1.A Commissioning Report on the Magnetic ResonanceCompatible Geneva Brachytherapy Applicator
Yoonsuk HUH ; Hyojun PARK ; Jin JEGAL ; Inbum LEE ; Jaeman SON ; Seonghee KANG ; Chang Heon CHOI ; Jung-in KIM ; Hyeongmin JIN
Progress in Medical Physics 2024;35(4):116-124
		                        		
		                        			 Purpose:
		                        			Brachytherapy is essential for treating gynecological cancers as it offers precise radiation delivery to tumors while minimizing radiation exposure to surrounding healthy tissues. The Geneva applicator, introduced in 2020 as a replacement for older models like the Utrecht applicator, enhances MRI-based brachytherapy with improved imaging capabilities and more accurate applicator placement. In 2021, updates to non-reimbursement policies in Korea for MRI-based 3D brachytherapy planning further promoted the adoption of advanced techniques such as the Geneva applicator. This study aims to commission the Geneva applicator, focusing on wall thickness, dummy marker positions, and source dwell positions to ensure accurate dose delivery and safety. 
		                        		
		                        			Methods:
		                        			The commissioning process involved measuring wall thickness in both the longitudinal and transverse directions for the tandem and lunar-shaped ovoid tubes and comparing thesemeasurements with the manufacturer’s specifications. Dummy marker positions were verifiedusing CT imaging, with a focus on alignment tolerances of ±1 mm. Source dwell positions were planned using the Oncentra treatment planning system, with measurements taken using EBT4 film and analyzed with RIT software. 
		                        		
		                        			Results:
		                        			Wall thickness measurements and dummy marker positions were within the specified tolerance ranges, confirming their accuracy. The source dwell positions, measured and analyzedthrough multiple tests, were all within the ±1 mm tolerance, ensuring the applicator’s reliability. 
		                        		
		                        			Conclusions
		                        			The Geneva applicator met all standards for safe and effective use in brachytherapy.The use of a 3D-printed holder was crucial for precise alignment and measurement. With updated reimbursement policies in Korea for MRI-based brachytherapy, the Geneva applicator is expected to significantly impact the future of advanced brachytherapy treatments and research. 
		                        		
		                        		
		                        		
		                        	
2.Evaluations of a Commercial CLEANBOLUS-WHITE for Clinical Application
Geum Bong YU ; Jung-in KIM ; Jaeman SON
Progress in Medical Physics 2024;35(1):10-15
		                        		
		                        			 Purpose:
		                        			This study aimed to comprehensively investigate the diverse characteristics of a novel commercial bolus, CLEANBOLUS-WHITE (CBW), to ascertain its suitability for clinical application. 
		                        		
		                        			Methods:
		                        			The evaluation of CBW encompassed both physical and biological assessments. Physical parameters such as mass density and shore hardness were measured alongside analyses of element composition. Biological evaluations included assessments for skin irritation and cytotoxicity. Dosimetric properties were examined by calculating surface dose and beam quality using a treatment planning system (TPS). Additionally, doses were measured at maximum and reference depths, and the results were compared with those obtained using a solid water phantom. The effect of air gap on dose measurement was also investigated by comparing measured doses on the RANDO phantom, under the bolus, with doses calculated from the TPS. 
		                        		
		                        			Results:
		                        			Biological evaluation confirmed that CBW is non-cytotoxic, nonirritant, and nonsensitizing. The bolus exhibited a mass density of 1.02 g/cm 3 and 14 shore 00. Dosimetric evaluations revealed that using the 0.5 cm CBW resulted in less than a 1% difference compared to using the solid water phantom. Furthermore, beam quality calculations in the TPS indicated increased surface dose with the bolus. The air gap effect on dose measurement was deemed negligible, with a difference of approximately 1% between calculated and measured doses, aligning with measurement uncertainty. 
		                        		
		                        			Conclusions
		                        			CBW demonstrates outstanding properties for clinical utilization. The dosimetric evaluation underscores a strong agreement between calculated and measured doses, validating its reliability in both planning and clinical settings. 
		                        		
		                        		
		                        		
		                        	
3.A Commissioning Report on the Magnetic ResonanceCompatible Geneva Brachytherapy Applicator
Yoonsuk HUH ; Hyojun PARK ; Jin JEGAL ; Inbum LEE ; Jaeman SON ; Seonghee KANG ; Chang Heon CHOI ; Jung-in KIM ; Hyeongmin JIN
Progress in Medical Physics 2024;35(4):116-124
		                        		
		                        			 Purpose:
		                        			Brachytherapy is essential for treating gynecological cancers as it offers precise radiation delivery to tumors while minimizing radiation exposure to surrounding healthy tissues. The Geneva applicator, introduced in 2020 as a replacement for older models like the Utrecht applicator, enhances MRI-based brachytherapy with improved imaging capabilities and more accurate applicator placement. In 2021, updates to non-reimbursement policies in Korea for MRI-based 3D brachytherapy planning further promoted the adoption of advanced techniques such as the Geneva applicator. This study aims to commission the Geneva applicator, focusing on wall thickness, dummy marker positions, and source dwell positions to ensure accurate dose delivery and safety. 
		                        		
		                        			Methods:
		                        			The commissioning process involved measuring wall thickness in both the longitudinal and transverse directions for the tandem and lunar-shaped ovoid tubes and comparing thesemeasurements with the manufacturer’s specifications. Dummy marker positions were verifiedusing CT imaging, with a focus on alignment tolerances of ±1 mm. Source dwell positions were planned using the Oncentra treatment planning system, with measurements taken using EBT4 film and analyzed with RIT software. 
		                        		
		                        			Results:
		                        			Wall thickness measurements and dummy marker positions were within the specified tolerance ranges, confirming their accuracy. The source dwell positions, measured and analyzedthrough multiple tests, were all within the ±1 mm tolerance, ensuring the applicator’s reliability. 
		                        		
		                        			Conclusions
		                        			The Geneva applicator met all standards for safe and effective use in brachytherapy.The use of a 3D-printed holder was crucial for precise alignment and measurement. With updated reimbursement policies in Korea for MRI-based brachytherapy, the Geneva applicator is expected to significantly impact the future of advanced brachytherapy treatments and research. 
		                        		
		                        		
		                        		
		                        	
4.Evaluations of a Commercial CLEANBOLUS-WHITE for Clinical Application
Geum Bong YU ; Jung-in KIM ; Jaeman SON
Progress in Medical Physics 2024;35(1):10-15
		                        		
		                        			 Purpose:
		                        			This study aimed to comprehensively investigate the diverse characteristics of a novel commercial bolus, CLEANBOLUS-WHITE (CBW), to ascertain its suitability for clinical application. 
		                        		
		                        			Methods:
		                        			The evaluation of CBW encompassed both physical and biological assessments. Physical parameters such as mass density and shore hardness were measured alongside analyses of element composition. Biological evaluations included assessments for skin irritation and cytotoxicity. Dosimetric properties were examined by calculating surface dose and beam quality using a treatment planning system (TPS). Additionally, doses were measured at maximum and reference depths, and the results were compared with those obtained using a solid water phantom. The effect of air gap on dose measurement was also investigated by comparing measured doses on the RANDO phantom, under the bolus, with doses calculated from the TPS. 
		                        		
		                        			Results:
		                        			Biological evaluation confirmed that CBW is non-cytotoxic, nonirritant, and nonsensitizing. The bolus exhibited a mass density of 1.02 g/cm 3 and 14 shore 00. Dosimetric evaluations revealed that using the 0.5 cm CBW resulted in less than a 1% difference compared to using the solid water phantom. Furthermore, beam quality calculations in the TPS indicated increased surface dose with the bolus. The air gap effect on dose measurement was deemed negligible, with a difference of approximately 1% between calculated and measured doses, aligning with measurement uncertainty. 
		                        		
		                        			Conclusions
		                        			CBW demonstrates outstanding properties for clinical utilization. The dosimetric evaluation underscores a strong agreement between calculated and measured doses, validating its reliability in both planning and clinical settings. 
		                        		
		                        		
		                        		
		                        	
5.Optimizing target and diaphragmatic configuration, and dosimetric benefits using continuous positive airway pressure in stereotactic ablative radiotherapy for lung tumors
Jung Bin PARK ; Joo Ho LEE ; Ji Hyun CHANG ; Jaeman SON ; Seho KWON ; Su Yun CHOI ; Hyun-Woo SHIN ; Tosol YU ; Hak Jae KIM
Radiation Oncology Journal 2024;42(3):200-209
		                        		
		                        			 Purpose:
		                        			This study aimed to evaluate the impact of facilitating target delineation of continuous positive airway pressure (CPAP) in patients undergoing stereotactic ablative radiation therapy (SABR) for lung tumors by lung expansion and respiratory motion management. 
		                        		
		                        			Materials and Methods:
		                        			We performed a prospective single-institutional trial of patients who were diagnosed with either primary lung cancer or lung metastases and received SABR with a dose of 40 to 60 Gy in 4 fractions. Four-dimensional computed tomography simulations were conducted for each patient: once without CPAP and again with CPAP.  
		                        		
		                        			Results:
		                        			Thirty-two patients with 39 tumors were analyzed, after the withdrawal of five patients due to discomfort. For 26 tumors separated from the diaphragm, CPAP significantly increased the superoinferior distance between the tumor and the diaphragm (5.96 cm vs. 8.06 cm; p < 0.001). For 13 tumors located adjacent to the diaphragm, CPAP decreased the overlap of planning target volume (PTV) with the diaphragm significantly (6.32 cm3 vs. 4.09 cm3; p = 0.002). PTV showed a significant reduction with CPAP (25.06 cm3 vs. 22.52 cm3, p = 0.017). In dosimetric analyses, CPAP expanded lung volume by 58.4% with a significant reduction in mean dose and V5 to V40. No more than grade 2 adverse events were reported. 
		                        		
		                        			Conclusion
		                        			This trial demonstrated significant improvement of CPAP in target delineation uncertainties for lung SABR, with dosimetric benefits, a favorable safety profile and tolerability. Further investigation is warranted to explore the role of CPAP as a novel strategy for respiratory motion management. 
		                        		
		                        		
		                        		
		                        	
6.Optimizing target and diaphragmatic configuration, and dosimetric benefits using continuous positive airway pressure in stereotactic ablative radiotherapy for lung tumors
Jung Bin PARK ; Joo Ho LEE ; Ji Hyun CHANG ; Jaeman SON ; Seho KWON ; Su Yun CHOI ; Hyun-Woo SHIN ; Tosol YU ; Hak Jae KIM
Radiation Oncology Journal 2024;42(3):200-209
		                        		
		                        			 Purpose:
		                        			This study aimed to evaluate the impact of facilitating target delineation of continuous positive airway pressure (CPAP) in patients undergoing stereotactic ablative radiation therapy (SABR) for lung tumors by lung expansion and respiratory motion management. 
		                        		
		                        			Materials and Methods:
		                        			We performed a prospective single-institutional trial of patients who were diagnosed with either primary lung cancer or lung metastases and received SABR with a dose of 40 to 60 Gy in 4 fractions. Four-dimensional computed tomography simulations were conducted for each patient: once without CPAP and again with CPAP.  
		                        		
		                        			Results:
		                        			Thirty-two patients with 39 tumors were analyzed, after the withdrawal of five patients due to discomfort. For 26 tumors separated from the diaphragm, CPAP significantly increased the superoinferior distance between the tumor and the diaphragm (5.96 cm vs. 8.06 cm; p < 0.001). For 13 tumors located adjacent to the diaphragm, CPAP decreased the overlap of planning target volume (PTV) with the diaphragm significantly (6.32 cm3 vs. 4.09 cm3; p = 0.002). PTV showed a significant reduction with CPAP (25.06 cm3 vs. 22.52 cm3, p = 0.017). In dosimetric analyses, CPAP expanded lung volume by 58.4% with a significant reduction in mean dose and V5 to V40. No more than grade 2 adverse events were reported. 
		                        		
		                        			Conclusion
		                        			This trial demonstrated significant improvement of CPAP in target delineation uncertainties for lung SABR, with dosimetric benefits, a favorable safety profile and tolerability. Further investigation is warranted to explore the role of CPAP as a novel strategy for respiratory motion management. 
		                        		
		                        		
		                        		
		                        	
7.A Commissioning Report on the Magnetic ResonanceCompatible Geneva Brachytherapy Applicator
Yoonsuk HUH ; Hyojun PARK ; Jin JEGAL ; Inbum LEE ; Jaeman SON ; Seonghee KANG ; Chang Heon CHOI ; Jung-in KIM ; Hyeongmin JIN
Progress in Medical Physics 2024;35(4):116-124
		                        		
		                        			 Purpose:
		                        			Brachytherapy is essential for treating gynecological cancers as it offers precise radiation delivery to tumors while minimizing radiation exposure to surrounding healthy tissues. The Geneva applicator, introduced in 2020 as a replacement for older models like the Utrecht applicator, enhances MRI-based brachytherapy with improved imaging capabilities and more accurate applicator placement. In 2021, updates to non-reimbursement policies in Korea for MRI-based 3D brachytherapy planning further promoted the adoption of advanced techniques such as the Geneva applicator. This study aims to commission the Geneva applicator, focusing on wall thickness, dummy marker positions, and source dwell positions to ensure accurate dose delivery and safety. 
		                        		
		                        			Methods:
		                        			The commissioning process involved measuring wall thickness in both the longitudinal and transverse directions for the tandem and lunar-shaped ovoid tubes and comparing thesemeasurements with the manufacturer’s specifications. Dummy marker positions were verifiedusing CT imaging, with a focus on alignment tolerances of ±1 mm. Source dwell positions were planned using the Oncentra treatment planning system, with measurements taken using EBT4 film and analyzed with RIT software. 
		                        		
		                        			Results:
		                        			Wall thickness measurements and dummy marker positions were within the specified tolerance ranges, confirming their accuracy. The source dwell positions, measured and analyzedthrough multiple tests, were all within the ±1 mm tolerance, ensuring the applicator’s reliability. 
		                        		
		                        			Conclusions
		                        			The Geneva applicator met all standards for safe and effective use in brachytherapy.The use of a 3D-printed holder was crucial for precise alignment and measurement. With updated reimbursement policies in Korea for MRI-based brachytherapy, the Geneva applicator is expected to significantly impact the future of advanced brachytherapy treatments and research. 
		                        		
		                        		
		                        		
		                        	
8.Evaluations of a Commercial CLEANBOLUS-WHITE for Clinical Application
Geum Bong YU ; Jung-in KIM ; Jaeman SON
Progress in Medical Physics 2024;35(1):10-15
		                        		
		                        			 Purpose:
		                        			This study aimed to comprehensively investigate the diverse characteristics of a novel commercial bolus, CLEANBOLUS-WHITE (CBW), to ascertain its suitability for clinical application. 
		                        		
		                        			Methods:
		                        			The evaluation of CBW encompassed both physical and biological assessments. Physical parameters such as mass density and shore hardness were measured alongside analyses of element composition. Biological evaluations included assessments for skin irritation and cytotoxicity. Dosimetric properties were examined by calculating surface dose and beam quality using a treatment planning system (TPS). Additionally, doses were measured at maximum and reference depths, and the results were compared with those obtained using a solid water phantom. The effect of air gap on dose measurement was also investigated by comparing measured doses on the RANDO phantom, under the bolus, with doses calculated from the TPS. 
		                        		
		                        			Results:
		                        			Biological evaluation confirmed that CBW is non-cytotoxic, nonirritant, and nonsensitizing. The bolus exhibited a mass density of 1.02 g/cm 3 and 14 shore 00. Dosimetric evaluations revealed that using the 0.5 cm CBW resulted in less than a 1% difference compared to using the solid water phantom. Furthermore, beam quality calculations in the TPS indicated increased surface dose with the bolus. The air gap effect on dose measurement was deemed negligible, with a difference of approximately 1% between calculated and measured doses, aligning with measurement uncertainty. 
		                        		
		                        			Conclusions
		                        			CBW demonstrates outstanding properties for clinical utilization. The dosimetric evaluation underscores a strong agreement between calculated and measured doses, validating its reliability in both planning and clinical settings. 
		                        		
		                        		
		                        		
		                        	
9.Optimizing target and diaphragmatic configuration, and dosimetric benefits using continuous positive airway pressure in stereotactic ablative radiotherapy for lung tumors
Jung Bin PARK ; Joo Ho LEE ; Ji Hyun CHANG ; Jaeman SON ; Seho KWON ; Su Yun CHOI ; Hyun-Woo SHIN ; Tosol YU ; Hak Jae KIM
Radiation Oncology Journal 2024;42(3):200-209
		                        		
		                        			 Purpose:
		                        			This study aimed to evaluate the impact of facilitating target delineation of continuous positive airway pressure (CPAP) in patients undergoing stereotactic ablative radiation therapy (SABR) for lung tumors by lung expansion and respiratory motion management. 
		                        		
		                        			Materials and Methods:
		                        			We performed a prospective single-institutional trial of patients who were diagnosed with either primary lung cancer or lung metastases and received SABR with a dose of 40 to 60 Gy in 4 fractions. Four-dimensional computed tomography simulations were conducted for each patient: once without CPAP and again with CPAP.  
		                        		
		                        			Results:
		                        			Thirty-two patients with 39 tumors were analyzed, after the withdrawal of five patients due to discomfort. For 26 tumors separated from the diaphragm, CPAP significantly increased the superoinferior distance between the tumor and the diaphragm (5.96 cm vs. 8.06 cm; p < 0.001). For 13 tumors located adjacent to the diaphragm, CPAP decreased the overlap of planning target volume (PTV) with the diaphragm significantly (6.32 cm3 vs. 4.09 cm3; p = 0.002). PTV showed a significant reduction with CPAP (25.06 cm3 vs. 22.52 cm3, p = 0.017). In dosimetric analyses, CPAP expanded lung volume by 58.4% with a significant reduction in mean dose and V5 to V40. No more than grade 2 adverse events were reported. 
		                        		
		                        			Conclusion
		                        			This trial demonstrated significant improvement of CPAP in target delineation uncertainties for lung SABR, with dosimetric benefits, a favorable safety profile and tolerability. Further investigation is warranted to explore the role of CPAP as a novel strategy for respiratory motion management. 
		                        		
		                        		
		                        		
		                        	
10.A Commissioning Report on the Magnetic ResonanceCompatible Geneva Brachytherapy Applicator
Yoonsuk HUH ; Hyojun PARK ; Jin JEGAL ; Inbum LEE ; Jaeman SON ; Seonghee KANG ; Chang Heon CHOI ; Jung-in KIM ; Hyeongmin JIN
Progress in Medical Physics 2024;35(4):116-124
		                        		
		                        			 Purpose:
		                        			Brachytherapy is essential for treating gynecological cancers as it offers precise radiation delivery to tumors while minimizing radiation exposure to surrounding healthy tissues. The Geneva applicator, introduced in 2020 as a replacement for older models like the Utrecht applicator, enhances MRI-based brachytherapy with improved imaging capabilities and more accurate applicator placement. In 2021, updates to non-reimbursement policies in Korea for MRI-based 3D brachytherapy planning further promoted the adoption of advanced techniques such as the Geneva applicator. This study aims to commission the Geneva applicator, focusing on wall thickness, dummy marker positions, and source dwell positions to ensure accurate dose delivery and safety. 
		                        		
		                        			Methods:
		                        			The commissioning process involved measuring wall thickness in both the longitudinal and transverse directions for the tandem and lunar-shaped ovoid tubes and comparing thesemeasurements with the manufacturer’s specifications. Dummy marker positions were verifiedusing CT imaging, with a focus on alignment tolerances of ±1 mm. Source dwell positions were planned using the Oncentra treatment planning system, with measurements taken using EBT4 film and analyzed with RIT software. 
		                        		
		                        			Results:
		                        			Wall thickness measurements and dummy marker positions were within the specified tolerance ranges, confirming their accuracy. The source dwell positions, measured and analyzedthrough multiple tests, were all within the ±1 mm tolerance, ensuring the applicator’s reliability. 
		                        		
		                        			Conclusions
		                        			The Geneva applicator met all standards for safe and effective use in brachytherapy.The use of a 3D-printed holder was crucial for precise alignment and measurement. With updated reimbursement policies in Korea for MRI-based brachytherapy, the Geneva applicator is expected to significantly impact the future of advanced brachytherapy treatments and research. 
		                        		
		                        		
		                        		
		                        	
            