Research on Maintenance Management of Equipment of Magnetic Resonance Imaging Based on RAM.
10.3969/j.issn.1671-7104.2022.03.024
- Author:
Cheng-Chen CHU
1
;
Zhi-Yong JI
1
;
Bin LI
1
Author Information
1. Department of Medical Equipment, Shanghai Sixth People's Hospital Affiliated to Shanghai Jiao Tong University, Shanghai,
- Publication Type:Journal Article
- Keywords:
availability;
magnetic resonance imaging equipment;
maintainability;
reliability
- MeSH:
Equipment Safety;
Equipment and Supplies, Hospital;
Magnetic Resonance Imaging;
Reproducibility of Results;
Safety Management
- From:
Chinese Journal of Medical Instrumentation
2022;46(3):350-354
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE:To optimize the maintenance quality management of MRI equipment and ensure the quality and safety of its clinical use.
METHODS:The data of failure time and repair time of a MRI equipment in three years were collected by magnetic resonance repair report system, and then the reliability, availability and maintainability(RAM) were studied and analyzed.
RESULTS:The results of reliability analysis showed that the communication module was the key subsystem of the MRI equipment. The results of usability analysis showed that RF module was a key subsystem of MRI equipment. Maintainability results showed that the proportion of the MRI equipment not fully utilized due to maintenance-related problems was 2.58%. In order to improve the availability of MRI equipment, the maintenance time of MRI equipment should be shortened.
CONCLUSIONS:RAM-based analysis of MRI equipment can help hospital equipment managers to carry out the work of operation optimization, maintenance strategy formulation and safety management of MRI equipment.