Comparisons of electrolytes were measured by  Point-of-care testing and auto-analyzer
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:Электролитийн хэмжээг цусны хийн анализатор болон автомат анализатороор  харьцуулан тодорхойлсон дүн
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Bayarmaa E
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Byambasuren B
			        		
			        		
			        		
			        			2
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Ulziisaikhan E
			        		
			        		
			        		
			        			3
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Bayarjavkhlan Ch
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. MNUMS;Gurvan gal hospital
			        		
			        			2. MNUMS
			        		
			        			3. Mobio laboratory
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		electrolyte;
			        		
			        		
			        		
				        		point-of-care testing;
			        		
			        		
			        		
				        		blood gas analyzer;
			        		
			        		
			        		
				        		potassium;
			        		
			        		
			        		
				        		sodium;
			        		
			        		
			        		
				        		whole blood
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Health Laboratory
	            		
	            		 2017;6(1):27-31
	            	
            	
 
            
            
            	- CountryMongolia
 
            
            
            	- Language:Mongolian
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Introduction:Electrolyte values are measured by two different analyzers: arterial blood gas (Point of care) and auto-analyzers. Those two has different methods to measure electrolytes and have several pros and cons. We evaluated if there was agreement between whole blood electrolytes measured by a point-of-care device and serum electrolytes measured using indirect ion-selective electrodes. 
				        	
				        
				        	Materials and methods:An observational cohort study was conducted in 50 paired venous samples from patients admitted in Gurvan gal central hospital. Those were analyzed on OPTC blood gas devise and Roche c-501 auto-analyzer. Statistical analyses were performed using paired t-test and persons’ correlation test. 
				        	
				        
				        	Results:Sodium mean range was 138.54 mmol/l (SD=3.69) by blood gas analyser, but by the automated analyser mean range was 140.75 mmol/l (SD=4.45). Mean difference of the normal sodium group was 1.77 (SD=1.65, p=0.039), and hyposodium group was 4.4 (SD=0.33, p=0.007). Pottasium mean range was 3.13 mmol/l (SD=0.53) by blood gas analyser, but by the automated analyser mean range was 4.42 mmol/l (SD=0.45). Mean difference of the normal sodium group was 0.18 (p<0.001), and hypokalemi group was 1.44 (p<0.001). 
				        	
				        
				        	Conclusion:Clinicians should be aware of the difference between whole blood and serum electrolytes. A correction factor needs to be determined at each laboratory. 
				        	
				        
				    
			     
	        
	        
	        	- Full text:HL-2017-6(1)-27-31.pdf