- Author:
Rou WU
1
;
Jilun YE
1
,
2
,
3
;
Xu ZHANG
1
;
Jie LIU
1
;
Maojie YUAN
1
;
Bin WEN
1
Author Information
- Publication Type:Journal Article
- Keywords: ADS125H02; adaptive; isolation module; wide input range
- MeSH: Heart Rate
- From: Chinese Journal of Medical Instrumentation 2021;45(3):284-287
- CountryChina
- Language:Chinese
- Abstract: In order to solve the problem of wide input range of photocurrent, aiming at the shortage of existing pulse signal acquisition system, an adaptive pulse signal acquisition system based on ADS125H02 is designed, which has high collection accuracy, low noise, anti-interference, wide input range, high integration, etc. The measurement range of the system can reach ±10 V, and the accuracy is 0.009 μV. Experimental tests show that the design of the system can extract weak pulse signals, and has good practicality.