1.Comparison of ZOOMit-DWI sequence and conventional DWI sequence in endometrial cancer.
Shixiong TANG ; Chun FU ; Hongliang CHEN ; Enhua XIAO ; Yicheng LONG ; Dujun BIAN
Journal of Central South University(Medical Sciences) 2023;48(1):76-83
OBJECTIVES:
Magnetic resonance diffusion-weighted imaging (DWI) has important clinical value in diagnosis and curative effect evaluation on endometrial carcinoma. How to improve the detection rate of endometrial small lesions by DWI is the research focus of MRI technology. This study aims to analyze the image quality of small field MRI ZOOMit-DWI sequence and conventional single-shot echo-planar imaging (SS-EPI) DWI sequence in the scanning of endometrial carcinoma, and to explore the clinical value of ZOOMit-DWI sequence.
METHODS:
A total of 37 patients with endometrial carcinoma diagnosed by operation and pathology in the Second Xiangya Hospital of Central South University from July 2019 to May 2021 were collected. All patients were scanned with MRI ZOOMit-DWI sequence and SS-EPI DWI sequence before operation. Two radiologists subjectively evaluated the anatomical details, artifacts, geometric deformation and focus definition of the 2 groups of DWI images. At the same time, the signal intensity were measured and the signal-to-noise ratio (SNR), contrast to noise ratio (CNR), and apparent diffusion coefficient (ADC) of the 2 DWI sequences were calculated for objective evaluation. The differences of subjective score, objective score and ADC value of the 2 DWI sequences were analyzed.
RESULTS:
The SNR of the ZOOMit-DWI group was significantly higher than that of the SS-EPI DWI group (301.96±141.85 vs 94.66±41.26), and the CNR of the ZOOMit-DWI group was significantly higher than that of the SS-EPI DWI group (185.05±105.45 vs 57.91±31.54, P<0.05). There was no significant difference in noise standard deviation between the ZOOMit-DWI group and the SS-EPI DWI group (P>0.05). The subjective score of anatomical detail and focus definition in the ZOOMit-DWI group was significantly higher than that of the SS-EPI DWI group (both P<0.05). The subjective score of artifacts and geometric deformation of ZOOMit-DWI group was significantly lower than that of the SS-EPI DWI group (both P<0.05). ADC had no significant difference between the ZOOMit-DWI group and the SS-EPI DWI group (P>0.05).
CONCLUSIONS
The image quality of ZOOMit-DWI is significantly higher than that of conventional SS-EPI DWI. In the MRI DWI examination of endometrial carcinoma, ZOOMit-DWI can effectively reduce the geometric deformation and artifacts of the image, which is more conducive to clinical diagnosis and treatment.
Female
;
Humans
;
Signal-To-Noise Ratio
;
Endometrial Neoplasms/diagnostic imaging*
;
Diffusion Magnetic Resonance Imaging/methods*
;
Endometrium
;
Echo-Planar Imaging/methods*
;
Reproducibility of Results
2.Accelerated Resting-State Functional Magnetic Resonance Imaging Using Multiband Echo-Planar Imaging with Controlled Aliasing.
Hyung Suk SEO ; Kyung Eun JANG ; Dingxin WANG ; In Seong KIM ; Yongmin CHANG
Investigative Magnetic Resonance Imaging 2017;21(4):223-232
PURPOSE: To report the use of multiband accelerated echo-planar imaging (EPI) for resting-state functional MRI (rs-fMRI) to achieve rapid high temporal resolution at 3T compared to conventional EPI. MATERIALS AND METHODS: rs-fMRI data were acquired from 20 healthy right-handed volunteers by using three methods: conventional single-band gradient-echo EPI acquisition (Data 1), multiband gradient-echo EPI acquisition with 240 volumes (Data 2) and 480 volumes (Data 3). Temporal signal-to-noise ratio (tSNR) maps were obtained by dividing the mean of the time course of each voxel by its temporal standard deviation. The resting-state sensorimotor network (SMN) and default mode network (DMN) were estimated using independent component analysis (ICA) and a seed-based method. One-way analysis of variance (ANOVA) was performed between the tSNR map, SMN, and DMN from the three data sets for between-group analysis. P < 0.05 with a family-wise error (FWE) correction for multiple comparisons was considered statistically significant. RESULTS: One-way ANOVA and post-hoc two-sample t-tests showed that the tSNR was higher in Data 1 than Data 2 and 3 in white matter structures such as the striatum and medial and superior longitudinal fasciculus. One-way ANOVA revealed no differences in SMN or DMN across the three data sets. CONCLUSION: Within the adapted metrics estimated under specific imaging conditions employed in this study, multiband accelerated EPI, which substantially reduced scan times, provides the same quality image of functional connectivity as rs-fMRI by using conventional EPI at 3T. Under employed imaging conditions, this technique shows strong potential for clinical acceptance and translation of rs-fMRI protocols with potential advantages in spatial and/or temporal resolution. However, further study is warranted to evaluate whether the current findings can be generalized in diverse settings.
Dataset
;
Echo-Planar Imaging*
;
Magnetic Resonance Imaging*
;
Methods
;
Signal-To-Noise Ratio
;
Volunteers
;
White Matter
3.Effect of Age on Cortical Activation during Swallowing: An fMRI Study.
Hyun Im MOON ; Youjin JUNG ; Sunyoung CHOI ; Wooseok TAE ; Sung Bom PYUN
Journal of the Korean Dysphagia Society 2016;6(1):26-33
OBJECTIVE: Understanding the neural functional organization of swallowing in the elderly is essential when diagnosing and treating older adults with swallowing difficulties. While brain-imaging studies in young adults have implicated multiple cortical regions in swallowing, only a few investigations were performed on older subjects. In this study, we aimed to compare neural activation in regions for swallowing between healthy young and older adults and to better understand neural control of deglutition, complex sensory-motor process which occurs as a result of old age. METHOD: Fifteen young and fifteen older healthy individuals without a swallowing problem were examined with functional magnetic resonance imaging (fMRI) during voluntary saliva swallowing. Functional image data was obtained with a T2 gradient-echo, echo planar imaging (EPI) pulse sequence optimized for blood-oxygen level dependent (BOLD) contrast. Two samples t-test was conducted to perform group comparison (younger adults versus older adults) for the areas in which the activation was larger for the swallowing condition than the non-swallow condition. RESULT: Both groups showed activations in areas involved in the motor control and execution. In both groups, main regions of activation included bilateral prefrontal cortex, primary somatosensory cortex, insula, basal ganglia, and cerebellum. Between-group comparisons revealed statistically stronger activations in the prefrontal cortex and middle temporal gyrus of older adults during swallowing. CONCLUSION: This study provides evidence that swallowing requires larger and more widespread areas of neural control in older adults group, especially in prefrontal cortex and inferior frontal gyrus. These findings suggest that more demanding swallowing tasks are necessary for elderly patients because of their inefficient neural network due to their age.
Adult
;
Aged
;
Basal Ganglia
;
Cerebellum
;
Deglutition*
;
Echo-Planar Imaging
;
Humans
;
Magnetic Resonance Imaging*
;
Methods
;
Prefrontal Cortex
;
Saliva
;
Somatosensory Cortex
;
Temporal Lobe
;
Young Adult
4.Readout-Segmented Echo-Planar Imaging in Diffusion-Weighted MR Imaging in Breast Cancer: Comparison with Single-Shot Echo-Planar Imaging in Image Quality.
Yun Ju KIM ; Sung Hun KIM ; Bong Joo KANG ; Chang Suk PARK ; Hyeon Sook KIM ; Yo Han SON ; David Andrew PORTER ; Byung Joo SONG
Korean Journal of Radiology 2014;15(4):403-410
OBJECTIVE: The purpose of this study was to compare the image quality of standard single-shot echo-planar imaging (ss-EPI) and that of readout-segmented EPI (rs-EPI) in patients with breast cancer. MATERIALS AND METHODS: Seventy-one patients with 74 breast cancers underwent both ss-EPI and rs-EPI. For qualitative comparison of image quality, three readers independently assessed the two sets of diffusion-weighted (DW) images. To evaluate geometric distortion, a comparison was made between lesion lengths derived from contrast enhanced MR (CE-MR) images and those obtained from the corresponding DW images. For assessment of image parameters, signal-to-noise ratio (SNR), lesion contrast, and contrast-to-noise ratio (CNR) were calculated. RESULTS: The rs-EPI was superior to ss-EPI in most criteria regarding the qualitative image quality. Anatomical structure distinction, delineation of the lesion, ghosting artifact, and overall image quality were significantly better in rs-EPI. Regarding the geometric distortion, lesion length on ss-EPI was significantly different from that of CE-MR, whereas there were no significant differences between CE-MR and rs-EPI. The rs-EPI was superior to ss-EPI in SNR and CNR. CONCLUSION: Readout-segmented EPI is superior to ss-EPI in the aspect of image quality in DW MR imaging of the breast.
Adult
;
Aged
;
Aged, 80 and over
;
Artifacts
;
Breast/pathology
;
Breast Neoplasms/pathology
;
Contrast Media/diagnostic use
;
Diffusion Magnetic Resonance Imaging/methods/*standards
;
Echo-Planar Imaging/methods/*standards
;
Female
;
Humans
;
Image Enhancement/methods/*standards
;
Middle Aged
;
Observer Variation
;
Retrospective Studies
;
Sensitivity and Specificity
;
Signal-To-Noise Ratio
5.Renal arterial spin labeling magnetic resonance imaging in normal adults: a study with a 3.0 T scanner.
Fan ZHANG ; Xuelin ZHANG ; Li YANG ; Jie SHEN ; Wei GAO
Journal of Southern Medical University 2013;33(10):1478-1482
OBJECTIVETo analyze the renal relative blood flow value (rBFV) and image quality in normal adults using single-shot fast spin echo, flow sensitive invention recovery (SSFSE-FAIR) magnetic resonance (MR) sequence and echo planar imaging, and flow sensitive invention recovery (EPI-FAIR) MR sequence, and assess its value for clinical application in routine renal examination.
METHODSForty volunteers (25 male and 15 female adults, aged 30 to 62 years) with normal renal function were included in this prospective study. All the subjects underwent 3.0 Tesla MR scanning using 3 MR scan modes, namely breath-holding EPI-FAIR, breath-holding SSFSE-FAIR and free breathing SSFSE-FAIR.
RESULTSSSFSE-FAIR without breath-holding was capable of differentiating the renal cortex and medulla with the corresponding rBFVs of 111.48∓9.23 and 94.98∓3.38, respectively. Breath-holding SSFSE-FAIR and EPI-FAIR failed to distinguish the borders of the renal cortex and medulla. The EPI-FAIR rBFV of mixed cortex and medulla value was 178.50∓17.17 (95%CI: 167.59, 189.41).
CONCLUSIONBreath-holding SSFSE-FAIR and EPI-FAIR can not distinguish the renal cortex and medulla due to a poor spatial resolution but can be used for rough evaluation of renal blood perfusion. Free breathing SSFSE-FAIR with an improved spatial resolution allows evaluation of the status of renal perfusion of the cortex and medulla.
Adult ; Echo-Planar Imaging ; methods ; Female ; Humans ; Kidney ; blood supply ; physiology ; Magnetic Resonance Imaging ; methods ; Male ; Middle Aged ; Prospective Studies ; Renal Circulation
6.Experimental Evaluation of Accelerated T1rho Relaxation Quantification in Human Liver Using Limited Spin-Lock Times.
Feng ZHAO ; Min DENG ; Jing YUAN ; Gao Jun TENG ; Anil T AHUJA ; Yi Xiang J WANG
Korean Journal of Radiology 2012;13(6):736-742
OBJECTIVE: It was reported lately that to obtain consistent liver T1rho measurement, at 3T MRI using six spin-lock times (SLTs), is feasible. In this study, the feasibility of using three or two SLT points to measure liver T1rho relaxation time was explored. MATERIALS AND METHODS: Seventeen healthy volunteers underwent 36 examinations. Three representative axial slices were selected to cut through the upper, middle, and lower liver. A rotary echo spin-lock pulse was implemented in a 2D fast field echo sequence. Spin-lock frequency was 500 Hz and the spin-lock times of 1, 10, 20, 30, 40, and 50 milliseconds (ms) were used for T1rho mapping. T1rho maps were constructed by using all 6 SLT points, three SLT points of 1, 20, and 50 ms, or two SLTs of 1 and 50 ms, respectively. Intra-class correlation coefficient (ICC) and Bland and Altman plot were used to assess the measurement agreement. RESULTS: Two examinations were excluded, due to motion artifact at the SLT of 50 ms. With the remaining 34 examinations, the ICC for 6-SLT vs. 3-SLT T1rho measurements was 0.922, while the ICC for 6-SLT vs. 2-SLT T1rho measurement was 0.756. The Bland and Altman analysis showed a mean difference of 0.19 (95% limits of agreement: -1.34, 1.73) for 6-SLT vs. 3-SLT T1rho measurement, and the mean difference of 0.89 (95% limits of agreement: -1.67, 3.45) for 6-SLT vs. 2-SLT T1rho measurement. The scan re-scan reproducibility ICC (n = 11 subjects) was 0.755, 0.727, and 0.528 for 6-SLT measurement, 3-SLT measurement, and 2-SLT measurement, respectively. CONCLUSION: Adopting 3 SLTs of 1, 20, and 50 ms can be an acceptable alternative for the liver T1rho measurement, while 2 SLTs of 1 and 50 ms do not provide reliable measurement.
Adult
;
Eating
;
Echo-Planar Imaging/*methods
;
Fasting
;
Female
;
Humans
;
Liver/*anatomy & histology
;
Magnetic Resonance Imaging/*methods
;
Male
;
Young Adult
7.Diffusion-weighted magnetic resonance imaging with short T1 inversion recovery-echo planar imaging combined with dual-head coincidence single photon emission computed tomography for diagnosing solitary pulmonary nodule.
Wei WANG ; Jin-Zhi XU ; Tong ZHANG ; Bao-Zhong SHEN
Chinese Medical Journal 2010;123(24):3717-3721
8.MR Spectroscopy and Perfusion MR Imaging Findings of Intracranial Foreign Body Granuloma: a Case Report.
Seung Won JANG ; Sang Joon KIM ; Sun Mi KIM ; Jeong Hyun LEE ; Choong Gon CHOI ; Deok Hee LEE ; Eun Ju KIM ; Jung Kyo LEE
Korean Journal of Radiology 2010;11(3):359-363
We report a case of intracranial foreign body granuloma that showed features of a high grade tumor on magnetic resonance (MR) imaging. However, the relative cerebral blood volume was not increased in the enhancing mass on perfusion MRI and the choline/creatine ratio only slightly increased on MR spectroscopy. The results suggest that the lesion is benign in nature. Perfusion MRI and MR spectroscopy may be helpful to differentiate a foreign body granuloma from a neoplastic condition.
Adult
;
Blood Volume
;
Brain/*pathology/surgery
;
Contrast Media/diagnostic use
;
Diagnosis, Differential
;
Echo-Planar Imaging/*methods
;
Female
;
Follow-Up Studies
;
Gadolinium/diagnostic use
;
Granuloma, Foreign-Body/*pathology/surgery
;
Humans
;
Image Enhancement/methods
;
Magnetic Resonance Angiography/*methods
;
Magnetic Resonance Spectroscopy/*methods
9.Apparent Diffusion Coefficient Value of Diffusion-Weighted Imaging for Hepatocellular Carcinoma: Correlation with the Histologic Differentiation and the Expression of Vascular Endothelial Growth Factor.
Suk Hee HEO ; Yong Yeon JEONG ; Sang Soo SHIN ; Jin Woong KIM ; Hyo Soon LIM ; Jae Hyuk LEE ; Yang Seok KOH ; Chol Kyoon CHO ; Heoung Keun KANG
Korean Journal of Radiology 2010;11(3):295-303
OBJECTIVE: To evaluate whether the histopathological differentiation and the expression of vascular endothelial growth factor (VEGF) of hepatocellular carcinoma (HCC) do show correlation with the apparent diffusion coefficient (ADC) value on diffusion-weighted imaging (DWI). MATERIALS AND METHODS: Twenty-seven HCCs from 27 patients who had undergone preoperative liver MRI (1.5T) and surgical resection were retrospectively reviewed. DWI was obtained with a single-shot, echo-planar imaging sequence in the axial plane (b values: 0 and 1,000 sec/mm2). On DWIs, the ADC value of the HCCs was measured by one radiologist, who was kept 'blinded' to the histological findings. Histopathologically, the differentiation was classified into well (n = 9), moderate (n = 9) and poor (n = 9). The expression of VEGF was semiquantitatively graded as grade 0 (n = 8), grade 1 (n = 9) and grade 2 (n = 10). We analyzed whether the histopathological differentiation and the expression of VEGF of the HCC showed correlation with the ADC value on DWI. RESULTS: The mean ADC value of the poorly-differentiated HCCs (0.9 +/- 0.13x10(-3) mm2/s) was lower than those of the well-differentiated HCCs (1.2 +/- 0.22x10(-3) mm2/s) (p = 0.031) and moderately-differentiated HCCs (1.1 +/- 0.01x10(-3) mm2/s) (p = 0.013). There was a significant correlation between the differentiation and the ADC value of the HCCs (r = -0.51, p = 0.012). The mean ADC of the HCCs with a VEGF expression grade of 0, 1 and 2 was 1.1 +/- 0.17, 1.1 +/- 0.21 and 1.1 +/- 0.18x10(-3) mm2/s, respectively. The VEGF expression did not show correlation with the ADC value of the HCCs (r = 0.07, p = 0.74). CONCLUSION: The histopathological differentiation of HCC shows inverse correlation with the ADC value. Therefore, DWI with ADC measurement may be a valuable tool for noninvasively predicting the differentiation of HCC.
Adult
;
Aged
;
Carcinoma, Hepatocellular/metabolism/*pathology
;
Cell Differentiation
;
Contrast Media/diagnostic use
;
Diffusion Magnetic Resonance Imaging/*methods
;
Echo-Planar Imaging/methods
;
Female
;
Gadolinium DTPA/diagnostic use
;
Humans
;
Image Enhancement/methods
;
Liver/metabolism/pathology
;
Liver Neoplasms/metabolism/*pathology
;
Male
;
Middle Aged
;
Predictive Value of Tests
;
Retrospective Studies
;
Sensitivity and Specificity
;
Severity of Illness Index
;
Vascular Endothelial Growth Factor A/*metabolism
10.Image quality stability of whole-body diffusion weighted imaging.
Yun-bin CHEN ; Chun-miao HU ; Jing ZHONG ; Fei SUN
Chinese Medical Sciences Journal 2009;24(2):122-126
OBJECTIVETo assess the reproducibility of whole-body diffusion weighted imaging (WB-DWI) technique in healthy volunteers under normal breathing with background body signal suppression.
METHODSWB-DWI was performed on 32 healthy volunteers twice within two-week period using short TI inversion-recovery diffusion-weighted echo-planar imaging sequence and built-in body coil. The volunteers were scanned across six stations continuously covering the entire body from the head to the feet under normal breathing. The bone apparent diffusion coefficient (ADC) and exponential ADC (eADC) of regions of interest (ROIs) were measured. We analyzed correlation of the results using paired-t-test to assess the reproducibility of the WB-DWI technique.
RESULTSWe were successful in collecting and analyzing data of 64 WB-DWI images. There was no significant difference in bone ADC and eADC of 824 ROIs between the paired observers and paired scans (P>0.05). Most of the images from all stations were of diagnostic quality.
CONCLUSIONThe measurements of bone ADC and eADC have good reproducibility. WB-DWI technique under normal breathing with background body signal suppression is adequate.
Adult ; Aged ; Diffusion Magnetic Resonance Imaging ; methods ; Echo-Planar Imaging ; methods ; Female ; Humans ; Image Processing, Computer-Assisted ; methods ; Male ; Middle Aged ; Reproducibility of Results ; Respiration ; Whole Body Imaging ; methods