Detecting resorcinol in workplace air by gas chromatography with capillary column 
	    		
		   		
		   			
		   		
	    	
    	 
    	10.11763/j.issn.2095-2619.2019.04.016
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiaojuan LI
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yingxia HE
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Baoli ZHU
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Jiangsu Provincial Center for Disease Prevention and Control Nanjing, Jiangsu 210009, China
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Resorcinol;
			        		
			        		
			        		
				        		Capillary column;
			        		
			        		
			        		
				        		Gas chromatography;
			        		
			        		
			        		
				        		Workplace;
			        		
			        		
			        		
				        		Air
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Occupational Medicine
	            		
	            		 2019;46(04):474-479
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	 OBJECTIVE: To establish a method for detecting resorcinol in workplace air by gas chromatography with capillary column. METHODS: The resorcinol in workplace air was collected into muiti-hole absorbing tubes with distilled water and detected by capillary chromatographic column by direct injection. RESULTS: The good linear range of resorcinol was 1.7-200.0 mg/L. The correlation coefficient was 0.999 9. The detection limit was 0.5 mg/L and the lower limit of quantitation was 1.7 mg/L. The minimum detection concentration was 0.7 mg/m~3(sample volume was 7.5 L). The standard recovery rate was 98.5%-102.6%. The within-run relative standard deviation(RSD) was 0.7%-3.6% and the between-run RSD was 1.8%-5.7%. CONCLUSION: This method has high sensitivity, accuracy and can effectively remove interference, which is suitable for determination of resorcinol in workplace air.