting non-expert general practitioners (GPs) in detecting carotid plaques (CPs) in community populations.
Methods:
This prospective parallel controlled trial recruited 111 consecutive community residents. All of them underwent examinations by non-expert GPs and specialist doctors using handheld US devices (setting A, setting B, and setting C). The results of setting C with specialist doctors were considered the gold standard. Carotid intima-media thickness (CIMT) and the features of CPs were measured and recorded. The diagnostic performance of GPs in distinguishing CPs was evaluated using a receiver operating characteristic curve. Inter-observer agreement was compared using the intragroup correlation coefficient (ICC). Questionnaires were completed to evaluate clinical benefits.
Results:
Among the 111 community residents, 80, 96, and 112 CPs were detected in settings A, B, and C, respectively. Setting B exhibited better diagnostic performance than setting A for detecting CPs (area under the curve, 0.856 vs. 0.749; P<0.01). Setting B had better consistency with setting C than setting A in CIMT measurement and the assessment of CPs (ICC, 0.731 to 0.923). Moreover, measurements in setting B required less time than the other two settings (44.59 seconds vs. 108.87 seconds vs. 126.13 seconds, both P<0.01).
Conclusion
Using an intelligent handheld US device, GPs can perform CP screening and achieve a diagnostic capability comparable to that of specialist doctors.
6.Three 2,3-diketoquinoxaline alkaloids with hepatoprotective activity from Heterosmilax yunnanensis
Rong-rong DU ; Xin-yi GUO ; Wen-jie QIN ; Hua SUN ; Xiu-mei DUAN ; Xiang YUAN ; Ya-nan YANG ; Kun LI ; Pei-cheng ZHANG
Acta Pharmaceutica Sinica 2024;59(2):413-417
Three 2,3-diketoquinoxaline alkaloids were isolated from
7.Introduction and application of non-parametric regression method in medical research
Ya-Hang LIU ; Yong-Fu YU ; Guo-You QIN
Fudan University Journal of Medical Sciences 2024;51(2):280-284
This article introduced the basic theory of non-parametric regression and its application in medical and public health research for methodological reference.We conducted Cox proportional hazard models with restricted cubic splines using chronic disease management data from a Center for Disease Control and Prevention.We aimed to explore the separate and combined effects of mean fasting blood glucose level and glucose variability on all-cause mortality among individuals with type 2 diabetes.A non-linear association was observed between glucose variability and the risk of all-cause mortality.The association between glucose variability and all-cause mortality was stronger at higher mean fasting blood glucose levels compared to lower levels.The non-parametric regression methods comprehensively explored dose-response relationships between continuous exposure and outcome,revealing the combined effects of continuous exposures,which provided recommendations for targeted interventions.The method showed promising application value in medical and public health research.
8.Investigation on a suspected outbreak of carbapenem-resistant Acinetobacter baumannii nosocomial infection in an intensive care unit
Mei HUANG ; Xiaobo GUI ; Ya YANG ; Feng LU ; Juanxiu QIN ; Yan LI ; Shuyi ZHANG ; Wenqin ZHOU ; Xiaofang FU ; Haiqun BAN
Shanghai Journal of Preventive Medicine 2024;36(5):435-438
ObjectiveTo investigate a suspected outbreak of carbapenem-resistant Acinetobacter baumannii (CRAB) nosocomial infection in an intensive care unit (ICU) and provide scientific evidence for prevention and control of multi-drug resistant nosocomial infection. MethodsClinical and epidemiological data of 4 patients with CRAB infection were retrospectively investigated in the ICU of Renji Hospital in November 2021. Environmental hygiene monitoring and multilocus sequence typing (MLST) were conducted and intervention measures were taken. ResultsA total of 4 cases with CRAB infection were identified, among which 1 case was determined to be community-acquired and3 cases were hospital-acquired. The isolated strains shared the same drug resistance, and were all classified into ST368 type. In the surface and hand samples (n=40), 2 CRAB strains were detected in the air filter beside the bed of the first case, with a detection rate of 5%. After adopting comprehensive prevention and control strategies, including environmental cleaning and disinfection, hand hygiene, staff management and training, and supervision, no similar case with CRAB infection was found. ConclusionThis suspected outbreak of CRAB nosocomial infection may be induced by inadequate environmental cleaning and disinfection, and inadequate implementation of hand hygiene. Timely identification, investigation, and targeted measures remain crucial to effective control of possible nosocomial infection.
9.Research progress on the application of membrane biomimetic nanocarriers in targeted therapy of lung diseases
Wen-jing BAI ; Rong GUO ; Lin XIONG ; Leng-jing ZHU ; Jia-xin LI ; Ya-shi WANG ; Man LI ; Qin HE
Acta Pharmaceutica Sinica 2024;59(10):2730-2740
Pulmonary disease is one of the major threats to human health. However, the current clinical treatment drugs for lung diseases generally have problems such as low lung delivery efficiency, fast clearance rate and obvious toxic side effects. Recently, membrane biomimetic nanocarriers have attracted more and more attention. Due to their advantages of high targeting, long cycle time, good biocompatibility and strong immune escape ability, membrane biomimetic nanocarriers have become a major research hotspot in targeted therapy of lung diseases. In this review, we discuss the main preparation methods of membrane biomimetic nanoparticles, the characteristics of membrane biomimetic nanocarriers from different cell sources and their application in the targeted therapy of lung diseases. At the same time, according to the characteristics of different membranes, the shortcomings, current technical limitations and future prospects are discussed. This review is expected to provide references for the design of membrane biomimetic nanocarriers and their potential applications in the treatment of lung diseases.
10.Chemical constituents from Ganoderma angustisporum and their α-glucosidase inhibitory activities
Ya-Qin HUO ; Yu-Xi WANG ; Yu-Lian WEI ; Yi-Xuan ZHANG ; Hai-Sheng YUAN
Chinese Traditional Patent Medicine 2024;46(1):132-137
AIM To study the chemical constituents from Ganoderma angustisporum J.H.Xing,B.K.Cui&Y.C.Dai and their α-glucosidase inhibitory activities.METHODS The ethyl acetate extract from G.angustisporum was isolated and purified by silica gel,ODS,TLC and HPLC,then the structures of obtained compounds were identified by physicochemical properties and spectral data.pNPG method was used to evaluate their α-glucosidase inhibitory activities.RESULTS Seven compounds were isolated and identified as N-acetyl-L-phenylalanine ethyl ester(1),N-acetyl-L-phenylalanine methyl ester(2),4-hydroxy-17R-methylincisterol(3),6,8-di-O-methylaverufin(4),aversin(5),methyl 2-(4-hydroxyphenyl)aceate(6),5-toluene-1,3-diol(7).Compounds 1-2,4-7 showed inhibitory activities of α-glucosidase with IC50 values being(33.80±0.47),(45.45±7.95),(48.80±5.86),(39.48±2.82),(41.47±6.68),(55.38±10.12)μmol/L,and compound 1 showed good inhibitory activity.CONCLUSION Compound 1 is a new natural product.Compounds 2-7 are isolated from genus Ganoderma for the first time.Compounds 1-2,4-7 have α-glucosidase inhibitory activities.

Result Analysis
Print
Save
E-mail