1.Validation of Devices for the Five Times Sit To Stand Test:Comparing Plantar Pressure and Head Motion Analysis with Manual Measurement
Sanghyun JEE ; Chan Woong JANG ; Kyoungmin PARK ; Sanghoon SHIN ; Min-Chul PAEK ; Jung Hyun PARK
Yonsei Medical Journal 2025;66(1):51-57
Purpose:
This study aims to evaluate a new method for the five times sit to stand test (FTSST), crucial for addressing frailty in an aging population. It utilizes a smart insole for plantar pressure analysis and a marker-less motion capture device for head height analysis.
Materials and Methods:
Thirty-five participants aged 50 years or older underwent FTSST assessment using three methods: manual measurement with a stopwatch (FTSST-M), plantar pressure analysis with smart insoles (FTSST-P), and head height analysis with a marker-less motion capture device (FTSST-H). Simultaneous measurements using three methods were done. Correlation between results of these methods were analyzed using intraclass correlation coefficient (ICC) and κ coefficient. Comprehensive clinical examinations were conducted with ethical approval.
Results:
Participants’ mean scores for FTSST-M, FTSST-P, and FTSST-H were 2.43±1.20, 2.43±1.29, and 2.37±1.31, respectively. Correlations of the times and corresponding scores between FTSST-P and FTSST-M, as well as FTSST-H and FTSST-M, exceeded 0.9 (ICC and κ coefficients, p<0.001). Using an FTSST score of 3 or less to indicate vulnerability, the κ value for vulnerability classification between two measurements was 0.886 (p<0.001).
Conclusion
This study showed strong correlation between FTSST results using smart insoles and marker-less motion capture, compared to conventional methods. These findings highlight the potential of these technologies for precise FTSST measurements, offering convenience and cost-effectiveness. Simultaneous use of these devices enables diverse analyses, enhancing our understanding of frailty.
2.Validation of Devices for the Five Times Sit To Stand Test:Comparing Plantar Pressure and Head Motion Analysis with Manual Measurement
Sanghyun JEE ; Chan Woong JANG ; Kyoungmin PARK ; Sanghoon SHIN ; Min-Chul PAEK ; Jung Hyun PARK
Yonsei Medical Journal 2025;66(1):51-57
Purpose:
This study aims to evaluate a new method for the five times sit to stand test (FTSST), crucial for addressing frailty in an aging population. It utilizes a smart insole for plantar pressure analysis and a marker-less motion capture device for head height analysis.
Materials and Methods:
Thirty-five participants aged 50 years or older underwent FTSST assessment using three methods: manual measurement with a stopwatch (FTSST-M), plantar pressure analysis with smart insoles (FTSST-P), and head height analysis with a marker-less motion capture device (FTSST-H). Simultaneous measurements using three methods were done. Correlation between results of these methods were analyzed using intraclass correlation coefficient (ICC) and κ coefficient. Comprehensive clinical examinations were conducted with ethical approval.
Results:
Participants’ mean scores for FTSST-M, FTSST-P, and FTSST-H were 2.43±1.20, 2.43±1.29, and 2.37±1.31, respectively. Correlations of the times and corresponding scores between FTSST-P and FTSST-M, as well as FTSST-H and FTSST-M, exceeded 0.9 (ICC and κ coefficients, p<0.001). Using an FTSST score of 3 or less to indicate vulnerability, the κ value for vulnerability classification between two measurements was 0.886 (p<0.001).
Conclusion
This study showed strong correlation between FTSST results using smart insoles and marker-less motion capture, compared to conventional methods. These findings highlight the potential of these technologies for precise FTSST measurements, offering convenience and cost-effectiveness. Simultaneous use of these devices enables diverse analyses, enhancing our understanding of frailty.
3.Validation of Devices for the Five Times Sit To Stand Test:Comparing Plantar Pressure and Head Motion Analysis with Manual Measurement
Sanghyun JEE ; Chan Woong JANG ; Kyoungmin PARK ; Sanghoon SHIN ; Min-Chul PAEK ; Jung Hyun PARK
Yonsei Medical Journal 2025;66(1):51-57
Purpose:
This study aims to evaluate a new method for the five times sit to stand test (FTSST), crucial for addressing frailty in an aging population. It utilizes a smart insole for plantar pressure analysis and a marker-less motion capture device for head height analysis.
Materials and Methods:
Thirty-five participants aged 50 years or older underwent FTSST assessment using three methods: manual measurement with a stopwatch (FTSST-M), plantar pressure analysis with smart insoles (FTSST-P), and head height analysis with a marker-less motion capture device (FTSST-H). Simultaneous measurements using three methods were done. Correlation between results of these methods were analyzed using intraclass correlation coefficient (ICC) and κ coefficient. Comprehensive clinical examinations were conducted with ethical approval.
Results:
Participants’ mean scores for FTSST-M, FTSST-P, and FTSST-H were 2.43±1.20, 2.43±1.29, and 2.37±1.31, respectively. Correlations of the times and corresponding scores between FTSST-P and FTSST-M, as well as FTSST-H and FTSST-M, exceeded 0.9 (ICC and κ coefficients, p<0.001). Using an FTSST score of 3 or less to indicate vulnerability, the κ value for vulnerability classification between two measurements was 0.886 (p<0.001).
Conclusion
This study showed strong correlation between FTSST results using smart insoles and marker-less motion capture, compared to conventional methods. These findings highlight the potential of these technologies for precise FTSST measurements, offering convenience and cost-effectiveness. Simultaneous use of these devices enables diverse analyses, enhancing our understanding of frailty.
4.Validation of Devices for the Five Times Sit To Stand Test:Comparing Plantar Pressure and Head Motion Analysis with Manual Measurement
Sanghyun JEE ; Chan Woong JANG ; Kyoungmin PARK ; Sanghoon SHIN ; Min-Chul PAEK ; Jung Hyun PARK
Yonsei Medical Journal 2025;66(1):51-57
Purpose:
This study aims to evaluate a new method for the five times sit to stand test (FTSST), crucial for addressing frailty in an aging population. It utilizes a smart insole for plantar pressure analysis and a marker-less motion capture device for head height analysis.
Materials and Methods:
Thirty-five participants aged 50 years or older underwent FTSST assessment using three methods: manual measurement with a stopwatch (FTSST-M), plantar pressure analysis with smart insoles (FTSST-P), and head height analysis with a marker-less motion capture device (FTSST-H). Simultaneous measurements using three methods were done. Correlation between results of these methods were analyzed using intraclass correlation coefficient (ICC) and κ coefficient. Comprehensive clinical examinations were conducted with ethical approval.
Results:
Participants’ mean scores for FTSST-M, FTSST-P, and FTSST-H were 2.43±1.20, 2.43±1.29, and 2.37±1.31, respectively. Correlations of the times and corresponding scores between FTSST-P and FTSST-M, as well as FTSST-H and FTSST-M, exceeded 0.9 (ICC and κ coefficients, p<0.001). Using an FTSST score of 3 or less to indicate vulnerability, the κ value for vulnerability classification between two measurements was 0.886 (p<0.001).
Conclusion
This study showed strong correlation between FTSST results using smart insoles and marker-less motion capture, compared to conventional methods. These findings highlight the potential of these technologies for precise FTSST measurements, offering convenience and cost-effectiveness. Simultaneous use of these devices enables diverse analyses, enhancing our understanding of frailty.
5.Validation of Devices for the Five Times Sit To Stand Test:Comparing Plantar Pressure and Head Motion Analysis with Manual Measurement
Sanghyun JEE ; Chan Woong JANG ; Kyoungmin PARK ; Sanghoon SHIN ; Min-Chul PAEK ; Jung Hyun PARK
Yonsei Medical Journal 2025;66(1):51-57
Purpose:
This study aims to evaluate a new method for the five times sit to stand test (FTSST), crucial for addressing frailty in an aging population. It utilizes a smart insole for plantar pressure analysis and a marker-less motion capture device for head height analysis.
Materials and Methods:
Thirty-five participants aged 50 years or older underwent FTSST assessment using three methods: manual measurement with a stopwatch (FTSST-M), plantar pressure analysis with smart insoles (FTSST-P), and head height analysis with a marker-less motion capture device (FTSST-H). Simultaneous measurements using three methods were done. Correlation between results of these methods were analyzed using intraclass correlation coefficient (ICC) and κ coefficient. Comprehensive clinical examinations were conducted with ethical approval.
Results:
Participants’ mean scores for FTSST-M, FTSST-P, and FTSST-H were 2.43±1.20, 2.43±1.29, and 2.37±1.31, respectively. Correlations of the times and corresponding scores between FTSST-P and FTSST-M, as well as FTSST-H and FTSST-M, exceeded 0.9 (ICC and κ coefficients, p<0.001). Using an FTSST score of 3 or less to indicate vulnerability, the κ value for vulnerability classification between two measurements was 0.886 (p<0.001).
Conclusion
This study showed strong correlation between FTSST results using smart insoles and marker-less motion capture, compared to conventional methods. These findings highlight the potential of these technologies for precise FTSST measurements, offering convenience and cost-effectiveness. Simultaneous use of these devices enables diverse analyses, enhancing our understanding of frailty.
6.Complication of the Cervical Thrust Technique in a Patient with Athetoid Cerebral Palsy: A Case Report
Myungsang KIM ; Han Eol CHO ; Min Chul PAEK ; Jung Hyun PARK
Clinical Pain 2022;21(1):38-40
Manual therapy has been widely practiced in almost all countries worldwide. With its potential complications, there are contraindications for thrust technique. A 56-year-old woman diagnosed with athetoid cerebral palsy visited a local clinic due to a tingling sensation in right hand that lasted for a year and aggravated to both upper extremities. The patient underwent three sessions of cervical manual therapy including thrust technique. No evaluation was performed. The patient immediately felt weakness in all extremities after performing cervical thrust technique. Magnetic resonance imaging showed atlantoaxial instability and cervical myelopathy. Approximately one-third of adults with cerebral palsy reported chronic musculoskeletal pain and they often experience neck pain. Particularly in athetoid cerebral palsy, malalignment or instability of the cervical spine is prevalent and often results in myelopathy. Therefore, musculoskeletal evaluation is necessary to identify cervical instability in case of cervical thrust technique, and it should be performed by relevant medical professionals.
7.Ultrasound Guided Thoracic Paravertebral Space Block for Chronic Intractable Upper Back Pain
Myungsang KIM ; Min Chul PAEK ; Han Eol CHO ; Jung Hyun PARK
Clinical Pain 2021;20(2):141-144
There are some cases of myofascial pain syndrome (MPS) with chronic upper back pain that does not respond to dry needling or trigger point injection, well-known treatments for MPS. A 67-year-old female developed a stabbing upper back pain with trigger point at left T7∼8 levels 10 years ago. She complained of the pain with Numeral Rating Scale (NRS) 8 points. Myofascial release technique and trigger point injection had no effect. Under ultrasound guidance 20 ml of 1% lidocaine was injected into thoracic paravertebral space. Immediately, the pain was reduced to NRS 4 points. One week later, the second block was performed in the same way as the first, and the pain was reduced to NRS 2 points. The stabbing pain disappeared, and oral opioids were discontinued. Ultrasound guided thoracic paravertebral space block is an effective and safe treatment for refractory MPS with chronic upper back pain.
8.Erratum: Correction of Author Name and Affiliation in the Article “Artificial Intelligence in Health Care: Current Applications and Issues”
Chan-Woo PARK ; Sung Wook SEO ; Noeul KANG ; BeomSeok KO ; Byung Wook CHOI ; Chang Min PARK ; Dong Kyung CHANG ; Hwiyoung KIM ; Hyunchul KIM ; Hyunna LEE ; Jinhee JANG ; Jong Chul YE ; Jong Hong JEON ; Joon Beom SEO ; Kwang Joon KIM ; Kyu-Hwan JUNG ; Namkug KIM ; Seungwook PAEK ; Soo-Yong SHIN ; Soyoung YOO ; Yoon Sup CHOI ; Youngjun KIM ; Hyung-Jin YOON
Journal of Korean Medical Science 2020;35(48):e425-
9.Artificial Intelligence in Health Care: Current Applications and Issues
Chan-Woo PARK ; Sung Wook SEO ; Noeul KANG ; Beom Seok KO ; Byung Wook CHOI ; Chang Min PARK ; Dong Kyung CHANG ; Hwiuoung KIM ; Hyun chul KIM ; Hyun na LEE ; Jin hee JANG ; Jong Chul YE ; Jong Hong JEON ; Joon Beom SEO ; Kwang Joon KIM ; Kyu-Hwan JUNG ; Namkug KIM ; Seung wook PAEK ; Soo-Yong SHIN ; So young YOO ; Yoon Sup CHOI ; Youngjun KIM ; Hyung-Jin YOON
Journal of Korean Medical Science 2020;35(42):e379-
In recent years, artificial intelligence (AI) technologies have greatly advanced and become a reality in many areas of our daily lives. In the health care field, numerous efforts are being made to implement the AI technology for practical medical treatments. With the rapid developments in machine learning algorithms and improvements in hardware performances, the AI technology is expected to play an important role in effectively analyzing and utilizing extensive amounts of health and medical data. However, the AI technology has various unique characteristics that are different from the existing health care technologies. Subsequently, there are a number of areas that need to be supplemented within the current health care system for the AI to be utilized more effectively and frequently in health care. In addition, the number of medical practitioners and public that accept AI in the health care is still low;moreover, there are various concerns regarding the safety and reliability of AI technologyimplementations. Therefore, this paper aims to introduce the current research and application status of AI technology in health care and discuss the issues that need to be resolved.
10.Intensity-modulated radiotherapy for stage I glottic cancer: a short-term outcomes compared with three-dimensional conformal radiotherapy
Ick Joon CHO ; Woong Ki CHUNG ; Joon Kyoo LEE ; Min Cheol LEE ; Jayeong PAEK ; Yong Hyub KIM ; Jae Uk JEONG ; Mee Sun YOON ; Ju Young SONG ; Taek Keun NAM ; Sung Ja AHN ; Dong Hoon LEE ; Tae Mi YOON ; Sang Chul LIM
Radiation Oncology Journal 2019;37(4):271-278
PURPOSE: To investigate the differences in treatment outcomes between two radiation techniques, intensity-modulated radiotherapy (IMRT) and three-dimensional conformal radiotherapy (3DCRT).MATERIALS AND METHODS: We retrospectively analyzed 160 (IMRT = 23, 3DCRT = 137) patients with stage I glottic cancer treated from January 2005 through December 2016. The IMRT was performed with TomoTherapy (16 patients), volumetric-modulated arc therapy (6 patients), and step-and-shoot technique (1 patient), respectively. The 3DCRT was performed with bilateral parallel opposing fields. The median follow-up duration was 30 months (range, 31 to 42 months) in the IMRT group and 65 months (range, 20 to 143 months) in the 3DCRT group.RESULTS: The 5-year overall survival and 3-year local control rates of the 160 patients were 95.7% and 91.4%, respectively. There was no significant difference in 3-year local control rates between the IMRT and 3DCRT groups (94.4% vs. 91.0%; p = 0.587). Thirteen of 137 patients in the 3DCRT group had recurrences. In the IMRT group, one patient had a recurrence at the true vocal cord. Patients treated with IMRT had less grade 2 skin reaction than the 3DCRT group, but this had no statistical significance (4.3% vs. 21.2%; p = 0.080).CONCLUSION: IMRT had comparable outcomes with 3DCRT, and a trend of less acute skin reaction in stage I glottic cancer patients
Follow-Up Studies
;
Glottis
;
Humans
;
Laryngeal Neoplasms
;
Radiotherapy, Conformal
;
Radiotherapy, Intensity-Modulated
;
Recurrence
;
Retrospective Studies
;
Skin
;
Vocal Cords

Result Analysis
Print
Save
E-mail