It has become an industry consensus that self-assembled nanoparticles (SAN) are formed by molecular recognition of chemical components in traditional Chinese medicine during the decoction process. The insoluble components in the decoction are mostly in the form of nanoparticles, which can improve the problem of poor water solubility. However, the transfer rate of these insoluble components in the decoction is still very low, which limits the efficacy of the drug. This study aimed to refine the traditional decoction self-assembly phenomenon. The self-assembled nanoparticles were constructed by micro-precipitation method (MP-SAN), and characterized by particle size, zeta potential, stability index and morphology. The formation of MP-SAN and alterations in related physicochemical properties were evaluated using modern spectroscopic and thermal analysis techniques. The quality value transmitting pattern of lignan components within the MP-SAN was assessed
2.Performance test of high-flow nasal cannula oxygen therapy device at different altitudes
Chun-Wei HE ; Ya-Ting WANG ; Yi-Long ZHOU ; Hui WANG ; Zi-Yu FU ; De-Dong MA
Chinese Medical Equipment Journal 2024;45(6):49-58
Objective To investigate the performance of brands and types of high-flow nasal cnnula oxygen therapy(HFNC)devices at different altitudes.Methods Four different models of HFNC devices,including R-80S bi-level non-invasive ventilator integrated with HFNC device,HF-60A HFNC device,HFT-300 HFNC device and H-80A HFNC device,were connected with the gas flow meter,simularted head and QuickLung and then put into a low-pressure chamber.The flow rates of the HFNC devices were set to 10,15,20,25,30,35,40,45,50,55 and 60 L/min,and the simulated altitudes of the low-pressure chamber were set to 6 000,5 000,4 000,3 000,2 000,1 000 and 0 m.The actual output airway flow rates,airway pressure changes and trends of the four HFNC devices were recorded at different setting altitudes and flow rates.SPSS 25.0 software was used for statistical analysis.Results The actual output airway flow rates of the four HFNC devices showed an increasing trend as the altitude rose with the simulated altitude of 6 000 m and the setting flow rate kept constant,which increased slowly and even went to decrease when the altitude and flow rate exceeded some limits.The degree of changes in the flow rate with the increasing altitude varied,and there was no uniform pattern.With the rising of altitude,the actual output airway pressure of the four HFNC devices with the flow rate raning from 10 to 35 L/min also increased gradually,which showed a decreasing trend(turning point)after going up to some certain value when the flow rate exceeded 35 L/min,and the altitude where the turning point appeared was lowered as the flow rate increased.Conclusion The actual output airway flow rates and airway pressure during HFNC rise at a high-altitude environment,and generally considerations have to be taken on required airway pressure,patient comfort and the altitude of the patient's usual place of residence when setting the flow rates of the HFNC device.[Chinese Medical Equipment Journal,2024,45(6):49-58]
3.Intraoperative slide rail CT assistance in percutaneous sacroiliac joint screws for the treatment of pelvic posterior ring injury
Bin SHENG ; Yi-Wei WANG ; Yu-Si WANG ; De-Long LIU ; Zhan-Yu YANG ; Rui GUAN ; Chao LIU
China Journal of Orthopaedics and Traumatology 2024;37(5):438-444
Objective To compare the clinical efficacy of intraoperative slide rail CT combined with C-arm X-ray assis-tance and just C-arm for percutaneous screw in the treatment of pelvic posterior ring injury.Methods A retrospective analysis was performed on the patient data of 76 patients with posterior pelvic ring injury admitted to the Department of Orthopedic Trauma from December 2018 to February 2022.Among them,39 patients in the CT group were treated with C-arm combined with slide rail CT-assisted inline fixation including 23 males and 16 females with an average age of(44.98±7.33)years old;and the other 37 patients in the C-arm group were treated with intraline fixation treatment under only C-arm fluoroscopy in-cluding 24 males and 13 females with an average age of(44.37±10.82)years old.Among them,42 patients with anterior ring fractures were treated with percutaneous inferior iliac spines with internal fixation(INFIX)or suprapubic support screws to fix the anterior pelvic ring.Postoperative follow-up time,operation time,complications of the two groups were compared.Results of Matta reduction criteria,Majed efficacy evaluation,the CT grading and the rate of secondary surgical revision were com-pared.Results The nailing time of(32.63±7.33)min in CT group was shorter than that of(52.95±10.64)min in C-arm group(t=-9.739,P<0.05).The follow-up time between CT group(11.97±1.86)months and C-arm group(12.03±1.71)months were not statistically significant(P>0.05).The postoperative complication rates between two groups were not statistically significant(x2=0.159,P>0.05).Results of Matta reduction criteria(Z=2.79,P<0.05),Majeed efficacy evaluation(Z=2.79,P<0.05),CT grading(Z=2.83,P<0.05)in CT group were better than those in C-arm group(P<0.05);the secondary surgical revision rate in the CT group was significantly lower than that in the C-arm group(x2=5.641,P<0.05).Conclusion Compared with traditional C-arm fluoroscopy,intraoperative slide rail CT combined with C-arm assisted percutaneous sacroiliac joint screw placement surgery has the characteristics of short operation time,high accuracy and safety,and significant decrease in postoperative sec-ondary revision rate,and is one of the effective methods for re-establishing the stability of the posterior ring of pelvic fracture.
4.Study on the objectivity and biological mechanism of Psoralea corylifolia Linn.'s 'Great dryness damages the liver'
Ming-liang ZHANG ; Xu ZHAO ; Wei-xia LI ; Xiao-yan WANG ; Yu-long CHEN ; De-xin KONG ; Cheng-zhao WU ; Xiao-fei CHEN ; Zhao-fang BAI ; Ming NIU ; Jia-bo WANG ; Yan-ling ZHAO ; Xiao-he XIAO ; Jin-fa TANG
Acta Pharmaceutica Sinica 2023;58(4):1014-1023
According to the theory of 'Xingben Dazao'
5.Research progress on the effect mechanism of acupuncture-moxibustion for pressure injuries.
Hong-Na YIN ; Yang CUI ; Zhong-Ren SUN ; De-Long WANG ; Xin-Yu ZHOU ; Jian-Tao YIN ; Shuo ZHANG ; Gui-Jun LIU
Chinese Acupuncture & Moxibustion 2023;43(8):970-976
Evidence shows that acupuncture-moxibustion could promote the healing of pressure injuries (PI), but its action mechanism is not fully understood. This review summarizes the basic research literature of acupuncture-moxibustion for PI and identifies that the mechanism of acupuncture-moxibustion for PI is related with regulation of related signaling pathway target proteins, improvement of inflammatory response, modulation of vascular microenvironment, attenuation of oxidative stress damage, and inhibition of cell apoptosis. The review also points out the current limitations and future research directions. It emphasizes the need for further exploration of the upstream regulatory mechanism, specific cellular molecules, and the interactions among these molecules. A multi-level, multi-target, and multi-dimensional approach is required to fully understand the mechanism underlying the promotion of PI healing by acupuncture-moxibustion.
Humans
;
Moxibustion
;
Pressure Ulcer
;
Acupuncture Therapy
;
Acupuncture
;
Apoptosis
6.Clinical Value Evaluation of DR Corneal Suture Ring Localization and CT 3D Reconstruction Localization of Orbital Foreign Body
Li-xu GUO ; Qing-yu CHEN ; Zhao-hui YUAN ; Chong-de LONG ; Shi-huai NIE
Journal of Sun Yat-sen University(Medical Sciences) 2023;44(6):1016-1021
ObjectiveTo investigate the clinical value of DR corneal suture ring localization and CT 3D reconstruction localization of orbital foreign bodies. MethodsRetrospective analysis was performed on patients (51 cases) suspected of ocular foreign bodies admitted to our hospital from January 2016 to December 2020 At the same time, DR corneal suture ring localization and CT three-dimensional reconstruction localization of orbital foreign bodies were performed, and the detection rate of foreign bodies by the two methods was calculated to analyze the location of ocular foreign bodies. ResultsThere were 38 cases of intraorbital foreign body identified by DR corneal suture ring localization method, and 46 cases of intraorbital foreign body identified by CT three-dimensional reconstruction localization method. The accuracy of CT three-dimensional reconstruction localization method was 90.20%, which was higher than that of DR corneal suture ring localization method 74.51% (P<0.05). Intraocular foreign bodies in 23 cases could be distinguished by DR corneal suture ring localization, and 25 cases by CT 3D reconstruction localization. The maximum diameter of intraocular foreign bodies that could be distinguished by CT 3D reconstruction localization was (2.65±0.14) mm. The diameter of intraspherical foreign body was (2.94±0.36) mm (P<0.05) lower than that which could be distinguished by DR corneal suture ring localization method. The results of DR corneal suture ring localization of orbital foreign body showed no difference compared with the results of CT 3D reconstruction localization (P>0.05). Forty cases of high density images inside and outside the eye could be clearly distinguished by CT 3D reconstruction. By using DR corneal suture ring localization method, 23 cases were confirmed to be intraocular high-density shadows, and 15 cases could not be confirmed to be intraocular high-density shadows (P < 0.05). ConclusionsFor the location of intraorbital foreign bodies, CT three-dimensional reconstruction can be used as a conventional method for locating ocular foreign bodies with high resolution and accuracy, and can detect fine metal particles inside the eyeball. Meanwhile, for a few small foreign bodies in the iris root, ciliary body and lateral suspension ligament, it is necessary to locate orbital foreign bodies with DR corneal suture ring.
7.Population structure of food-borne Staphylococcus aureus in China.
Ya Hui GUO ; Zi Long HE ; Qing Long JI ; Hai Jian ZHOU ; Fan Liang MENG ; Xiao Feng HU ; Xiao Yue WEI ; Jun Cai MA ; Yu Hua YANG ; Wei ZHAO ; Li Jin LONG ; Xin WANG ; Jia Ming FAN ; Xiao Jie YU ; Jian Zhong ZHANG ; De HUA ; Xiao Mei YAN ; Hai Bin WANG
Chinese Journal of Epidemiology 2023;44(6):982-989
Objective: To understand the population structure of food-borne Staphylococcus (S.) aureus in China. Methods: Whole genome sequencing was used to analyze 763 food-borne S. aureus strains from 16 provinces in China from 2006 to 2020. Multilocus sequence typing (MLST), staphylococcal protein A gene (spa) typing, and staphylococcal chromosome cassettemec (SCCmec) typing were conducted, and minimum spanning tree based on ST types (STs) was constructed by BioNumerics 7.5 software. Thirty-one S. aureus strains isolated from imported food products were also included in constructing the genome phylogenetic tree. Results: A total of 90 STs (20 novel types) and 160 spa types were detected in the 763 S. aureus isolates. The 72 STs (72/90, 80.0%) were related to 22 clone complexes. The predominant clone complexes were CC7, CC1, CC5, CC398, CC188, CC59, CC6, CC88, CC15, and CC25, accounting for 82.44% (629/763) of the total. The STs and spa types in the predominant clone complexes changed over the years. The methicillin-resistant S. aureus (MRSA) detection rate was 7.60%, and 7 SCCmec types were identified. The ST59-t437-Ⅳa (17.24%, 10/58), ST239-t030-Ⅲ (12.07%, 7/58), ST59-t437-Ⅴb (8.62%, 5/58), ST338-t437-Ⅴb (6.90%, 4/58) and ST338-t441-Ⅴb (6.90%, 4/58) were the main types in MRSA strains. The genome phylogenetic tree had two clades, and the strains with the same CC, ST, and spa types clustered together. All CC7 methicillin sensitive S. aureus strains were included in Clade1, while 21 clone complexes and all MRSA strains were in Clade2. The MRSA strains clustered according to the SCCmec and STs. The strains from imported food products in CC398, CC7, CC30, CC12, and CC188 had far distances from Chinese strains in the tree. Conclusions: In this study, the predominant clone complexes of food-borne strains were CC7, CC1, CC5, CC398, CC188, CC59, CC6, CC88, CC15, and CC25, which overlapped with the previously reported clone complexes of hospital and community-associated strains in China, suggesting that close attention needs to be paid to food, a vehicle of pathogen transmission in community and food poisoning.
Humans
;
Staphylococcus aureus/genetics*
;
Methicillin-Resistant Staphylococcus aureus/genetics*
;
Multilocus Sequence Typing
;
Phylogeny
;
Staphylococcal Infections/epidemiology*
;
China/epidemiology*
8.Percutaneous kyphoplasty assisted by three dimensional printing percutaneous guide plate for the treatment of osteoporotic vertebral compression fractures.
Jiang-Long LIAO ; Li DENG ; De-Guang LI ; Hao-Min SUN ; Jian LI ; Yan SU ; Bo JIANG ; Lyu-Yu LI
China Journal of Orthopaedics and Traumatology 2023;36(5):445-449
OBJECTIVE:
To verify the safety of three dimensional printing percutaneous guide plate assisted percutaneous kyphoplasty(PKP) in the treatment of osteoporotic vertebral compression fractures(OVCFs).
METHODS:
The clinical data of 60 patients with OVCFs treated by PKP from November 2020 to August 2021 were retrospectively analyzed. There were 24 males and 36 females, aged from 72 to 86 years old with an average of (76.5±7.9) years. Routine percutaneous kyphoplasty was performed in 30 cases (conventional group) and three dimensional printing percutaneous guide plate assisted PKP was performed in 30 cases (guide plate group). Intraoperative pedicle puncture time (puncture needle to posterior vertebral body edge) and number of fluoroscopy, total operation time, total number of fluoroscopy, amount of bone cement injection, and complication (spinal canal leakage of bone cement) were observed. The visual analogue scale (VAS) and the anterior edge compression rate of the injured vertebra were compared before operation and 3 days after operation between two groups.
RESULTS:
All 60 patients were successfully operated without complication of spinal canal leakage of bone cement. In the guide plate group, the pedicle puncture time was(10.23±3.15) min and the number of fluoroscopy was(4.77±1.07) times, the total operation time was (33.83±4.21) min, the total number of fluoroscopy was(12.27±2.61) times;and in the conventional group, the pedicle puncture time was (22.83±3.09) min and the number of fluoroscopy was (10.93±1.62) times, the total operation time was(44.33±3.57) min, the total number of fluoroscopy was(19.20±2.67) times. There were statistically significant differences in the pedicle puncture time, intraoperative number of fluoroscopy, the total operation time, and the total number of fluoroscopy between the two groups(P<0.05). There was no significant difference in amount of bone cement injection between the two groups(P>0.05). There were no significant differences in VAS and the anterior edge compression rate of the injured vertebra at 3 days after operation between two groups(P>0.05).
CONCLUSION
Three dimensional printing percutaneous guide plate assisted percutaneous kyphoplasty is safe and reliable, which can reduce the number of fluoroscopy, shorten the operation time, and decrease the radiation exposure of patients and medical staff, and conforms to the concept of precise orthopaedic management.
Male
;
Female
;
Humans
;
Aged
;
Aged, 80 and over
;
Kyphoplasty/methods*
;
Fractures, Compression/surgery*
;
Spinal Fractures/surgery*
;
Bone Cements
;
Retrospective Studies
;
Treatment Outcome
;
Osteoporotic Fractures/surgery*
9.Platelet Transfusion Strategies for MASPAT-Matched Platelet Transfusion Failed Patient with Allogeneic Hematopoietic Stem Cell Transplantation.
Lu YANG ; Chun-Ya MA ; Li-Hui FU ; Sheng-Fei TAI ; Ming-Zi MA ; Xiao-Long ZHONG ; Bin FAN ; Xiao-Xing WANG ; De-Qing WANG ; Yang YU
Journal of Experimental Hematology 2023;31(3):850-854
OBJECTIVE:
To investigate the causes of ineffectiveness of platelet transfusion with monoclonal antibody solid phase platelet antibody test (MASPAT) matching in patients with allogeneic hematopoietic stem cell transplantation and explore the strategies of platelet transfusion.
METHODS:
A case of donor-specific HLA antibodies (DSA) induced by transfusion which ultimately resulted in transplantation failure and ineffective platelet transfusion with MASPAT matching was selected, and the causes of ineffective platelet transfusion and platelet transfusion strategy were retrospectively analyzed.
RESULTS:
The 32-year-old female patient was diagnosed as acute myeloid leukemia (high risk) in another hospital with the main symptoms of fever and leukopenia, who should be admitted for hematopoietic stem cell transplantation after remission by chemotherapy. In the course of chemotherapy, DSA was generated due to platelet transfusion, and had HLA gene loci incompatible with the donor of the first transplant, leading to the failure of the first transplant. The patient received platelet transfusion for several times before and after transplantation, and the results showed that the effective rate of MASPAT matched platelet transfusion was only 35.3%. Further analysis showed that the reason for the ineffective platelet transfusion was due to the missed detection of antibodies by MASPAT method. During the second hematopoietic stem cell transplantation, the DSA-negative donor was selected, and the matching platelets but ineffective transfusion during the primary transplantation were avoided. Finally, the patient was successfully transplanted and discharged from hospital.
CONCLUSIONS
DSA can cause graft failure or render the graft ineffective. For the platelet transfusion of patients with DSA, the platelet transfusion strategy with matching type only using MASPAT method will miss the detection of antibodies, resulting in invalid platelet transfusion.
Female
;
Humans
;
Adult
;
Platelet Transfusion
;
Antibodies, Monoclonal
;
Retrospective Studies
;
HLA Antigens
;
Hematopoietic Stem Cell Transplantation
10.Catheter ablation versus medical therapy for atrial fibrillation with prior stroke history: a prospective propensity score-matched cohort study.
Wen-Li DAI ; Zi-Xu ZHAO ; Chao JIANG ; Liu HE ; Ke-Xin YAO ; Yu-Feng WANG ; Ming-Yang GAO ; Yi-Wei LAI ; Jing-Rui ZHANG ; Ming-Xiao LI ; Song ZUO ; Xue-Yuan GUO ; Ri-Bo TANG ; Song-Nan LI ; Chen-Xi JIANG ; Nian LIU ; De-Yong LONG ; Xin DU ; Cai-Hua SANG ; Jian-Zeng DONG ; Chang-Sheng MA
Journal of Geriatric Cardiology 2023;20(10):707-715
BACKGROUND:
Patients with atrial fibrillation (AF) and prior stroke history have a high risk of cardiovascular events despite anticoagulation therapy. It is unclear whether catheter ablation (CA) has further benefits in these patients.
METHODS:
AF patients with a previous history of stroke or systemic embolism (SE) from the prospective Chinese Atrial Fibrillation Registry study between August 2011 and December 2020 were included in the analysis. Patients were matched in a 1:1 ratio to CA or medical treatment (MT) based on propensity score. The primary outcome was a composite of all-cause death or ischemic stroke (IS)/SE.
RESULTS:
During a total of 4.1 ± 2.3 years of follow-up, the primary outcome occurred in 111 patients in the CA group (3.3 per 100 person-years) and in 229 patients in the MT group (5.7 per 100 person-years). The CA group had a lower risk of the primary outcome compared to the MT group [hazard ratio (HR) = 0.59, 95% CI: 0.47-0.74, P < 0.001]. There was a significant decreasing risk of all-cause mortality (HR = 0.43, 95% CI: 0.31-0.61, P < 0.001), IS/SE (HR = 0.73, 95% CI: 0.54-0.97, P = 0.033), cardiovascular mortality (HR = 0.32, 95% CI: 0.19-0.54, P < 0.001) and AF recurrence (HR = 0.33, 95% CI: 0.30-0.37, P < 0.001) in the CA group compared to that in the MT group. Sensitivity analysis generated consistent results when adjusting for time-dependent usage of anticoagulants.
CONCLUSIONS
In AF patients with a prior stroke history, CA was associated with a lower combined risk of all-cause death or IS/SE. Further clinical trials are warranted to confirm the benefits of CA in these patients.

Result Analysis
Print
Save
E-mail