1.In vitro heat insulation efficacy of 5% dextrose versus 0.9% saline during radiofrequency ablation
Yanping MA ; Jinfen WANG ; Tao WU ; Bowen ZHENG ; Tinghui YIN ; Yufan LIAN ; Jie REN
Ultrasonography 2024;43(5):376-383
Purpose:
This study compared the efficacy of heat insulation between 5% dextrose and 0.9% saline in radiofrequency ablation (RFA). Accordingly, temperature variations and maximum temperatures were assessed at identical distances and heat field distributions.
Methods:
Cubes of porcine liver tissue, measuring 10 mm across, were selected to precisely align the ablation boundary with the tissue boundary. An 18-gauge electrode with a 7-mm tip was inserted into each cube (10 per group) in a stainless-steel cup containing 40 mL of 5% dextrose or 0.9% saline. Fixed ablation was performed for 3 minutes using continuous mode at 30 W, simulating the typical thermal environment during thyroid RFA. Real-time temperature measurements were recorded by sensors positioned 0, 1, 3, and 5 mm from the cube’s edge. A comparative analysis was conducted to assess the maximum temperature, temperature variation, and duration of temperatures exceeding 42℃.
Results:
In both groups, the temperature curve declined with increasing distance from the edge of the ablated tissue. However, 0.9% saline exhibited higher maximum temperatures at 1, 3, and 5 mm compared to 5% dextrose (1 mm: 44.55°C±5.25°C vs. 34.68°C±3.07°C; 3 mm: 39.64°C±2.53°C vs. 29.22°C±2.21°C; 5 mm: 38.86°C±2.14°C vs. 28.74°C±2.51°C; all P<0.001). Considering a nerve injury threshold of 42°C, the 0.9% saline also displayed a greater proportion of samples reaching this temperature and a longer duration of temperatures exceeding it (P<0.05).
Conclusion
The heat insulation efficacy of 5% dextrose at 1-5 mm exceeds that of 0.9% saline at identical distances and in a common thermal environment during thyroid RFA.
2.In vitro heat insulation efficacy of 5% dextrose versus 0.9% saline during radiofrequency ablation
Yanping MA ; Jinfen WANG ; Tao WU ; Bowen ZHENG ; Tinghui YIN ; Yufan LIAN ; Jie REN
Ultrasonography 2024;43(5):376-383
Purpose:
This study compared the efficacy of heat insulation between 5% dextrose and 0.9% saline in radiofrequency ablation (RFA). Accordingly, temperature variations and maximum temperatures were assessed at identical distances and heat field distributions.
Methods:
Cubes of porcine liver tissue, measuring 10 mm across, were selected to precisely align the ablation boundary with the tissue boundary. An 18-gauge electrode with a 7-mm tip was inserted into each cube (10 per group) in a stainless-steel cup containing 40 mL of 5% dextrose or 0.9% saline. Fixed ablation was performed for 3 minutes using continuous mode at 30 W, simulating the typical thermal environment during thyroid RFA. Real-time temperature measurements were recorded by sensors positioned 0, 1, 3, and 5 mm from the cube’s edge. A comparative analysis was conducted to assess the maximum temperature, temperature variation, and duration of temperatures exceeding 42℃.
Results:
In both groups, the temperature curve declined with increasing distance from the edge of the ablated tissue. However, 0.9% saline exhibited higher maximum temperatures at 1, 3, and 5 mm compared to 5% dextrose (1 mm: 44.55°C±5.25°C vs. 34.68°C±3.07°C; 3 mm: 39.64°C±2.53°C vs. 29.22°C±2.21°C; 5 mm: 38.86°C±2.14°C vs. 28.74°C±2.51°C; all P<0.001). Considering a nerve injury threshold of 42°C, the 0.9% saline also displayed a greater proportion of samples reaching this temperature and a longer duration of temperatures exceeding it (P<0.05).
Conclusion
The heat insulation efficacy of 5% dextrose at 1-5 mm exceeds that of 0.9% saline at identical distances and in a common thermal environment during thyroid RFA.
3.In vitro heat insulation efficacy of 5% dextrose versus 0.9% saline during radiofrequency ablation
Yanping MA ; Jinfen WANG ; Tao WU ; Bowen ZHENG ; Tinghui YIN ; Yufan LIAN ; Jie REN
Ultrasonography 2024;43(5):376-383
Purpose:
This study compared the efficacy of heat insulation between 5% dextrose and 0.9% saline in radiofrequency ablation (RFA). Accordingly, temperature variations and maximum temperatures were assessed at identical distances and heat field distributions.
Methods:
Cubes of porcine liver tissue, measuring 10 mm across, were selected to precisely align the ablation boundary with the tissue boundary. An 18-gauge electrode with a 7-mm tip was inserted into each cube (10 per group) in a stainless-steel cup containing 40 mL of 5% dextrose or 0.9% saline. Fixed ablation was performed for 3 minutes using continuous mode at 30 W, simulating the typical thermal environment during thyroid RFA. Real-time temperature measurements were recorded by sensors positioned 0, 1, 3, and 5 mm from the cube’s edge. A comparative analysis was conducted to assess the maximum temperature, temperature variation, and duration of temperatures exceeding 42℃.
Results:
In both groups, the temperature curve declined with increasing distance from the edge of the ablated tissue. However, 0.9% saline exhibited higher maximum temperatures at 1, 3, and 5 mm compared to 5% dextrose (1 mm: 44.55°C±5.25°C vs. 34.68°C±3.07°C; 3 mm: 39.64°C±2.53°C vs. 29.22°C±2.21°C; 5 mm: 38.86°C±2.14°C vs. 28.74°C±2.51°C; all P<0.001). Considering a nerve injury threshold of 42°C, the 0.9% saline also displayed a greater proportion of samples reaching this temperature and a longer duration of temperatures exceeding it (P<0.05).
Conclusion
The heat insulation efficacy of 5% dextrose at 1-5 mm exceeds that of 0.9% saline at identical distances and in a common thermal environment during thyroid RFA.
4.In vitro heat insulation efficacy of 5% dextrose versus 0.9% saline during radiofrequency ablation
Yanping MA ; Jinfen WANG ; Tao WU ; Bowen ZHENG ; Tinghui YIN ; Yufan LIAN ; Jie REN
Ultrasonography 2024;43(5):376-383
Purpose:
This study compared the efficacy of heat insulation between 5% dextrose and 0.9% saline in radiofrequency ablation (RFA). Accordingly, temperature variations and maximum temperatures were assessed at identical distances and heat field distributions.
Methods:
Cubes of porcine liver tissue, measuring 10 mm across, were selected to precisely align the ablation boundary with the tissue boundary. An 18-gauge electrode with a 7-mm tip was inserted into each cube (10 per group) in a stainless-steel cup containing 40 mL of 5% dextrose or 0.9% saline. Fixed ablation was performed for 3 minutes using continuous mode at 30 W, simulating the typical thermal environment during thyroid RFA. Real-time temperature measurements were recorded by sensors positioned 0, 1, 3, and 5 mm from the cube’s edge. A comparative analysis was conducted to assess the maximum temperature, temperature variation, and duration of temperatures exceeding 42℃.
Results:
In both groups, the temperature curve declined with increasing distance from the edge of the ablated tissue. However, 0.9% saline exhibited higher maximum temperatures at 1, 3, and 5 mm compared to 5% dextrose (1 mm: 44.55°C±5.25°C vs. 34.68°C±3.07°C; 3 mm: 39.64°C±2.53°C vs. 29.22°C±2.21°C; 5 mm: 38.86°C±2.14°C vs. 28.74°C±2.51°C; all P<0.001). Considering a nerve injury threshold of 42°C, the 0.9% saline also displayed a greater proportion of samples reaching this temperature and a longer duration of temperatures exceeding it (P<0.05).
Conclusion
The heat insulation efficacy of 5% dextrose at 1-5 mm exceeds that of 0.9% saline at identical distances and in a common thermal environment during thyroid RFA.
5.In vitro heat insulation efficacy of 5% dextrose versus 0.9% saline during radiofrequency ablation
Yanping MA ; Jinfen WANG ; Tao WU ; Bowen ZHENG ; Tinghui YIN ; Yufan LIAN ; Jie REN
Ultrasonography 2024;43(5):376-383
Purpose:
This study compared the efficacy of heat insulation between 5% dextrose and 0.9% saline in radiofrequency ablation (RFA). Accordingly, temperature variations and maximum temperatures were assessed at identical distances and heat field distributions.
Methods:
Cubes of porcine liver tissue, measuring 10 mm across, were selected to precisely align the ablation boundary with the tissue boundary. An 18-gauge electrode with a 7-mm tip was inserted into each cube (10 per group) in a stainless-steel cup containing 40 mL of 5% dextrose or 0.9% saline. Fixed ablation was performed for 3 minutes using continuous mode at 30 W, simulating the typical thermal environment during thyroid RFA. Real-time temperature measurements were recorded by sensors positioned 0, 1, 3, and 5 mm from the cube’s edge. A comparative analysis was conducted to assess the maximum temperature, temperature variation, and duration of temperatures exceeding 42℃.
Results:
In both groups, the temperature curve declined with increasing distance from the edge of the ablated tissue. However, 0.9% saline exhibited higher maximum temperatures at 1, 3, and 5 mm compared to 5% dextrose (1 mm: 44.55°C±5.25°C vs. 34.68°C±3.07°C; 3 mm: 39.64°C±2.53°C vs. 29.22°C±2.21°C; 5 mm: 38.86°C±2.14°C vs. 28.74°C±2.51°C; all P<0.001). Considering a nerve injury threshold of 42°C, the 0.9% saline also displayed a greater proportion of samples reaching this temperature and a longer duration of temperatures exceeding it (P<0.05).
Conclusion
The heat insulation efficacy of 5% dextrose at 1-5 mm exceeds that of 0.9% saline at identical distances and in a common thermal environment during thyroid RFA.
7.Effects of Aneurysmal Neck Angle on Stent Displacement after Endovascular Repair of Abdominal Aortic Aneurysm
Yiming ZHAO ; Haoyao CAO ; Jianjin YUE ; Jiarong WANG ; Ding YUAN ; Yongtao WEI ; Tinghui ZHENG
Journal of Medical Biomechanics 2023;38(3):E472-E478
Objective To study the effects of aneurysmal neck angle on stent displacement after endovascular repair of abdominal aortic aneurysm (AAA). Methods The CT images of 28 patients were selected to establish preoperative AAA model, postoperative AAA model and covered stent model respectively, and the models were divided into non-severe angulation group ( n = 14) and severe angulation group ( n = 14) according to the preoperative angle of tumor neck. The geometric shape of each model was measured, and the changes of AAA geometric parameters and postoperative stent displacements before and after surgery were analyzed. The displacement force of the model during the first follow-up was calculated by hemodynamic simulation. Results Significant differences were found in tumor length, maximum diameter, displacement force, tumor neck length and tumor volume between two groups of patients (P<0. 05), while there was no significant difference between COG (the center of gravity) displacement and proximal displacement (P> 0. 05). For the incidence of internal leakage, there were 2 cases in non-severe angulation group and 4 cases in severe angulation group (P>0. 05).Conclusions Severe neck angulation can lead to a significant increase in support displacement force and decrease in proximal anchorage zone, and thus increase the possibility of support displacement. It is suggested that doctors should strengthen postoperative follow-up for patients with severe neck angulation and be vigilant of the occurrence of long-term internal leakage in clinic.
8.Preparation method and characteristics of biomimetic microbubbles
Tingting SHA ; Xiaoyan MIAO ; Rongqin ZHENG ; Zheng ZHANG ; Weifeng YAO ; Tinghui YIN
Chinese Journal of Ultrasonography 2022;31(2):161-168
Objective:To evaluate the feasibility and applicability of using phospholipid-hybridization method for preparing biomimetic microbubbles (Bio-MBs) ultrasound contrast agents.Methods:Leukocyte biomimetic microbubbles (MB leu), platelet biomimetic microbubbles (MB pla) and erythrocyte biomimetic microbubbles (MB ery) were prepared by multiple steps: film-hydration, phospholipid-hybridization, mechanical oscillation. The size and zeta potential of Bio-MBs were measured by dynamic light scattering. A laser scanning confocal microscopy experiment was performed to confirm the presence of membrane proteins on the shell of Bio-MBs. The fluorescence of FITC-labeled typical membrane protein was evaluated using a flow cytometer. Sodium dodecyl sulfate polyacrylamide gel electrophoresis was used to characterize the membrane protein. Biosafety of Bio-MBs was evaluated by CCK-8 counting kit, blood and major organs. The contrast enhancement effect and stability were observed in vitro and in vivo. An in vivo fluorescence imaging system was performed to evaluate the distribution of Bio-MBs. The application value of biomimetic microbubbles was measured by ultrasound molecular imaging by using ischemia-reperfusion rat models and acute hepatitis rat models. Results:Bio-MBs with spherical shape distributed homogenously, without obvious aggregation. The membrane proteins were successfully integrated into the shell of Bio-MBs.The diameter of three Bio-MBs was similar to that of control microbubbles (MB con) ( P>0.05), three Bio-MBs had a lower zeta potential than MB con ( P<0.05). The Bio-MBs had an appreciable performance in vitro and in vivo biosafety. The Bio-MBs retained the main proteins inherited from cell membrane. Contrast enhanced ultrasound imaging in vitro and in vivo showed that the Bio-MBs had a stable imaging ability.MB leu and MB pla have good targeted imaging effect in two disease models. Conclusions:A series of Bio-MBs ultrasound contrast agents, which have high stability, biosafety and targeted imaging efficiency, were successfully prepared by using phospholipid-hybridization method. This fabrication method for obtaining Bio-MBs can be applied to different clinical scenarios with different cell types in the future.
10.Comparison of stent displacement and displacement force after endovascular aneurysm repair with cross-limb or parallel-limb stent.
Jianjin YUE ; Yiming ZHAO ; Jiarong WANG ; Yubo FAN ; Tinghui ZHENG
Journal of Biomedical Engineering 2022;39(4):645-650
This study aims to investigate whether displacement force on stents can accurately represents the displacement of the stent after endovascular aneurysm repair (EVAR) by comparing the measured stent displacement with the displacement forces calculated by computational fluid dynamics (CFD). And the effect of cross-limb and parallel-limb EVAR on stent displacements is further studied. Based on our objective, in this study, ten cross-limb EVAR patients and ten parallel-limb EVAR patients in West China Hospital of Sichuan University were enrolled. Patient-specific models were first reconstructed based on the computed tomography angiography images, then the stent displacements were measured, and the displacement forces acting on the stents were calculated by CFD. Finally, the
Aortic Aneurysm, Abdominal/surgery*
;
Blood Vessel Prosthesis
;
Blood Vessel Prosthesis Implantation
;
Endovascular Procedures/methods*
;
Humans
;
Prosthesis Design
;
Retrospective Studies
;
Stents
;
Treatment Outcome