1.Preclinical Image Quality Evaluation of Simultaneous Vision Intraocular Lenses.
Qin FENG ; Qi HAO ; Ting SONG ; Qionghui CHEN
Chinese Journal of Medical Instrumentation 2023;47(1):93-98
Three different preclinical evaluation methods of MTF through-frequency response, MTF through-focus-response and expected visual acuity were used to compare and analyze the imaging differences of IOLs with four different optical designs. The research work could be used in the simultaneous vision IOLs in the optical design stage and verify the optical quality of the IOLs, the results can predict the visual representation of the patients better. The evaluation results can provide reference for IOL manufacturers and users in product design, development, validation and application selection.
Humans
;
Prosthesis Design
;
Lenses, Intraocular
;
Vision, Ocular
;
Visual Acuity
2.Study on Clear Optic Evaluation Method of Intraocular Lenses.
Qin FENG ; Qi HAO ; Qionghui CHEN ; Yongjie LUO ; Ting SONG
Chinese Journal of Medical Instrumentation 2022;46(5):565-569
The study explored the methods of determining the diameter of clear optic of intraocular lenses by analyzing the design principles and influencing factors of clear optic. Measure and compare the clear optic diameters of intraocular lenses from -10.0 D to 36.0 D respectively. The measurement results of the two methods have good consistency and can be used in the measurement of clear optic. But considering the accuracy and convenience of actual measurement, profile projector measurement principle is more suitable for conventional mass production testing. The Schlieren technique method can provide more useful information in product development and feature description stage, and can be used as an auxiliary verification method for clear optic diameter measurement.
Eye
;
Lenses, Intraocular
;
Optics and Photonics
;
Prosthesis Design
3.Advances in interdisciplinary medical and engineering research of intraocular lens surface modifications to prevent posterior capsule opacification.
Journal of Central South University(Medical Sciences) 2022;47(12):1754-1762
Posterior capsule opacification (PCO), a common complication after cataract surgery, impacts a patient's long-term visual quality to various degrees. Although a neodymium:yttrium aluminum garnet (Nd:YAG) laser posterior capsulotomy is a very effective treatment, it may lead to a serial of complications. Accordingly, the search for simple, safe, and effective methods to prevent PCO has received widespread attention. Various researchers are committed to the interdisciplinary collaboration between medicine and engineering fields, such as functionalizing the surface of the intraocular lens (IOL) via supercritical fluid impregnation, coating the surface of the IOL, high-concentration drug immersion, and application of a drug delivery system, to effectively reduce the incidence and severity of PCO.
Humans
;
Capsule Opacification/surgery*
;
Lens Implantation, Intraocular
;
Cataract/etiology*
;
Lens Capsule, Crystalline/surgery*
;
Lenses, Intraocular/adverse effects*
;
Treatment Outcome
;
Postoperative Complications
;
Prosthesis Design
4.Comparison of Three Formulas for Intraocular Lens Power Formula Accuracy
Ki Woong LEE ; Jinsoo KIM ; Dong Hyun KIM
Journal of the Korean Ophthalmological Society 2020;61(1):27-33
PURPOSE: To compare the accuracy of three intraocular lens (IOL) power calculation formulas (SRK/T, Barrett Universal II, and T2) in cataract surgery patients.METHODS: In total, 73 eyes of 73 patients who underwent uneventful cataract surgery were retrospectively reviewed. IOL power was determined using SRK/T, Barrett Universal II, and T2 preoperatively. The findings were compared with the actual refractive outcome to obtain the prediction error. The mean prediction error (ME) and mean absolute error (MAE) of each formula were compared. The MAE was defined as the difference between the postoperative spherical equivalence (SE) and the preoperatively predicted SE. The ME and MAE of each formula 3 months after surgery were compared with preoperatively predicted SE. Eyes were classified into subgroups based on axial length (AL) and average keratometry (K).RESULTS: The ME and MAE for the three formulas were SRK/T [−0.08 ± 0.45 diopters (D) and 0.35 ± 0.40 D, respectively], Barrett Universal II (−0.01 ± 0.44 D and 0.33 ± 0.30 D, respectively), and T2 (0.04 ± 0.45 D and −0.34 ± 0.30 D, respectively), but no statistically significant differences were detected. Similar results were obtained in groups with a long AL or a large average K. In groups with an AL ≥ 26 mm or with an average K ≥ 47 D, the Barrett Universal II formula yielded the smallest standard deviation and a ME closest to zero, but these differences were not statistically significant.CONCLUSIONS: No significant differences were observed between the three formulas regarding ME or MAE. However, recent formulas such as the Barrett Universal II could provide certain benefits in predicting IOL power for patients with a long AL (> 26 mm) or larger average K. Further research with a larger sample size is recommended for more evaluation.
Cataract
;
Humans
;
Lenses, Intraocular
;
Retrospective Studies
;
Sample Size
5.Anterior Chamber and Lens Position before and after Phacoemulsification According to Axial Length
Suk Hoon JUNG ; Seonjoo KIM ; So Hyang CHUNG
Journal of the Korean Ophthalmological Society 2020;61(1):17-26
PURPOSE: In this study, we evaluated changes in the anterior chamber structure and lens position before and after phacoemulsification in eyes grouped by axial length (AL).METHODS: This study included 65 eyes (16 short eyes [AL < 22.5 mm], 33 normal eyes [22.5 mm < AL < 25.0 mm], and 16 long eyes [AL > 25.5 mm]) that underwent cataract surgery. Pre- and postoperative anterior chamber depth (ACD) was measured using Pentacam® and IOL Master®,. In addition, we evaluated the anterior chamber angle (ACA), anterior chamber volume (ACV), epithelium-iris distance, and iris-lens (intraocular lens [IOL]) distance.RESULTS: The change in ACD was significantly smaller in long eyes (Pentacam®,, p = 0.000; IOL Master®,, p = 0.001). The change in ACA was significantly larger in short eyes (p = 0.000), and the change in ACV was significantly smaller in long eyes (p = 0.000). The change in the epithelium–iris distance was significantly smaller in long eyes (p = 0.000), and the change in the iris-lens (IOL) distance was significantly smaller in short eyes (p = 0.000).CONCLUSIONS: In short eyes, changes in ACD, ACA, and ACV were found to be larger than those of other groups as the iris moved backward. In long eyes, greater backward movement of the IOL was observed. Therefore, the appropriate IOL power should be chosen, considering the postoperative position of the IOL during cataract surgery of short and long eyes.
Anterior Chamber
;
Cataract
;
Iris
;
Lenses, Intraocular
;
Phacoemulsification
6.Manual Preparation of Donor Lenticule Using Artificial Anterior Chamber for Descemet's Membrane Stripping Endothelial Keratoplasty
Gye Jung KIM ; Min Chul SHIN ; Ho Sik HWANG
Journal of the Korean Ophthalmological Society 2020;61(2):209-213
PURPOSE: To report a patient with a pseudophakic bullous keratopathy (PBK) who underwent Descemet's membrane stripping endothelial keratoplasty (DSEK) with manual preparation of the donor corneal graft.CASE SUMMARY: A 61-year-old female presented with visual disturbance in her right eye. Five months prior, she was treated with phacoemulsification and intraocular lens exchange surgery of the right eye, and a very severe corneal edema was revealed by slit-lamp examination. We diagnosed PBK and planned DSEK with manual preparation of a donor corneal graft because of the non-availability of a microkeratome or a femtosecond laser. After making the corneal graft using an artificial anterior chamber, crescent knife and cornea dissector, the keratoplasty proceeded using the graft. Three months after surgery, her graft was well-maintained on the right eye. The patient's visual acuity was 0.3, and the corneal endothelial cell count was 1,844/mm².CONCLUSIONS: Manual preparation of the donor corneal graft for DSEK is suitable as a second choice treatment method when the availability of surgical devices is limited.
Anterior Chamber
;
Cornea
;
Corneal Edema
;
Corneal Transplantation
;
Descemet Membrane
;
Endothelial Cells
;
Female
;
Humans
;
Lenses, Intraocular
;
Methods
;
Middle Aged
;
Phacoemulsification
;
Tissue Donors
;
Transplants
;
Visual Acuity
7.Measuring Defocus Curves of Monofocal, Multifocal and Extended Depth-of-focus Intraocular Lenses Using Optical Bench Test
Yong Ho KOO ; Chang Su LEE ; Jin Soo KIM ; Min Chul SHIN ; Eun Chul KIM ; Man Soo KIM ; Ho Sik HWANG
Journal of the Korean Ophthalmological Society 2020;61(2):153-158
PURPOSE: To determine the through-focus optical bench test performance of monofocal, bifocal, and extended depth-of-focus intraocular lenses (IOLs), and to measure their defocus curves.METHODS: A model eye was placed on an optical bench to test three different IOLs (TECNIS ZXR00, ZMB00, and ZCB00; Abbott Medical Optics, Santa Ana, CA, USA). The focus was changed by inserting trial lenses from +1.00 diopters to −4.00 diopters, in increments of +0.25 diopters. The 1951 United States Air Force Resolution chart was used to determine the quality of the images. The degree of similarity with reference images was given by the cross-correlation coefficient, and defocus curves were drawn and compared.RESULTS: Bifocal IOLs showed lower image quality with the addition of minus diopter trial lenses, but showed good image quality at near distance. Bifocal IOLs also showed a ‘double peak’ in their defocus curve. Monofocal IOLs showed a lower image quality and cross-correlation coefficient with addition of lower-diopter trial lenses. The extended depth of focus IOLs showed a single peak in their defocus curve, but had a wider range of diopters and better image quality than monofocal IOLs.CONCLUSIONS: Bifocal IOLs showed a double peak defocus curve, and extended depth of focus IOLs showed a wider diopter range and better image quality than monofocal IOLs.
Lenses, Intraocular
;
United States
8.A comparative evaluation of visual, refractive, and patient-reported outcomes of three diffractive trifocal intraocular lenses
Robert Edward T. Ang ; Janice Marie N. Jordan-Yu ; Mark Sylvester F. Agas ; Ryan S. Torres ; Emerson M. Cruz
Philippine Journal of Ophthalmology 2020;45(1):28-40
OBJECTIVE: To compare the visual, refractive, and patient-reported outcomes of eyes implanted with one of 3 trifocal intraocular lenses (IOLs).
METHODS: This is a cross-sectional, comparative, non-interventional study wherein subjects implanted with FineVision Micro F, AT LISA tri 839MP or AcrySof IQ PanOptix trifocal IOL after phacoemulsification were recruited. Manifest refraction, uncorrected and corrected visual acuity (VA) at distance, intermediate and near vision, contrast sensitivity, modulated transfer function (MTF) values and questionnaire answers were compared among the 3 groups using analysis of variance (ANOVA).
RESULTS: Fifty-seven (57) eyes were included in the study: 21 eyes with FineVision (group A), 21 eyes with LISA tri (group B), and 15 eyes with PanOptix IOL (group C). The post-operative mean manifest spherical equivalent was -0.01D, -0.07D, and 0.05D, respectively (p=0.083). Uncorrected distance VA and best-corrected distance VA were similar among the groups. Groups A and C had better uncorrected and corrected intermediate VA at 80 cm and at 60 cm compared to group B. Group A had significantly better uncorrected near visual acuity than groups B and C (p=0.032). Mesopic contrast sensitivity testing showed group C had higher contrast sensitivities without glare in at the spatial frequency of 6 CPD (p=0.038) and with glare at 3 CPD (p=0.039) and at 12 CPD (p=0.009). MTF average height analysis showed that the group A had significantly superior resolution in far targets compared to groups B and C (p=0.001). At near targets, groups A and C had better resolutions compared to group B (p=0.017). There was no significant difference in patient satisfaction for far, intermediate and near VA among the groups.
CONCLUSION: Eyes implanted with any of the 3 trifocal IOL designs achieved excellent uncorrected and bestcorrected distance, intermediate and near vision. FineVision and PanOptix provided significantly better intermediate vision than LISA tri at both 80 cm and 60 cm testing distance. FineVision had better near visual outcomes than PanOptix and LISA tri. Patient satisfaction was high in all 3 trifocal IOLS
Lenses, Intraocular
;
Vision, Ocular
9.Application of femtosecond laser technology in the management of subluxated lens.
Ruihong JU ; Yun CHEN ; Wei YANG ; Mansha HE ; Yuanhong PAN ; Zheming WU
Journal of Southern Medical University 2019;39(7):843-849
OBJECTIVE:
To evaluate the application of femtosecond laser technology in the management of subluxated lens.
METHODS:
We retrospectively analyzed the data of the patients with subluxated lens undergoing femtosecond laser- assisted surgery at the Cataract Center of Guangzhou Aier Eye Hospital between March, 2017 and May, 2019. The LenSx femtosecond laser-assisted cataract surgery system was used to perform capsulotomy and lens fragmentation. According to the patients' eye condition, anterior vitrectomy was performed and capsular retractors was used. After phacoemulsification, I/A and insertion of the tension rings, the intraocular lens (IOL) was implanted into the capsular bag. The perioperative data, complications, visual acuity and intraocular pressure after the operation were recorded, and the stability of the capsular bag and IOLs were assessed.
RESULTS:
We analyzed the data of 25 cases (29 eyes) of subluxated lens, including 16 (16 eyes; 55.17%) as the result of traumatic lens subluxation, 5 (9 eyes; 31.03%) of Mafan syndromes, 1 case (1 eye; 3.45%) of high myopia and 3 cases (3 eyes; 10.34%) of unknown causes. Thirteen 13 eyes (44.83%) showed mild subluxation, 7 (24.14%) had moderate subluxation, and 9 (31.03%) had severe subluxation. Femtosecond laser- assisted capsulorhexis, lens fragmentation and phacoemulsification were successfully completed for 29 eyes, of which 28 eyes (96.55%) retained the complete capsular bag and with successful implantation of the capsular tension devices and IOLs. Nine eyes (31.03%) were treated with anterior segment vitrectomy; iris hooks were used for 2 eyes (6.90%) and capsular bag hooks for 9 eyes (31.03%). The best corrected visual acuity was significantly improved in 29 eyes after operation ( < 0.05). At 1 month after the surgery, 26 eyes (89.66%) showed stably centered IOLs, 2 eyes (6.90%) showed slight tilt of the IOLs, and 3 eyes (10.34%) had anterior capsular contraction. The intraoperative complications included subconjunctival hemorrhage (75.87%), incomplete capsulotomy (17.24%) and contracted pupils (13.79%).
CONCLUSIONS
The application of femtosecond laser assisted technology enhances the surgical safety and effectiveness for subluxated lens, facilitates the choice of individualized surgical options, and promotes maximum recovery of the patients' visual function.
Humans
;
Laser Therapy
;
Lens Implantation, Intraocular
;
Lens Subluxation
;
surgery
;
Lenses, Intraocular
;
Phacoemulsification
;
Retrospective Studies
10.An Integrative Review on Alterations of Sleep after Cataract Surgery
Journal of Korean Biological Nursing Science 2019;21(1):12-21
PURPOSE: The purpose of this study was to review previous literature with an aim to explore the sleep disturbance level after a patient undergoes cataract surgery using an integrative review. METHODS: We used the key words, ‘cataract surgery’, ‘sleep’, and ‘intraocular lens’ to find peer-reviewed publications in seven databases. Among 450 searched articles, eight articles were selected after exclusion of articles that did not meet the criteria. RESULTS: Five of the articles submitted that subjective sleep quality of the patients with blue-filtering intraocular lens (BF-IOL) implant improved as compared to that before surgery. The change of saliva melatonin concentration after BF-IOL implant did not coincide in two of the articles. Two of the articles reported an increased level of intrinsically photosensitive retinal ganglion cells response after BF-IOL implantation. CONCLUSION: The published studies stated that BF-IOL implant did not have any negative impact on quality of sleep among cataract patients suffering with poor sleep. Rregardless of intraocular lens type, cataract surgery may increase photoreception of intrinsically photosensitive retinal ganglion cells one year after surgery although the mechanism was not clear. It is necessary to identify various factors influencing the quality of sleep such as gender and activities among cataract patients with BF-IOL implant in the future.
Cataract
;
Humans
;
Lenses, Intraocular
;
Melatonin
;
Retinal Ganglion Cells
;
Saliva


Result Analysis