1.Surgical Outcomes of Weight-Bearing Shoulders:Arthroscopic Rotator Cuff Repair and Reverse Shoulder Arthroplasty
Su Cheol KIM ; Hyun Gon KIM ; Young Girl RHEE ; Sung Min RHEE ; Chul-Hyun CHO ; Du-Han KIM ; Hee Dong LEE ; Jae Chul YOO
Clinics in Orthopedic Surgery 2025;17(3):438-452
Background:
This study aimed to report the short- and midterm outcomes of arthroscopic rotator cuff repair (ARCR) and reverse shoulder arthroplasty (RSA) in weight-bearing shoulders.
Methods:
This retrospective multicenter study included 19 cases of ARCR and 10 cases of RSA performed in weight-bearing shoulders from 2009 to 2021. In the ARCR group, postoperative 6-month magnetic resonance imaging confirmed the tendon integrity. In the RSA group, scapular notching, acromial fracture, and implant failure were assessed using plain radiographs, and complications were recorded. In both groups, preoperative and postoperative range of motion and functional scores were documented, along with subjective satisfaction and arm use for weight-bearing on the shoulders. For patients followed up for > 5 years, a midterm analysis was performed.
Results:
The ARCR group included 8 men and 11 women (average age, 58.8 ± 8.0 years). Initially, Patte types 1, 2, and 3 were noted in 9, 8, and 2 patients, respectively, and 4 patients exhibited full-thickness subscapularis tears. Four patients showed supraspinatus retear, and 2 patients showed subscapularis retear. Retear of any rotator cuff was observed in 5 patients (26.3%). Twelve patients were followed up for > 5 years; 11 (91.7%) used their operated arm for weight-bearing and 9 (75.0%) were satisfied. The RSA group included 5 men and 5 women (average age, 74.3 ± 7.9 years). Procedures included RSAs for cuff tear arthropathy (n = 6), osteoarthritis (n = 3), and fracture nonunion (n = 1). No cases of dislocation, prosthesis loosening, or disassociation were observed throughout the follow-up. However, 1 patient required implant removal due to infection, and 4 patients showed stage 1 scapular notching. Five patients were followed up for > 5 years, all of whom expressed satisfaction and used their operated arms for weight-bearing, despite mean forward flexion (107.5° ± 12.6°) and American Shoulder and Elbow Surgeons score (61.5 ± 5.3) being less than reported patient acceptable symptomatic state (110° and 76, respectively).
Conclusions
Both ARCR and RSA showed promising outcomes in terms of weight-bearing on the operated arm and subjective satisfaction at short- and midterm follow-up. Therefore, neither of these surgeries should be considered contraindicated for patients with weight-bearing shoulder conditions.
2.Surgical Outcomes of Weight-Bearing Shoulders:Arthroscopic Rotator Cuff Repair and Reverse Shoulder Arthroplasty
Su Cheol KIM ; Hyun Gon KIM ; Young Girl RHEE ; Sung Min RHEE ; Chul-Hyun CHO ; Du-Han KIM ; Hee Dong LEE ; Jae Chul YOO
Clinics in Orthopedic Surgery 2025;17(3):438-452
Background:
This study aimed to report the short- and midterm outcomes of arthroscopic rotator cuff repair (ARCR) and reverse shoulder arthroplasty (RSA) in weight-bearing shoulders.
Methods:
This retrospective multicenter study included 19 cases of ARCR and 10 cases of RSA performed in weight-bearing shoulders from 2009 to 2021. In the ARCR group, postoperative 6-month magnetic resonance imaging confirmed the tendon integrity. In the RSA group, scapular notching, acromial fracture, and implant failure were assessed using plain radiographs, and complications were recorded. In both groups, preoperative and postoperative range of motion and functional scores were documented, along with subjective satisfaction and arm use for weight-bearing on the shoulders. For patients followed up for > 5 years, a midterm analysis was performed.
Results:
The ARCR group included 8 men and 11 women (average age, 58.8 ± 8.0 years). Initially, Patte types 1, 2, and 3 were noted in 9, 8, and 2 patients, respectively, and 4 patients exhibited full-thickness subscapularis tears. Four patients showed supraspinatus retear, and 2 patients showed subscapularis retear. Retear of any rotator cuff was observed in 5 patients (26.3%). Twelve patients were followed up for > 5 years; 11 (91.7%) used their operated arm for weight-bearing and 9 (75.0%) were satisfied. The RSA group included 5 men and 5 women (average age, 74.3 ± 7.9 years). Procedures included RSAs for cuff tear arthropathy (n = 6), osteoarthritis (n = 3), and fracture nonunion (n = 1). No cases of dislocation, prosthesis loosening, or disassociation were observed throughout the follow-up. However, 1 patient required implant removal due to infection, and 4 patients showed stage 1 scapular notching. Five patients were followed up for > 5 years, all of whom expressed satisfaction and used their operated arms for weight-bearing, despite mean forward flexion (107.5° ± 12.6°) and American Shoulder and Elbow Surgeons score (61.5 ± 5.3) being less than reported patient acceptable symptomatic state (110° and 76, respectively).
Conclusions
Both ARCR and RSA showed promising outcomes in terms of weight-bearing on the operated arm and subjective satisfaction at short- and midterm follow-up. Therefore, neither of these surgeries should be considered contraindicated for patients with weight-bearing shoulder conditions.
3.Surgical Outcomes of Weight-Bearing Shoulders:Arthroscopic Rotator Cuff Repair and Reverse Shoulder Arthroplasty
Su Cheol KIM ; Hyun Gon KIM ; Young Girl RHEE ; Sung Min RHEE ; Chul-Hyun CHO ; Du-Han KIM ; Hee Dong LEE ; Jae Chul YOO
Clinics in Orthopedic Surgery 2025;17(3):438-452
Background:
This study aimed to report the short- and midterm outcomes of arthroscopic rotator cuff repair (ARCR) and reverse shoulder arthroplasty (RSA) in weight-bearing shoulders.
Methods:
This retrospective multicenter study included 19 cases of ARCR and 10 cases of RSA performed in weight-bearing shoulders from 2009 to 2021. In the ARCR group, postoperative 6-month magnetic resonance imaging confirmed the tendon integrity. In the RSA group, scapular notching, acromial fracture, and implant failure were assessed using plain radiographs, and complications were recorded. In both groups, preoperative and postoperative range of motion and functional scores were documented, along with subjective satisfaction and arm use for weight-bearing on the shoulders. For patients followed up for > 5 years, a midterm analysis was performed.
Results:
The ARCR group included 8 men and 11 women (average age, 58.8 ± 8.0 years). Initially, Patte types 1, 2, and 3 were noted in 9, 8, and 2 patients, respectively, and 4 patients exhibited full-thickness subscapularis tears. Four patients showed supraspinatus retear, and 2 patients showed subscapularis retear. Retear of any rotator cuff was observed in 5 patients (26.3%). Twelve patients were followed up for > 5 years; 11 (91.7%) used their operated arm for weight-bearing and 9 (75.0%) were satisfied. The RSA group included 5 men and 5 women (average age, 74.3 ± 7.9 years). Procedures included RSAs for cuff tear arthropathy (n = 6), osteoarthritis (n = 3), and fracture nonunion (n = 1). No cases of dislocation, prosthesis loosening, or disassociation were observed throughout the follow-up. However, 1 patient required implant removal due to infection, and 4 patients showed stage 1 scapular notching. Five patients were followed up for > 5 years, all of whom expressed satisfaction and used their operated arms for weight-bearing, despite mean forward flexion (107.5° ± 12.6°) and American Shoulder and Elbow Surgeons score (61.5 ± 5.3) being less than reported patient acceptable symptomatic state (110° and 76, respectively).
Conclusions
Both ARCR and RSA showed promising outcomes in terms of weight-bearing on the operated arm and subjective satisfaction at short- and midterm follow-up. Therefore, neither of these surgeries should be considered contraindicated for patients with weight-bearing shoulder conditions.
4.Surgical Outcomes of Weight-Bearing Shoulders:Arthroscopic Rotator Cuff Repair and Reverse Shoulder Arthroplasty
Su Cheol KIM ; Hyun Gon KIM ; Young Girl RHEE ; Sung Min RHEE ; Chul-Hyun CHO ; Du-Han KIM ; Hee Dong LEE ; Jae Chul YOO
Clinics in Orthopedic Surgery 2025;17(3):438-452
Background:
This study aimed to report the short- and midterm outcomes of arthroscopic rotator cuff repair (ARCR) and reverse shoulder arthroplasty (RSA) in weight-bearing shoulders.
Methods:
This retrospective multicenter study included 19 cases of ARCR and 10 cases of RSA performed in weight-bearing shoulders from 2009 to 2021. In the ARCR group, postoperative 6-month magnetic resonance imaging confirmed the tendon integrity. In the RSA group, scapular notching, acromial fracture, and implant failure were assessed using plain radiographs, and complications were recorded. In both groups, preoperative and postoperative range of motion and functional scores were documented, along with subjective satisfaction and arm use for weight-bearing on the shoulders. For patients followed up for > 5 years, a midterm analysis was performed.
Results:
The ARCR group included 8 men and 11 women (average age, 58.8 ± 8.0 years). Initially, Patte types 1, 2, and 3 were noted in 9, 8, and 2 patients, respectively, and 4 patients exhibited full-thickness subscapularis tears. Four patients showed supraspinatus retear, and 2 patients showed subscapularis retear. Retear of any rotator cuff was observed in 5 patients (26.3%). Twelve patients were followed up for > 5 years; 11 (91.7%) used their operated arm for weight-bearing and 9 (75.0%) were satisfied. The RSA group included 5 men and 5 women (average age, 74.3 ± 7.9 years). Procedures included RSAs for cuff tear arthropathy (n = 6), osteoarthritis (n = 3), and fracture nonunion (n = 1). No cases of dislocation, prosthesis loosening, or disassociation were observed throughout the follow-up. However, 1 patient required implant removal due to infection, and 4 patients showed stage 1 scapular notching. Five patients were followed up for > 5 years, all of whom expressed satisfaction and used their operated arms for weight-bearing, despite mean forward flexion (107.5° ± 12.6°) and American Shoulder and Elbow Surgeons score (61.5 ± 5.3) being less than reported patient acceptable symptomatic state (110° and 76, respectively).
Conclusions
Both ARCR and RSA showed promising outcomes in terms of weight-bearing on the operated arm and subjective satisfaction at short- and midterm follow-up. Therefore, neither of these surgeries should be considered contraindicated for patients with weight-bearing shoulder conditions.
5.Evaluating the Influence of Scan Timing on Dosimetric Accuracy in EBT3 and EBT4 Radiochromic Films
Jin Dong CHO ; Su Chul HAN ; Jason Joon Bock LEE ; Hyebin LEE ; Heerim NAM
Progress in Medical Physics 2024;35(4):135-144
Purpose:
This study compares the dosimetric properties of EBT3 and EBT4 GAFchromic films in transmission and reflection scanning modes, focusing on dose response, sensitivity, and postirradiation stability.
Methods:
The EBT3 and EBT4 films were irradiated at doses of 0–10 Gy using a Varian TrueBeam linear accelerator at 6 MV. The films were scanned at intervals between 1 and 336 hours after irradiation in both transmission and reflection modes. Net optical density (NetOD) values from each scan were used to evaluate dose response and sensitivity, with calibration curves created for each film and scan mode. Dose differences between calculated and delivered doses were assessed over time.
Results:
The EBT3 and EBT4 films exhibited similar dose–response curves and stable NetOD values across both scanning modes. However, EBT4 exhibited reduced sensitivity variability in response to dose changes. After irradiation, NetOD values increased up to 24 hours before stabilizing, suggesting that a 24-hour scan time is sufficient for consistent measurements. Dose differences between films and modes remained within ±4%.
Conclusions
EBT4 offers comparable dosimetric performance to EBT3, with additional benefits, such as improved dose–response linearity and reduced sensitivity fluctuations. The findings indicate that EBT4 can serve as a reliable successor to EBT3.
6.Evaluating the Influence of Scan Timing on Dosimetric Accuracy in EBT3 and EBT4 Radiochromic Films
Jin Dong CHO ; Su Chul HAN ; Jason Joon Bock LEE ; Hyebin LEE ; Heerim NAM
Progress in Medical Physics 2024;35(4):135-144
Purpose:
This study compares the dosimetric properties of EBT3 and EBT4 GAFchromic films in transmission and reflection scanning modes, focusing on dose response, sensitivity, and postirradiation stability.
Methods:
The EBT3 and EBT4 films were irradiated at doses of 0–10 Gy using a Varian TrueBeam linear accelerator at 6 MV. The films were scanned at intervals between 1 and 336 hours after irradiation in both transmission and reflection modes. Net optical density (NetOD) values from each scan were used to evaluate dose response and sensitivity, with calibration curves created for each film and scan mode. Dose differences between calculated and delivered doses were assessed over time.
Results:
The EBT3 and EBT4 films exhibited similar dose–response curves and stable NetOD values across both scanning modes. However, EBT4 exhibited reduced sensitivity variability in response to dose changes. After irradiation, NetOD values increased up to 24 hours before stabilizing, suggesting that a 24-hour scan time is sufficient for consistent measurements. Dose differences between films and modes remained within ±4%.
Conclusions
EBT4 offers comparable dosimetric performance to EBT3, with additional benefits, such as improved dose–response linearity and reduced sensitivity fluctuations. The findings indicate that EBT4 can serve as a reliable successor to EBT3.
7.Liquid collagen from freshwater fish skin ameliorates hydration, roughness and elasticity in photo-aged skin: a randomized, controlled, clinical study
Joong Su LEE ; Yeo Cho YOON ; Ju Myung KIM ; Yun-Ho KIM ; Young-Hee KANG ; Yong Chul SHIN
Nutrition Research and Practice 2024;18(3):357-371
BACKGROUND/OBJECTIVES:
Collagen is commonly used in diverse forms as a functional component in skincare products. On the other hand, the effects of collagen on human skin are controversial. Dietary collagen hydrolysates from freshwater Pangasius hypophthalmus fish skin ameliorated photo-aged skin of hairless mice. This study conducted a randomized, double-blind, placebo-controlled clinical trial to determine if liquid fish collagen (CollagenTripep20™, Tripep20) as a drink strengthens skin health and quality.
SUBJECTS/METHODS:
In this clinical trial, 85 subjects aged 35–60 yrs were diagnosed with photo-aged skin. Eighty-five subjects were randomized to receive either Tripep20 (n = 44) or placebo (n = 41). Seventy-eight subjects fully participating for a 12-week period consumed 1,000 mg of Tripep20 (n = 41) or placebo (n = 37) in a 50-mL bottle as a daily drink. The intend-to-treat and per-protocol populations were 85 and 78, respectively. Skin hydration, wrinkles, and elasticity were assessed at 0 (baseline), 6, and 12 weeks during the study period.
RESULTS:
Skin hydration in the Tripep20 group was significantly higher from 6 weeks (P < 0.001) than the baseline. After 12 weeks, the Crow’s-feet visual score and skin roughness (Ra , Rq , and Rmax ) were significantly improved in the Tripep20 group than in the placebo group (P < 0.05). Consuming liquid collagen Tripep20 greatly enhanced skin elasticity (Gross R2, Net R5, and Biological elasticity R7) in 6 weeks compared to the placebo group. The Tripep20 group showed a significant increase in skin elasticity from the baseline after 6 and 12 weeks (P < 0.001). Neither abnormal symptoms nor adverse events were encountered during the study period in subjects ingesting Tripep20 or placebo. The changes in parameters related to hematology and clinical chemistry were within the normal ranges.
CONCLUSION
Oral consumption of liquid collagen Tripep20 was safe and well-tolerated.The results of this study show that freshwater fish-derived liquid collagen Tripep20 can be used as a healthy functional food ingredient to improve skin moisturizing, anti-wrinkling, and elasticity in an aging population.
8.The impact of severe depression on the survival of older patients with end-stage kidney disease
You Hyun JEON ; Jeong-Hoon LIM ; Yena JEON ; Yu-Kyung CHUNG ; Yon Su KIM ; Shin-Wook KANG ; Chul Woo YANG ; Nam-Ho KIM ; Hee-Yeon JUNG ; Ji-Young CHOI ; Sun-Hee PARK ; Chan-Duck KIM ; Yong-Lim KIM ; Jang-Hee CHO
Kidney Research and Clinical Practice 2024;43(6):818-828
Incidence of depression increases in patients with end-stage kidney disease (ESKD). We evaluated the association between depression and mortality among older patients with ESKD, which has not been studied previously. Methods: This nationwide prospective cohort study included 487 patients with ESKD aged >65 years, who were categorized into minimal, mild-to-moderate, and severe depression groups based on their Beck Depression Inventory-II (BDI-II) scores. Predisposing factors for high BDI-II scores and the association between the scores and survival were analyzed. Results: The severe depression group showed a higher modified Charlson comorbidity index value and lower serum albumin, phosphate, and uric acid levels than the other depression groups. The Kaplan-Meier curve revealed a significantly lower survival in the severe depression group than in the minimal and mild-to-moderate depression groups (p = 0.011). Multivariate Cox regression analysis confirmed that severe depression was an independent risk factor for mortality in the study cohort (hazard ratio, 1.39; 95% confidence interval, 1.01–1.91; p = 0.041). Additionally, BDI-II scores were associated with modified Charlson comorbidity index (p = 0.009) and serum albumin level (p = 0.004) in multivariate linear regression. Among the three depressive symptoms, higher somatic symptom scores were associated with increased mortality. Conclusion: Severe depression among older patients with ESKD increases mortality compared with minimal or mild-to-moderate depression, and patients with concomitant somatic symptoms require careful management of their comorbidities and nutritional status.
9.Comparison of Clinical Outcomes between Drug-Eluting Balloons and Drug-Eluting Stents in Patients with Small Coronary Artery Disease
Man Su KIM ; Seong Ho PARK ; Seok OH ; Dae Yong HYUN ; Seung Hun LEE ; Yong Hwan LIM ; Jun Ho AHN ; Kyung Hoon CHO ; Min Chul KIM ; Doo Sun SIM ; Young Joon HONG ; Ju Han KIM ; Youngkeun AHN ; Myung Ho JEONG
Korean Journal of Medicine 2024;99(5):253-262
Background/Aims:
Drug-eluting balloons (DEBs) represent a novel therapeutic approach for patients with small coronary artery disease. However, further studies are needed to compare the clinical efficacy of DEBs versus drug-eluting stents (DESs).
Methods:
In total, 492 patients (age, 67.9 ± 11.0 years; 339 men) with small coronary artery lesions (diameter < 2.75 mm) were randomly assigned to group I (DEB) (n = 104; age, 67.2 ± 10.7 years; 83 men) and group II (DES) (n = 388; age, 68.0 ± 11.1 years; 254 men). For inverse probability of treatment weighting (IPTW) analysis, the study population was stratified into groups I (n = 269) and II (n = 280). We compared the incidences of major adverse cardiac events (MACE) between the two groups during 12 months of clinical follow-up.
Results:
Group I had shorter device lengths (22.4 ± 5.8 mm) compared with group II (27.4 ± 9.3 mm; p < 0.001). Additionally, devices in group I were smaller in diameter (2.4 ± 0.1 mm) compared with those in group II (2.6 ± 0.1 mm; p < 0.001). Left ventricular ejection fraction (LVEF) was lower in group I (53.8% ± 12.6%) than in group II (58.6% ± 11.9%; p < 0.001). After IPTW, no significant differences in LVEF were observed between groups I and II. During 12 months of follow-up, the incidence of total MACE did not differ between the two groups.
Conclusions
No significant differences were observed in clinical efficacy between DEB and DES for the treatment of small coronary artery disease. Therefore, DEB can be considered a viable alternative to DES in patients with small coronary artery disease.
10.Evaluating the Influence of Scan Timing on Dosimetric Accuracy in EBT3 and EBT4 Radiochromic Films
Jin Dong CHO ; Su Chul HAN ; Jason Joon Bock LEE ; Hyebin LEE ; Heerim NAM
Progress in Medical Physics 2024;35(4):135-144
Purpose:
This study compares the dosimetric properties of EBT3 and EBT4 GAFchromic films in transmission and reflection scanning modes, focusing on dose response, sensitivity, and postirradiation stability.
Methods:
The EBT3 and EBT4 films were irradiated at doses of 0–10 Gy using a Varian TrueBeam linear accelerator at 6 MV. The films were scanned at intervals between 1 and 336 hours after irradiation in both transmission and reflection modes. Net optical density (NetOD) values from each scan were used to evaluate dose response and sensitivity, with calibration curves created for each film and scan mode. Dose differences between calculated and delivered doses were assessed over time.
Results:
The EBT3 and EBT4 films exhibited similar dose–response curves and stable NetOD values across both scanning modes. However, EBT4 exhibited reduced sensitivity variability in response to dose changes. After irradiation, NetOD values increased up to 24 hours before stabilizing, suggesting that a 24-hour scan time is sufficient for consistent measurements. Dose differences between films and modes remained within ±4%.
Conclusions
EBT4 offers comparable dosimetric performance to EBT3, with additional benefits, such as improved dose–response linearity and reduced sensitivity fluctuations. The findings indicate that EBT4 can serve as a reliable successor to EBT3.