1.Reinforcing treatment and evaluation workflow of stereotactic ablative body radiotherapy for refractory ventricular tachycardia
Hojin KIM ; Sangjoon PARK ; Jihun KIM ; Jin Sung KIM ; Dong Wook KIM ; Nalee KIM ; Jae-Sun UHM ; Daehoon KIM ; Hui-Nam PAK ; Chae-Seon HONG ; Hong In YOON
Radiation Oncology Journal 2024;42(4):319-329
Purpose:
Cardiac radioablation is a novel, non-invasive treatment for ventricular tachycardia (VT), involving a single fractional stereotactic ablative body radiotherapy (SABR) session with a prescribed dose of 25 Gy. This complex procedure requires a detailed workflow and stringent dose constraints compared to conventional radiation therapy. This study aims to establish a consistent institutional workflow for single-fraction cardiac VT-SABR, emphasizing robust plan evaluation and quality assurance.
Materials and Methods:
The study developed a consistent institutional workflow for VT-SABR, including computed tomography (CT) simulation, target volume definition, treatment planning, robust plan evaluation, quality assurance, and image-guided strategy. The workflow was implemented for two patients with cardiac arrhythmia. Accurate target volume definition using planning CT images and electronic anatomical mapping was critical. A four-dimensional (4D) cone-beam CT (CBCT) and breath-hold electrocardiographic gated CT images reliably detected target motion.
Results:
The resulting plans exhibited a conformity index greater than 0.7 and a gradient index around G4.0. Dose constraints for the planning target volume (PTV) aimed for 95% or higher PTV dose coverage, with a maximum dose of 200% or lower. However, one case did not meet the PTV dose coverage due to the proximity of the PTV to gastrointestinal organs. Plans adhered to dose constraints for organs at risk near the heart, but meeting constraints for specific cardiac sub-structures was challenging and dependent on PTV location.
Conclusion
The plans demonstrated robustness against respiratory motion and patient positional uncertainty through a robust evaluation function. The 4D and intra-fractional CBCT were effective in verifying target motion and setup stability.
2.Reinforcing treatment and evaluation workflow of stereotactic ablative body radiotherapy for refractory ventricular tachycardia
Hojin KIM ; Sangjoon PARK ; Jihun KIM ; Jin Sung KIM ; Dong Wook KIM ; Nalee KIM ; Jae-Sun UHM ; Daehoon KIM ; Hui-Nam PAK ; Chae-Seon HONG ; Hong In YOON
Radiation Oncology Journal 2024;42(4):319-329
Purpose:
Cardiac radioablation is a novel, non-invasive treatment for ventricular tachycardia (VT), involving a single fractional stereotactic ablative body radiotherapy (SABR) session with a prescribed dose of 25 Gy. This complex procedure requires a detailed workflow and stringent dose constraints compared to conventional radiation therapy. This study aims to establish a consistent institutional workflow for single-fraction cardiac VT-SABR, emphasizing robust plan evaluation and quality assurance.
Materials and Methods:
The study developed a consistent institutional workflow for VT-SABR, including computed tomography (CT) simulation, target volume definition, treatment planning, robust plan evaluation, quality assurance, and image-guided strategy. The workflow was implemented for two patients with cardiac arrhythmia. Accurate target volume definition using planning CT images and electronic anatomical mapping was critical. A four-dimensional (4D) cone-beam CT (CBCT) and breath-hold electrocardiographic gated CT images reliably detected target motion.
Results:
The resulting plans exhibited a conformity index greater than 0.7 and a gradient index around G4.0. Dose constraints for the planning target volume (PTV) aimed for 95% or higher PTV dose coverage, with a maximum dose of 200% or lower. However, one case did not meet the PTV dose coverage due to the proximity of the PTV to gastrointestinal organs. Plans adhered to dose constraints for organs at risk near the heart, but meeting constraints for specific cardiac sub-structures was challenging and dependent on PTV location.
Conclusion
The plans demonstrated robustness against respiratory motion and patient positional uncertainty through a robust evaluation function. The 4D and intra-fractional CBCT were effective in verifying target motion and setup stability.
3.Reinforcing treatment and evaluation workflow of stereotactic ablative body radiotherapy for refractory ventricular tachycardia
Hojin KIM ; Sangjoon PARK ; Jihun KIM ; Jin Sung KIM ; Dong Wook KIM ; Nalee KIM ; Jae-Sun UHM ; Daehoon KIM ; Hui-Nam PAK ; Chae-Seon HONG ; Hong In YOON
Radiation Oncology Journal 2024;42(4):319-329
Purpose:
Cardiac radioablation is a novel, non-invasive treatment for ventricular tachycardia (VT), involving a single fractional stereotactic ablative body radiotherapy (SABR) session with a prescribed dose of 25 Gy. This complex procedure requires a detailed workflow and stringent dose constraints compared to conventional radiation therapy. This study aims to establish a consistent institutional workflow for single-fraction cardiac VT-SABR, emphasizing robust plan evaluation and quality assurance.
Materials and Methods:
The study developed a consistent institutional workflow for VT-SABR, including computed tomography (CT) simulation, target volume definition, treatment planning, robust plan evaluation, quality assurance, and image-guided strategy. The workflow was implemented for two patients with cardiac arrhythmia. Accurate target volume definition using planning CT images and electronic anatomical mapping was critical. A four-dimensional (4D) cone-beam CT (CBCT) and breath-hold electrocardiographic gated CT images reliably detected target motion.
Results:
The resulting plans exhibited a conformity index greater than 0.7 and a gradient index around G4.0. Dose constraints for the planning target volume (PTV) aimed for 95% or higher PTV dose coverage, with a maximum dose of 200% or lower. However, one case did not meet the PTV dose coverage due to the proximity of the PTV to gastrointestinal organs. Plans adhered to dose constraints for organs at risk near the heart, but meeting constraints for specific cardiac sub-structures was challenging and dependent on PTV location.
Conclusion
The plans demonstrated robustness against respiratory motion and patient positional uncertainty through a robust evaluation function. The 4D and intra-fractional CBCT were effective in verifying target motion and setup stability.
4.Reinforcing treatment and evaluation workflow of stereotactic ablative body radiotherapy for refractory ventricular tachycardia
Hojin KIM ; Sangjoon PARK ; Jihun KIM ; Jin Sung KIM ; Dong Wook KIM ; Nalee KIM ; Jae-Sun UHM ; Daehoon KIM ; Hui-Nam PAK ; Chae-Seon HONG ; Hong In YOON
Radiation Oncology Journal 2024;42(4):319-329
Purpose:
Cardiac radioablation is a novel, non-invasive treatment for ventricular tachycardia (VT), involving a single fractional stereotactic ablative body radiotherapy (SABR) session with a prescribed dose of 25 Gy. This complex procedure requires a detailed workflow and stringent dose constraints compared to conventional radiation therapy. This study aims to establish a consistent institutional workflow for single-fraction cardiac VT-SABR, emphasizing robust plan evaluation and quality assurance.
Materials and Methods:
The study developed a consistent institutional workflow for VT-SABR, including computed tomography (CT) simulation, target volume definition, treatment planning, robust plan evaluation, quality assurance, and image-guided strategy. The workflow was implemented for two patients with cardiac arrhythmia. Accurate target volume definition using planning CT images and electronic anatomical mapping was critical. A four-dimensional (4D) cone-beam CT (CBCT) and breath-hold electrocardiographic gated CT images reliably detected target motion.
Results:
The resulting plans exhibited a conformity index greater than 0.7 and a gradient index around G4.0. Dose constraints for the planning target volume (PTV) aimed for 95% or higher PTV dose coverage, with a maximum dose of 200% or lower. However, one case did not meet the PTV dose coverage due to the proximity of the PTV to gastrointestinal organs. Plans adhered to dose constraints for organs at risk near the heart, but meeting constraints for specific cardiac sub-structures was challenging and dependent on PTV location.
Conclusion
The plans demonstrated robustness against respiratory motion and patient positional uncertainty through a robust evaluation function. The 4D and intra-fractional CBCT were effective in verifying target motion and setup stability.
5.Reinforcing treatment and evaluation workflow of stereotactic ablative body radiotherapy for refractory ventricular tachycardia
Hojin KIM ; Sangjoon PARK ; Jihun KIM ; Jin Sung KIM ; Dong Wook KIM ; Nalee KIM ; Jae-Sun UHM ; Daehoon KIM ; Hui-Nam PAK ; Chae-Seon HONG ; Hong In YOON
Radiation Oncology Journal 2024;42(4):319-329
Purpose:
Cardiac radioablation is a novel, non-invasive treatment for ventricular tachycardia (VT), involving a single fractional stereotactic ablative body radiotherapy (SABR) session with a prescribed dose of 25 Gy. This complex procedure requires a detailed workflow and stringent dose constraints compared to conventional radiation therapy. This study aims to establish a consistent institutional workflow for single-fraction cardiac VT-SABR, emphasizing robust plan evaluation and quality assurance.
Materials and Methods:
The study developed a consistent institutional workflow for VT-SABR, including computed tomography (CT) simulation, target volume definition, treatment planning, robust plan evaluation, quality assurance, and image-guided strategy. The workflow was implemented for two patients with cardiac arrhythmia. Accurate target volume definition using planning CT images and electronic anatomical mapping was critical. A four-dimensional (4D) cone-beam CT (CBCT) and breath-hold electrocardiographic gated CT images reliably detected target motion.
Results:
The resulting plans exhibited a conformity index greater than 0.7 and a gradient index around G4.0. Dose constraints for the planning target volume (PTV) aimed for 95% or higher PTV dose coverage, with a maximum dose of 200% or lower. However, one case did not meet the PTV dose coverage due to the proximity of the PTV to gastrointestinal organs. Plans adhered to dose constraints for organs at risk near the heart, but meeting constraints for specific cardiac sub-structures was challenging and dependent on PTV location.
Conclusion
The plans demonstrated robustness against respiratory motion and patient positional uncertainty through a robust evaluation function. The 4D and intra-fractional CBCT were effective in verifying target motion and setup stability.
6.Machine learning based potentiating impacts of 12‑lead ECG for classifying paroxysmal versus non‑paroxysmal atrial fibrillation
Sungsoo KIM ; Sohee KWON ; Mia K. MARKEY ; Alan C. BOVIK ; Sung‑Hwi HONG ; JunYong KIM ; Hye Jin HWANG ; Boyoung JOUNG ; Hui‑Nam PAK ; Moon‑Hyeong LEE ; Junbeom PARK
International Journal of Arrhythmia 2022;23(2):11-
Background:
Conventional modality requires several days observation by Holter monitor to differentiate atrial fibril‑ lation (AF) between Paroxysmal atrial fibrillation (PAF) and Non-paroxysmal atrial fibrillation (Non-PAF). Rapid and practical differentiating approach is needed.
Objective:
To develop a machine learning model that observes 10-s of standard 12-lead electrocardiograph (ECG) for real-time classification of AF between PAF versus Non-PAF.
Methods:
In this multicenter, retrospective cohort study, the model training and cross-validation was performed on a dataset consisting of 741 patients enrolled from Severance Hospital, South Korea. For cross-institutional validation, the trained model was applied to an independent data set of 600 patients enrolled from Ewha University Hospital, South Korea. Lasso regression was applied to develop the model.
Results:
In the primary analysis, the Area Under the Receiver Operating Characteristic Curve (AUC) on the test set for the model that predicted AF subtype only using ECG was 0.72 (95% CI 0.65–0.80). In the secondary analysis, AUC only using baseline characteristics was 0.53 (95% CI 0.45–0.61), while the model that employed both baseline characteris‑ tics and ECG parameters was 0.72 (95% CI 0.65–0.80). Moreover, the model that incorporated baseline characteristics, ECG, and Echocardiographic parameters achieved an AUC of 0.76 (95% CI 0.678–0.855) on the test set.
Conclusions
Our machine learning model using ECG has potential for automatic differentiation of AF between PAF versus Non-PAF achieving high accuracy. The inclusion of Echocardiographic parameters further increases model per‑ formance. Further studies are needed to clarify the next steps towards clinical translation of the proposed algorithm.
7.Added Value of Contrast Leakage Information over the CBV Value of DSC Perfusion MRI to Differentiate between Pseudoprogression and True Progression after Concurrent Chemoradiotherapy in Glioblastoma Patients
Elena PAK ; Seung Hong CHOI ; Chul-Kee PARK ; Tae Min KIM ; Sung-Hye PARK ; Jae-Kyung WON ; Joo Ho LEE ; Soon-Tae LEE ; Inpyeong HWANG ; Roh-Eul YOO ; Koung Mi KANG ; Tae Jin YUN
Investigative Magnetic Resonance Imaging 2022;26(1):10-19
Purpose:
To evaluate whether the added value of contrast leakage information from dynamic susceptibility contrast magnetic resonance imaging (DSC MRI) is a better prognostic imaging biomarker than the cerebral blood volume (CBV) value in distinguishing true progression from pseudoprogression in glioblastoma patients.
Materials and Methods:
Forty-nine glioblastoma patients who had undergone MRI after concurrent chemoradiotherapy with temozolomide were enrolled in this retrospective study. Twenty features were extracted from the normalized relative CBV (nCBV) and extraction fraction (EF) map of the contrast-enhancing region in each patient. After univariable analysis, we used multivariable stepwise logistic regression analysis to identify significant predictors for differentiating between pseudoprogression and true progression. Receiver operating characteristic (ROC) analysis was employed to determine the best cutoff values for the nCBV and EF features. Finally, leave-one-out cross-validation was used to validate the best predictor in differentiating between true progression and pseudoprogression.
Results:
Multivariable stepwise logistic regression analysis showed that MGMT (O 6 -methylguanine-DNA methyltransferase) and EF max were independent differentiating variables (P = 0.004 and P = 0.02, respectively). ROC analysis yielded the best cutoff value of 95.75 for the EF max value for differentiating the two groups (sensitivity, 61%; specificity, 84.6%; AUC, 0.681 ± 0.08; 95% CI, 0.524-0.837; P = 0.03). In the leave-one-out cross-validation of the EF max value, the cross-validated values for predicting true progression and pseudoprogression accuracies were 69.4% and 71.4%,respectively.
Conclusion
We demonstrated that contrast leakage information parameter from DSC MRI showed significance in differentiating true progression from pseudoprogression in glioblastoma patients.
8.Changes in Tear Lipid Layer Thickness after Cataract Surgery in Diabetic and Non-diabetic Patients
Ji Seon AN ; Ja Yoon MOON ; Jin Seok CHOI ; Eung Kweon KIM ; Kyu Hong PAK ; Sung Kun CHUNG
Journal of the Korean Ophthalmological Society 2022;63(11):903-909
Purpose:
To compare tear film lipid layer thickness (LLT) and dry eye parameters between non-diabetic and diabetic patients after cataract surgery.
Methods:
We retrospectively studied patients who underwent cataract surgery. We enrolled 65 eyes of 36 patients with cataracts but not diabetes (group I) and 35 eyes of 23 patients with both cataracts and diabetes (group II). The LLT, partial blink rate, tear break-up time (TBUT), and strip meniscometry tests were performed preoperatively and 1 and 2 months after surgery. The ocular surface disease index (OSDI) questionnaire was completed before and 2 months after surgery by both groups.
Results:
In group I, the strip meniscometry test value was 3.2 ± 1.4 mm before surgery, 2.3 ± 1.0 mm at 1 month, and 2.5 ± 1.0 mm at 2 months, thus significantly less than before surgery (p < 0.05). The TBUT result was 3.4 ± 1.6 seconds before surgery, 3.0 ± 1.5 seconds at 1 month, and 2.9 ± 1.5 seconds at 2 months, thus significantly less than before surgery (p < 0.05). The OSDI score was 30.8 ± 21.3 before surgery and 20.0 ± 15.8 at 2 months, thus significantly lower than before surgery (p < 0.05). In group II, the TBUT decreased significantly from the preoperative value of 3.6 ± 1.3 seconds to 3.0 ± 1.0 seconds at 1 month and 3.1 ± 1.0 seconds at 2 months (p < 0.05). The OSDI score was 30.0 ± 20.6 before surgery and 19.0 ± 16.0 at 2 months, thus significantly lower than before surgery (p < 0.05). Neither group evidenced any significant difference in the LLT before and after surgery.
Conclusions
There was a difference in dry eye parameters in each group before and after cataract surgery, but there was no difference between the two groups.
9.Patient-specific surgical options for breast cancer-related lymphedema: technical tips
Jin Geun KWON ; Dae Won HONG ; Hyunsuk Peter SUH ; Changsik John PAK ; Joon Pio HONG
Archives of Plastic Surgery 2021;48(3):246-253
In order to provide a physiological solution for patients with breast cancer-related lymphedema (BCRL), the surgeon must understand where and how the pathology of lymphedema occurred. Based on each patient’s pathology, the treatment plan should be carefully decided and individualized. At the authors’ institution, the treatment plan is made individually based on each patient’s symptoms and relative factors. Most early-stage patients first undergo decongestive therapy and then, depending on the efficacy of the treatment, a surgical approach is suggested. If the patient is indicated for surgery, all the points of lymphatic flow obstruction are carefully examined. Thus a BCRL patient can be considered for lymphaticovenous anastomosis (LVA), a lymph node flap, scar resection, or a combination thereof. LVA targets ectatic superficial collecting lymphatics, which are located within the deep fat layer, and preoperative mapping using ultrasonography is critical. If there is contracture on the axilla, axillary scar removal is indicated to relieve the vein pressure and allow better drainage. Furthermore, removing the scars and reconstructing the fat layer will allow a better chance for the lymphatics to regenerate. After complete removal of scar tissue, a regional fat flap or a superficial circumflex iliac artery perforator flap with lymph node transfer is performed. By deciding the surgical planning for BCRL based on each patient’s pathophysiology, optimal outcomes can be achieved. Depending on each patient’s pathophysiology, LVA, scar removal, vascularized lymph node transfer with a sufficient adipocutaneous flap, and simultaneous breast reconstruction should be planned.
10.Clinical Significance of Tear Film Osmolarity in Patients with Mild Dry Eye Syndrome
Mi Jeong KIM ; Se Young PARK ; Joo Young KWAG ; Jin Seok CHOI ; Kyu Hong PAK ; Sung Kun CHUNG
Journal of the Korean Ophthalmological Society 2021;62(3):295-299
Purpose:
To seek Pearson correlations of tear film osmolarity measured by the I-PEN® (I-MED Pharma Inc., Dollard-des-Ormeaux, Canada) with Schirmer test result, tear break-up time, and Ocular Surface Disease Index score in patients with mild dry eye syndrome.
Methods:
Patients with mild dry eye syndrome were divided into two groups according to Ocular Staining Score: group 1 (50 patients; 67 eyes) and group 2 (59 patients; 91 eyes), 90 patients and 158 eyes in total. The above mentioned correlations were derived.
Results:
No significant correlations were observed between tear film osmolarity and Schirmer test result (r < -0.01, p = 0.97), tear break-up time (r = 0.05, p = 0.54), or Ocular Surface Disease Index score (r = 0.03, p = 0.76). When the two groups were compared, the Ocular Surface Disease Index score significantly differed between groups (p < 0.01), whereas the Schirmer test result (p = 0.31), tear break-up time (p = 0.11), and tear film osmolarity (p = 0.12) did not.
Conclusions
No significant correlations were found between tear film osmolarity and other dry eye indicators in patients with mild dry eye syndrome. The diagnostic utility of tear film osmolarity in patients with moderate dry eye syndrome is should be evaluated.

Result Analysis
Print
Save
E-mail