1.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
		                        		
		                        			 Purpose:
		                        			This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment. 
		                        		
		                        			Materials and Methods:
		                        			Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan. 
		                        		
		                        			Results:
		                        			The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing. 
		                        		
		                        			Conclusion
		                        			MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing. 
		                        		
		                        		
		                        		
		                        	
2.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
		                        		
		                        			 Purpose:
		                        			This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment. 
		                        		
		                        			Materials and Methods:
		                        			Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan. 
		                        		
		                        			Results:
		                        			The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing. 
		                        		
		                        			Conclusion
		                        			MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing. 
		                        		
		                        		
		                        		
		                        	
3.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
		                        		
		                        			 Purpose:
		                        			This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment. 
		                        		
		                        			Materials and Methods:
		                        			Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan. 
		                        		
		                        			Results:
		                        			The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing. 
		                        		
		                        			Conclusion
		                        			MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing. 
		                        		
		                        		
		                        		
		                        	
4.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
		                        		
		                        			 Purpose:
		                        			This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment. 
		                        		
		                        			Materials and Methods:
		                        			Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan. 
		                        		
		                        			Results:
		                        			The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing. 
		                        		
		                        			Conclusion
		                        			MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing. 
		                        		
		                        		
		                        		
		                        	
5.Planning evaluation of stereotactic magnetic resonance–guided online adaptive radiosurgery for kidney tumors close to the organ at risk: is it valuable to wait for good timing to perform stereotactic radiosurgery?
Takaya YAMAMOTO ; Shohei TANAKA ; Noriyoshi TAKAHASHI ; Rei UMEZAWA ; Yu SUZUKI ; Keita KISHIDA ; So OMATA ; Kazuya TAKEDA ; Hinako HARADA ; Kiyokazu SATO ; Yoshiyuki KATSUTA ; Noriyuki KADOYA ; Keiichi JINGU
Radiation Oncology Journal 2025;43(1):40-48
		                        		
		                        			 Purpose:
		                        			This study aimed to investigate changes in target coverage using magnetic resonance–guided online adaptive radiotherapy (MRgoART) for kidney tumors and to evaluate the suitable timing of treatment. 
		                        		
		                        			Materials and Methods:
		                        			Among patients treated with 3-fraction MRgoART for kidney cancer, 18 tumors located within 1 cm of the gastrointestinal tract were selected. Stereotactic radiosurgery planning with a prescription dose of 26 Gy was performed using pretreatment simulation and three MRgoART timings with an adapt-to-shape method. The best MRgoART plan was defined as the plan achieving the highest percentage of planning target volume (PTV) coverage of 26 Gy. In clinical scenario simulation, MRgoART plans were evaluated in the order of actual treatment. Waiting for the next timing was done when the PTV coverage of 26 Gy did not achieve 95%–99% or did not increase by 5% or more compared to the pretreatment plan. 
		                        		
		                        			Results:
		                        			The median percentages of PTV receiving 26 Gy in pretreatment and the first, second, and third MRgoART were 82% (range, 19%), 63% (range, 7% to 99%), 88% (range, 31% to 99%), and 95% (range, 3% to 99%), respectively. Comparing pretreatment simulation plans with the best MRgoART plans showed a significant difference (p = 0.025). In the clinical scenario simulation, 16 of the 18 planning series, including nine plans with 95%–99% PTV coverage of 26 Gy and seven plans with increased PTV coverage by 5% or more, would be irradiated at a good timing. 
		                        		
		                        			Conclusion
		                        			MRgoART revealed dose coverage differences at each MRgoART timing. Waiting for optimal irradiation timing could be an option in case of suboptimal timing. 
		                        		
		                        		
		                        		
		                        	
6.Risk Factors for Loosening of S2 Alar Iliac Screw: Surgical Outcomes of Adult Spinal Deformity
Yasushi IIJIMA ; Toshiaki KOTANI ; Tsuyoshi SAKUMA ; Keita NAKAYAMA ; Tsutomu AKAZAWA ; Shunji KISHIDA ; Yuta MURAMATSU ; Yu SASAKI ; Keisuke UENO ; Tomoyuki ASADA ; Kosuke SATO ; Shohei MINAMI ; Seiji OHTORI
Asian Spine Journal 2020;14(6):864-871
		                        		
		                        			 Methods:
		                        			Cases of 50 patients with ASD who underwent long spinal fusion (>9 levels) with S2AI screws were retrospectively reviewed. Loosening of S2AI screws and S1 pedicle screws and bone fusion at the level of L5–S1 at 2 years after surgery were investigated using computed tomography. In addition, risk factors for loosening of S2AI screws were determined in patients with ASD.  
		                        		
		                        			Results:
		                        			At 2 years after surgery, 33 cases (66%) of S2AI screw loosening and six cases (12%) of S1 pedicle screw loosening were observed. In 40 of 47 cases (85%), bone fusion at L5–S1 was found. Pseudarthrosis at L5–S1 was not significantly associated with S2AI screw loosening (19.3% vs. 6.3%, p=0.23), but significantly higher in patients with S1 screw loosening (83.3% vs. 4.9%, p<0.001). On multivariate logistic regression analyses, high upper instrumented vertebra (UIV) level (T5 or above) (odds ratio [OR], 4.4; 95% confidence interval [CI], 1.0–18.6; p=0.045) and obesity (OR, 11.4; 95% CI, 1.2–107.2; p=0.033) were independent risk factors for S2AI screw loosening. 
		                        		
		                        			Conclusions
		                        			High UIV level (T5 or above) and obesity were independent risk factors for S2AI screw loosening in patients with lumbosacral fixation in surgery for ASD. The incidence of lumbosacral fusion is associated with S1 screw loosening, but not S2AI screw loosening. 
		                        		
		                        		
		                        		
		                        	
7.Households with insufficient bednets in a village with sufficient bednets: evaluation of household bednet coverage using bednet distribution index in Xepon district, Lao PDR
Daisuke Nonaka ; Tiengkham Pongvongsa ; Futoshi Nishimoto ; Phetsomphon Nansounthavong ; Yu Sato ; Hongwei Jiang ; Rie Takeuchi ; Kazuhiko Moji ; Panom Phongmany ; Jun Kobayashi
Tropical Medicine and Health 2015;advpub(0):-
		                        		
		                        			
		                        			InLao PDR, the National Malaria Control Program (NMCP) evaluates bednet coverage,often at the village level, using a coverage target of one net per 2.5 (or fewer)persons in a given population. However, in villages that meet the target, notall households necessarily meet the target or utilize all available bednets. Thisstudy explored households that fell short of the target and household utilizationof bednets in villages that met the target of bednet coverage set by the NMCP. Thepersons per net ratio (PPNR), which is defined as the population divided by thenumber of available bednets in a household/village, was used to determine whethera household/village met the NMCP target. Using a household survey, we collectedand analyzed the data of 635 households in 17 villages in Xepon district in2012. Households that fell short of the target (households with a PPNR of >2.5 or no bednet) existed in every village. The proportion of these households differedgreatly among the villages, ranging from 3.4–50%, with some households fallingfar short. Of the 635 households, 275 (43.5%) had at least one bednet that wasnot being used on the night preceding the survey and 131 (20.6%) had at leasttwo. In conclusion, in villages that met the NMCP target, a considerable numberof households fell short of the target. Available bednets were not fullyutilized in many of the surveyed households.
		                        		
		                        		
		                        		
		                        	
8.Households with Insufficient Bednets in a Village with Sufficient Bednets: Evaluation of Household Bednet Coverage Using Bednet Distribution Index in Xepon District, Lao PDR
Daisuke Nonaka ; Tiengkham Pongvongsa ; Futoshi Nishimoto ; Phetsomphon Nansounthavong ; Yu Sato ; Hongwei Jiang ; Rie Takeuchi ; Kazuhiko Moji ; Panom Phongmany ; Jun Kobayashi
Tropical Medicine and Health 2015;43(2):95-110
		                        		
		                        			
		                        			In Lao PDR, the National Malaria Control Program (NMCP) evaluates bednet coverage, often at the village level, using a coverage target of one net per 2.5 (or fewer) persons in a given population. However, in villages that meet the target, not all households necessarily meet the target or utilize all available bednets. This study explored households that fell short of the target and household utilization of bednets in villages that met the target of bednet coverage set by the NMCP. The person per net ratio (PPNR), which is defined as the population divided by the number of available bednets in a household/village, was used to determine whether a household/village met the NMCP target. Using a household survey, we collected and analyzed the data of 635 households in 17 villages in Xepon district in 2012. Households that fell short of the target (households with a PPNR of > 2.5 or no bednet) existed in every village. The proportion of these households differed greatly among the villages, ranging from 3.4–50%, with some households falling far short. Of the 635 households, 275 (43.5%) had at least one bednet that was not being used on the night preceding the survey and 131 (20.6%) had at least two. In conclusion, in villages that met the NMCP target, a considerable number of households fell short of the target, and the available bednets were not fully utilized in many of the surveyed households.
		                        		
		                        		
		                        		
		                        	
9.Multiple primary malignant neoplasms of three early cancer lesions: a case report.
Wan-jun ZHANG ; Xiao-ping QIAN ; Yu SHI ; Wen-sheng PAN ; Xiang XU ; Zai-yuan YE ; Liang-qin WU ; Takeshi TERAI ; Nobuhiro SATO ; Sumio WATANABE
Chinese Medical Journal 2011;124(8):1278-1280
		                        		
		                        			
		                        			Multiple primary malignant neoplasms (MPMNs) are rarely reported and it is important to give early diagnosis and proper therapy for these patients. Here reported a case of 62-year-old man with concomitant three early stage cancer lesions in upper gastrointestinal tract, all of which were detected by endoscopy. The first one was an IIc-type lesion at angular part of stomach under endoscopy, which was histologically confirmed to be a mucosal well-differentiated adenocarcinoma. The patient underwent a standard radical gastrectomy for the lesion after the failure of endoscopic treatment. The other two neoplasms were observed during follow-up and were indicated as early stage lesions by synthesizing information from endoscopy, endoscopic ultrasonography, computed tomography and biopsy. One displayed as a hyperemic patch (3 cm×4 cm in size) located at the part of esophagus 27 cm away from the incisor teeth and was proved to be moderately differentiated squamous cancer by histopathological examination. The other was an IIc-type lesion (3.0 cm×3.5 cm in size) located at the part of esophagus 36 cm away from the incisor teeth, and the biopsy result showed a poorly differentiated squamous carcinoma. Both the two lesions were treated with radical radiation because the patient refused surgery management. No recurrence of former lesions or occurrence of novel lesions were observed during post-treatment follow-up, suggesting radical radiation might be effective for this patient.
		                        		
		                        		
		                        		
		                        			Gastrointestinal Neoplasms
		                        			;
		                        		
		                        			diagnosis
		                        			;
		                        		
		                        			Humans
		                        			;
		                        		
		                        			Male
		                        			;
		                        		
		                        			Middle Aged
		                        			;
		                        		
		                        			Neoplasms, Multiple Primary
		                        			;
		                        		
		                        			diagnosis
		                        			;
		                        		
		                        			Upper Gastrointestinal Tract
		                        			;
		                        		
		                        			pathology
		                        			
		                        		
		                        	
10.The Anxiety for Medical Services System in Inbound Tourists to Japan
Shouko YAMAGISHI ; Yumiko SAKUMA ; Kiyoko MIYAUCHI ; Ayako MATUMOTO ; Saori HORIKAWA ; Yu SHIBUI ; Saori AOKI ; Chifumi SATO
Journal of International Health 2008;23(4):273-279
		                        		
		                        			
		                        			Background
 In recent years, inbound tourists to Japan are increasing. Although they rarely suffer diseases or injuries during travels, they may feel anxiety for medical services in foreign countries. This study was aimed to consider how inbound tourists visiting Japan feel about Japanese medical cares.
Methods
 Inbound tourists who came to the Tokyo Sightseeing Information Center were investigated. The questionnaire composed of 12 items relating to Japanese medical cares and levels was made in three languages; English, Chinese and Korean.
Results
 A total of 163 tourists including 98 English speakers, 39 Korean speakers and 26 Chinese speakers were enrolled in the study. (1) Native English speakers had a high anxiety about the communication with Japanese medical staffs (p<0.001). (2) Chinese and Korean speakers subjective estimated the Japanese medical level low (p <0.001). (3) Chinese and Korean speakers had a high anxiety for medical costs while traveling (p<0.001). (4) In a case of emergency, English speakers mostly “look for a drugstore and a hospital”, while Chinese and Korean speakers mostly “contact a travel agency and a tour conductor”.
Conclusion
 Considering foreign tourists' needs vary with their nationality and culture, Medical staffs should support them to have desired medical care in Japan.
		                        		
		                        		
		                        		
		                        	
            

Result Analysis
Print
Save
E-mail