An ultrasound-guided robot assisted system for percutaneous facet joint puncture: an experimental study
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.cn115530-20220824-00441
   		
        
        	
        		- VernacularTitle:超声引导下经皮关节突关节穿刺机器人辅助系统的实验研究
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Li GAO
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yuanzhi ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Shaobai WANG
			        		
			        		;
		        		
		        		
		        		
			        		Xufeng HU
			        		
			        		;
		        		
		        		
		        		
			        		Jing MI
			        		
			        		;
		        		
		        		
		        		
			        		Yaou LIU
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 内蒙古医科大学附属医院麻醉科,呼和浩特 010050
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Lumbar vertebra;
			        		
			        		
			        		
				        		Vertebral facet joint;
			        		
			        		
			        		
				        		Ultrasonic treatment;
			        		
			        		
			        		
				        		Robots
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Orthopaedic Trauma
	            		
	            		 2022;24(10):869-873
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To evaluate the accuracy of an ultrasound-guided robot-assisted system for percutaneous facet joint puncture in an animal experiment by registration of ultrasound images and two-dimensional X-ray fluorescence images.Methods:Six specimens of fresh adult sheep lumbar spine were prepared and soaked in gelatin solution. The specimens contained a total of 48 facet joints with 24 ones respectively on the left and right sides. Half of the joints were assigned by the random number table method into a robot-assisted group which were subjected to percutaneous facet joint puncture assisted by the ultrasound-guided robot-assisted system while the contralateral ones into a freehand group which were subjected to percutaneous facet joint puncture by freehand. In the robot-assisted group, the left facet joints were 3 L 1, 3 L 2, 3 L 3 and 3 L 4 ones while the right facet joints were 2 L 1, 3 L 2, 5 L 3 and 2 L 4 ones. In this self-control animal experiment, both the robot-assisted punctures and the freehand punctures were carried out in the same specimen. The puncture success rate, puncture needle adjustment, positioning deviation, orientation deviation and operation time were recorded. Results:The positioning deviation and orientation deviation in the robot-assisted group [(2.21 ± 1.12) mm and 1.51° ± 0.47°] were significantly lower than those in the freehand group [(3.26 ± 1.44) mm and 2.24° ± 0.89°] ( P < 0.05). All the articular facet joint punctures were successfully accomplished for the first time in the robot-assisted group while those required multiple adjustments in the freehand group. In the robot-assisted group, the total operation time [(463.84 ± 34.93) s] was significantly longer than that in the freehand group [(298.40 ± 27.48) s], but the puncture time [(37.97 ± 6.87) s] was significantly shorter than that in the freehand group [(261.61 ± 33.15) s] ( P < 0.05). Conclusion:The ultrasound-guided robot-assisted system for percutaneous facet joint puncture can lead to accurate puncture with limited needle adjustments.