A minimally invasive alternative for the treatment of nutcracker syndrome using individualized three-dimensional printed extravascular titanium stents 
	    		
		   		
		   			
		   		
	    	
    	 
    	10.1097/CM9.0000000000000255
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		He WANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yi-Tong GUO
			        		
			        		
			        		
			        			2
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yong JIAO
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Da-Li HE
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Bin WU
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Li-Jun YUAN
			        		
			        		
			        		
			        			2
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yan-Yan LI
			        		
			        		
			        		
			        			3
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yong YANG
			        		
			        		
			        		
			        			4
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Tie-Sheng CAO
			        		
			        		
			        		
			        			2
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Bo ZHANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Department of Urology, Tangdu Hospital, The Fourth Military Medical University, Xi’an, Shaanxi 710038, China
			        		
			        			2. Department of Ultrasound Diagnosis, Tangdu Hospital, The Fourth Military Medical University, Xi’an, Shaanxi 710038, China; 3D Printing Research Center of The Fourth Military Medical University, Xi’an, Shaanxi 710038, China
			        		
			        			3. Department of Radiology, Tangdu Hospital, The Fourth Military Medical University, Xi’an, Shaanxi 710038, China
			        		
			        			4. Department of Ultrasound Diagnosis, Tangdu Hospital, The Fourth Military Medical University, Xi’an, Shaanxi 710038, China
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Nutcracker syndrome;
			        		
			        		
			        		
				        		Extravascular stent;
			        		
			        		
			        		
				        		Three-dimensional printing;
			        		
			        		
			        		
				        		Titanium;
			        		
			        		
			        		
				        		Minimally invasive;
			        		
			        		
			        		
				        		Laparoscopy
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Medical Journal
	            		
	            		 2019;132(12):1454-1460
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:English
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	 Background:This study was designed to evaluate the clinical and radiographic outcomes of patients with nutcracker syndrome (NCS) who were treated with three-dimensional printing (3DP) extravascular titanium stents (EVTSs). The 3DP EVTS was expected to release the hypertension of the left renal vein (LRV) produced by its compression between the superior mesenteric artery (SMA) and the aorta without causing any complications.
				        	
				        
				        	Method:The pre-operative kidney model of each patient was printed out to enable surgical planning. After that, the EVTS was designed based on the LRV’s primitive physiologic structure using computer-aided design software, and each stent was printed out with a precision setting of 20 μm. Seventeen patients who had been suffering from NCS underwent laparoscopic 3DP EVTS placement. The surgical procedure was designed for the placement of EVTS, taking great care in positioning and fixing the stent. Surgical data, which included patient demographic characteristics as well as pre- and post-operative test results, were collected and analyzed.
				        	
				        
				        	Results:The mean duration of surgery was 75 ± 9 min, and the mean blood loss was 20 ± 5 mL. Computed tomography examinations revealed that the pre- and post-operative angle between the SMA and the aorta ranged from 18.7° ± 4.3° to 48.0° ± 8.8° (P < 0.05); in patients with left varicocele, the mean diameter of the left spermatic vein ranged from 3.7 ± 0.5 to 1.3 ± 0.2 mm (P < 0.05). Moreover, Doppler ultrasound examinations showed that the peak velocity of blood flow at the hilar area ranged from 12.4 ± 3.3 to 18.5 ± 3.4 cm/s (P < 0.05). No side effects were observed in the 24 to 42 months following surgery.
				        	
				        
				        	Conclusion:The findings after 2 years of follow-up suggest that the 3DP EVTS is a safe and effective minimally invasive alternative for the treatment of NCS.