3.In Vitro Experimental Study on Hemodynamics of Transcatheter Aortic Valve Replacement.
Chinese Journal of Medical Instrumentation 2023;47(4):383-390
The patient-specific aortic silicone model was established based on CTA data. The digital particle image velocimetry (DPIV) test method in the modified ViVitro pulsatile flow system was used to investigate the aortic hemodynamic performance and flow field characteristics before and after transcatheter aortic valve replacement (TAVR). The results showed that the hemodynamic parameters were consistent with the clinical data, which verified the accuracy of the model. From the comparative study of preoperative and postoperative effective orifice area (0.33 cm2 and 1.78 cm2), mean pressure difference (58 mmHg and 9 mmHg), percentage of regurgitation (52% and 8%), peak flow velocity (4.60 m/s and 1.81 m/s) and flow field distribution (eccentric jet and uniform jet), the immediate efficacy after TAVR is good. From the perspective of viscous shear stress and Reynolds shear stress, the risk of hemolysis and thrombotic problems was low in preoperative and postoperative patient-specific models. This study provides a set of reliable DPIV testing methods for aortic flow field, and provides biomechanical basis for the immediate and long-term effectiveness of TAVR from the perspective of hemodynamics and flow field characteristics. It has important application value in clinical diagnosis, surgical treatment and long-term evaluation.
Humans
;
Transcatheter Aortic Valve Replacement/methods*
;
Aortic Valve/surgery*
;
Heart Valve Prosthesis
;
Hemodynamics
;
Aortic Valve Stenosis/diagnosis*
;
Treatment Outcome
5.Long-term outcome of transcatheter repair of paravalvular leak post surgical mitral valve replacement.
Chao Fan XING ; Xin PAN ; Cheng WANG ; Lan MA ; Xiao Lei WANG ; Yan Jie LI ; Ben HE
Chinese Journal of Cardiology 2023;51(7):742-749
Objective: To explore the long-term clinical efficacy of transcatheter repair of mitral paravalvular leak (PVL) post surgical mitral valve replacement. Methods: This study is a retrospective study. Patients who completed transcatheter repair of paravalvular leak after surgical mitral valve replacement at Shanghai Chest Hospital, Shanghai Jiaotong University School of Medicine from March 2010 to December 2018 were included. Technical success was defined as the occluder being stably implanted in the paravalvular leak site without affecting the function of the mitral valve and surrounding tissues; and there were no intervention-related complications, such as new hemolysis or aggravated hemolysis, and echocardiography confirmed mitral paravalvular regurgitation reduced by more than 1 grade. Patients were followed up at 30 days, 1, and 3 years after the intervention. The main endpoints were all-cause death and re-surgery due to interventional failure or serious complications. The occurrence of occluder-mediated hemolysis and chronic renal insufficiency was recorded, and patients were monitored with echocardiography during follow up. Results: A total of 75 patients were included, aged (54.3±22.9) years old, and 38 patients were males. All patients had decreased cardiac function and/or hemolysis before intervention. Procedural success was achieved in 54 patients (72.0%). Incidence of device-mediated hemolysis was 18.7% (14/75). During the follow-up period, all-cause death occurred in 7 patients (9.3%), and 3 were cardiac deaths.The 3-year event-free survival rate was 81.3% (61/75). The need for cardiac surgery was 9.3% (7/75): 3 cases due to severe device-mediated hemolysis, 2 cases due to prosthetic valve failure and 2 cases due to moderate to severe residual regurgitation. The echocardiography follow-up results showed that the position of the occluder was stable, there was no impact on the artificial valve function and surrounding structures, and the residual regurgitation was stable without progressive increase in event-free patients. Compared with pre-intervention, the left ventricular end systolic diameter ((33.9±7.4)mm vs. (38.3±8.9) mm, P=0.036), end diastolic diameter ((53.7±8.3) mm vs. (58.4±9.1) mm, P=0.045) and left atrial diameter (59.3 (44.5, 90.7) mm vs. 64.3 (44.8, 96.6) mm, P=0.049) were significantly reduced, pulmonary artery systolic pressure was also significantly decreased ((36.5±15.8) mmHg vs. (46.3±14.9) mmHg, P=0.022, 1 mmHg=0.133 kPa). There was no significant difference between 3 years and 1 year after transcatheter repair of mitral paravalvular leak post surgical mitral valve replacement (all P>0.05). Conclusion: Transcatheter repair of mitral paravalvular leak post surgical mitral valve replacement is an effective treatment option in selective patients.
Male
;
Humans
;
Adult
;
Middle Aged
;
Aged
;
Female
;
Mitral Valve/surgery*
;
Heart Valve Prosthesis Implantation
;
Heart Valve Prosthesis
;
Retrospective Studies
;
Hemolysis
;
China
;
Mitral Valve Insufficiency/surgery*
;
Treatment Outcome
;
Cardiac Catheterization
;
Prosthesis Failure
9.Gender differences in mortality following tanscatheter aortic valve replacement (TAVR): a single-centre retrospective analysis from China.
Qi LIU ; Yali WANG ; Yijian LI ; Tianyuan XIONG ; Fei CHEN ; Yuanweixiang OU ; Xi WANG ; Yijun YAO ; Kaiyu JIA ; Yujia LIANG ; Xin WEI ; Xi LI ; Yong PENG ; Jiafu WEI ; Sen HE ; Qiao LI ; Wei MENG ; Guo CHEN ; Wenxia ZHOU ; Mingxia ZHENG ; Xuan ZHOU ; Zhengang ZHAO ; Chen MAO ; Feng YUAN
Chinese Medical Journal 2023;136(20):2511-2513
10.Preliminary experience of transcatheter pulmonary valve replacement using domestic balloon-expandable valve.
Zhen Gang ZHAO ; Rui Tao LI ; Xin WEI ; Yong PENG ; Jia Fu WEI ; Sen HE ; Qiao LI ; Xiao LI ; Yi Jian LI ; Xiang LI ; Xuan ZHOU ; Ming Xia ZHENG ; Guo CHEN ; Qi AN ; Mao CHEN ; Yuan FENG
Chinese Journal of Cardiology 2023;51(8):825-831
Objectives: To evaluate the feasibility and preliminary clinical results of transcatheter pulmonary valve replacement (TPVR) with the domestically-produced balloon-expandable Prizvalve system. Methods: This is a prospective single-center observational study. Patients with postoperative right ventricular outflow tract (RVOT) dysfunction, who were admitted to West China Hospital of Sichuan University from September 2021 to March 2023 and deemed anatomically suitable for TPVR with balloon-expandable valve, were included. Clinical, imaging, procedural and follow-up data were analyzed. The immediate procedural results were evaluated by clinical implant success rate, which is defined as successful valve implantation with echocardiography-assessed pulmonary regurgitation
Male
;
Humans
;
Pulmonary Valve/surgery*
;
Heart Valve Prosthesis/adverse effects*
;
Heart Valve Prosthesis Implantation
;
Constriction, Pathologic/surgery*
;
Prospective Studies
;
Ventricular Outflow Obstruction/surgery*
;
Treatment Outcome
;
Cardiac Catheterization/methods*
;
Transcatheter Aortic Valve Replacement

Result Analysis