6.Enhanced accuracy in gravity-based intravenous infusion using pulse oximeter drop counting and measured single-drop weights
Daeseok OH ; Myoung Jin KO ; Jae Hwan KIM ; Yeiheum PARK ; Sungho MOON
Anesthesia and Pain Medicine 2025;20(3):230-241
Background:
Intravenous (IV) fluid therapy is essential and widely used; however, it is associated with high error rates, largely due to human factors, necessitating constant and careful monitoring by medical staff. Gravity-based systems are prone to errors, whereas electronic pumps, though more accurate, are limited by size, cost, and complexity. In this study, the impact of single-drop weight measurement and real-time light source monitoring on the accuracy of gravity-based infusion systems was evaluated.
Methods:
Gravity-based IV sets with IV infusion flow regulators (IIFRs) from three manufacturers were tested using 1,000 ml of 0.9% saline. The drops per min and the drop weight were recorded using a pulse oximeter, which served as a light source. The flow rates from the pulse oximeter group (PO) were compared with those from the manufacturer’s drop volume (C) and the IIFR groups. The mean absolute percentage error (MAPE) of predicted versus actual volumes was analyzed along with correlations between the residual volume and drop rate.
Results:
The PO group values were statistically closer to those of the actual measurements than the C and IIFR groups values (P < 0.05), demonstrating higher accuracy and lower MAPE, except at 300 ml/h when than those of the C group, independent of residual volume. The residual volume strongly correlated with the drop rate (r > 0.9).
Conclusions
Real-time drop measurements using light sources combined with single-drop weight assessment improve the accuracy of these systems. Integrating pulse oximeters into IV sets may enhance clinical precision and reduce provider workload.
7.Comprehensive ultrasonographic evaluation of normal and fibrotic kidneys in a mouse model with an ultra-high-frequency transducer
Myoung Seok LEE ; Jeong Yeon CHO ; Min Hoan MOON ; Jeonghwan LEE ; Jung Pyo LEE ; Nayeon SHIN ; Wencheng JIN ; Ara CHO
Ultrasonography 2024;43(5):314-326
Purpose:
This study aimed to establish baseline morphological and functional data for normal mouse kidneys via a clinical 33 MHz ultra-high-frequency (UHF) transducer, compare the data with the findings from fibrotic mice, and assess correlations between ultrasonography (US) parameters and fibrosis-related markers.
Methods:
This retrospective study aggregated data from three separate experiments (obstructive nephropathy, diabetic nephropathy, and acute-to-chronic kidney injury models). Morphological parameters (kidney size, parenchymal thickness [PT]) and functional (shear-wave speed [SWS], stiffness, resistive index [RI], and microvascular imaging-derived vascular index [VI]) were assessed and compared between normal and fibrotic mouse kidneys. Semi-quantitative histopathologic scores were calculated and molecular markers (epithelial cadherin), Collagen 1A1 [Col1A1], transforming growth factor-β, and α-smooth muscle actin [α-SMA]) were evaluated using western blots. Correlations with US parameters were explored.
Results:
Clinical UHF US successfully imaged the kidneys of the experimental mice. A three-layer configuration was prevalent in the normal mouse kidney parenchyma (34/35) but was blurred in most fibrotic mouse kidneys (33/40). US parameters, including size (11.14 vs. 10.70 mm), PT (2.07 vs. 1.24 mm), RI (0.64 vs. 0.77), VI (22.55% vs. 11.47%, only for non-obstructive kidneys), SWS (1.67 vs. 2.06 m/s), and stiffness (8.23 vs. 12.92 kPa), showed significant differences between normal and fibrotic kidneys (P<0.001). These parameters also demonstrated strong discriminative ability in receiver operating characteristic curve analysis (area under the curve, 0.76 to 0.95; P<0.001). PT, VI, and RI were significantly correlated with histological fibrosis markers (ρ=-0.64 to -0.68 for PT and VI, ρ=0.71-0.76 for RI, P<0.001). VI exhibited strong negative correlations with Col1A1 (ρ=-0.76, P=0.006) and α-SMA (ρ=-0.75, P=0.009).
Conclusion
Clinical UHF US effectively distinguished normal and fibrotic mouse kidneys, indicating the potential of US parameters, notably VI, as noninvasive markers for tracking fibrosis initiation and progression in mouse kidney fibrosis models.
8.Comprehensive ultrasonographic evaluation of normal and fibrotic kidneys in a mouse model with an ultra-high-frequency transducer
Myoung Seok LEE ; Jeong Yeon CHO ; Min Hoan MOON ; Jeonghwan LEE ; Jung Pyo LEE ; Nayeon SHIN ; Wencheng JIN ; Ara CHO
Ultrasonography 2024;43(5):314-326
Purpose:
This study aimed to establish baseline morphological and functional data for normal mouse kidneys via a clinical 33 MHz ultra-high-frequency (UHF) transducer, compare the data with the findings from fibrotic mice, and assess correlations between ultrasonography (US) parameters and fibrosis-related markers.
Methods:
This retrospective study aggregated data from three separate experiments (obstructive nephropathy, diabetic nephropathy, and acute-to-chronic kidney injury models). Morphological parameters (kidney size, parenchymal thickness [PT]) and functional (shear-wave speed [SWS], stiffness, resistive index [RI], and microvascular imaging-derived vascular index [VI]) were assessed and compared between normal and fibrotic mouse kidneys. Semi-quantitative histopathologic scores were calculated and molecular markers (epithelial cadherin), Collagen 1A1 [Col1A1], transforming growth factor-β, and α-smooth muscle actin [α-SMA]) were evaluated using western blots. Correlations with US parameters were explored.
Results:
Clinical UHF US successfully imaged the kidneys of the experimental mice. A three-layer configuration was prevalent in the normal mouse kidney parenchyma (34/35) but was blurred in most fibrotic mouse kidneys (33/40). US parameters, including size (11.14 vs. 10.70 mm), PT (2.07 vs. 1.24 mm), RI (0.64 vs. 0.77), VI (22.55% vs. 11.47%, only for non-obstructive kidneys), SWS (1.67 vs. 2.06 m/s), and stiffness (8.23 vs. 12.92 kPa), showed significant differences between normal and fibrotic kidneys (P<0.001). These parameters also demonstrated strong discriminative ability in receiver operating characteristic curve analysis (area under the curve, 0.76 to 0.95; P<0.001). PT, VI, and RI were significantly correlated with histological fibrosis markers (ρ=-0.64 to -0.68 for PT and VI, ρ=0.71-0.76 for RI, P<0.001). VI exhibited strong negative correlations with Col1A1 (ρ=-0.76, P=0.006) and α-SMA (ρ=-0.75, P=0.009).
Conclusion
Clinical UHF US effectively distinguished normal and fibrotic mouse kidneys, indicating the potential of US parameters, notably VI, as noninvasive markers for tracking fibrosis initiation and progression in mouse kidney fibrosis models.
9.Comprehensive ultrasonographic evaluation of normal and fibrotic kidneys in a mouse model with an ultra-high-frequency transducer
Myoung Seok LEE ; Jeong Yeon CHO ; Min Hoan MOON ; Jeonghwan LEE ; Jung Pyo LEE ; Nayeon SHIN ; Wencheng JIN ; Ara CHO
Ultrasonography 2024;43(5):314-326
Purpose:
This study aimed to establish baseline morphological and functional data for normal mouse kidneys via a clinical 33 MHz ultra-high-frequency (UHF) transducer, compare the data with the findings from fibrotic mice, and assess correlations between ultrasonography (US) parameters and fibrosis-related markers.
Methods:
This retrospective study aggregated data from three separate experiments (obstructive nephropathy, diabetic nephropathy, and acute-to-chronic kidney injury models). Morphological parameters (kidney size, parenchymal thickness [PT]) and functional (shear-wave speed [SWS], stiffness, resistive index [RI], and microvascular imaging-derived vascular index [VI]) were assessed and compared between normal and fibrotic mouse kidneys. Semi-quantitative histopathologic scores were calculated and molecular markers (epithelial cadherin), Collagen 1A1 [Col1A1], transforming growth factor-β, and α-smooth muscle actin [α-SMA]) were evaluated using western blots. Correlations with US parameters were explored.
Results:
Clinical UHF US successfully imaged the kidneys of the experimental mice. A three-layer configuration was prevalent in the normal mouse kidney parenchyma (34/35) but was blurred in most fibrotic mouse kidneys (33/40). US parameters, including size (11.14 vs. 10.70 mm), PT (2.07 vs. 1.24 mm), RI (0.64 vs. 0.77), VI (22.55% vs. 11.47%, only for non-obstructive kidneys), SWS (1.67 vs. 2.06 m/s), and stiffness (8.23 vs. 12.92 kPa), showed significant differences between normal and fibrotic kidneys (P<0.001). These parameters also demonstrated strong discriminative ability in receiver operating characteristic curve analysis (area under the curve, 0.76 to 0.95; P<0.001). PT, VI, and RI were significantly correlated with histological fibrosis markers (ρ=-0.64 to -0.68 for PT and VI, ρ=0.71-0.76 for RI, P<0.001). VI exhibited strong negative correlations with Col1A1 (ρ=-0.76, P=0.006) and α-SMA (ρ=-0.75, P=0.009).
Conclusion
Clinical UHF US effectively distinguished normal and fibrotic mouse kidneys, indicating the potential of US parameters, notably VI, as noninvasive markers for tracking fibrosis initiation and progression in mouse kidney fibrosis models.
10.Comprehensive ultrasonographic evaluation of normal and fibrotic kidneys in a mouse model with an ultra-high-frequency transducer
Myoung Seok LEE ; Jeong Yeon CHO ; Min Hoan MOON ; Jeonghwan LEE ; Jung Pyo LEE ; Nayeon SHIN ; Wencheng JIN ; Ara CHO
Ultrasonography 2024;43(5):314-326
Purpose:
This study aimed to establish baseline morphological and functional data for normal mouse kidneys via a clinical 33 MHz ultra-high-frequency (UHF) transducer, compare the data with the findings from fibrotic mice, and assess correlations between ultrasonography (US) parameters and fibrosis-related markers.
Methods:
This retrospective study aggregated data from three separate experiments (obstructive nephropathy, diabetic nephropathy, and acute-to-chronic kidney injury models). Morphological parameters (kidney size, parenchymal thickness [PT]) and functional (shear-wave speed [SWS], stiffness, resistive index [RI], and microvascular imaging-derived vascular index [VI]) were assessed and compared between normal and fibrotic mouse kidneys. Semi-quantitative histopathologic scores were calculated and molecular markers (epithelial cadherin), Collagen 1A1 [Col1A1], transforming growth factor-β, and α-smooth muscle actin [α-SMA]) were evaluated using western blots. Correlations with US parameters were explored.
Results:
Clinical UHF US successfully imaged the kidneys of the experimental mice. A three-layer configuration was prevalent in the normal mouse kidney parenchyma (34/35) but was blurred in most fibrotic mouse kidneys (33/40). US parameters, including size (11.14 vs. 10.70 mm), PT (2.07 vs. 1.24 mm), RI (0.64 vs. 0.77), VI (22.55% vs. 11.47%, only for non-obstructive kidneys), SWS (1.67 vs. 2.06 m/s), and stiffness (8.23 vs. 12.92 kPa), showed significant differences between normal and fibrotic kidneys (P<0.001). These parameters also demonstrated strong discriminative ability in receiver operating characteristic curve analysis (area under the curve, 0.76 to 0.95; P<0.001). PT, VI, and RI were significantly correlated with histological fibrosis markers (ρ=-0.64 to -0.68 for PT and VI, ρ=0.71-0.76 for RI, P<0.001). VI exhibited strong negative correlations with Col1A1 (ρ=-0.76, P=0.006) and α-SMA (ρ=-0.75, P=0.009).
Conclusion
Clinical UHF US effectively distinguished normal and fibrotic mouse kidneys, indicating the potential of US parameters, notably VI, as noninvasive markers for tracking fibrosis initiation and progression in mouse kidney fibrosis models.

Result Analysis
Print
Save
E-mail