Performance Evaluation of MEDITAPE UC-11A Strip Test in Estimating the Urine Albumin-to-Creatinine Ratio and Urine Protein-to-Creatinine Ratio
	    		
		   		
		   			
		   		
	    	
    	- Author:
	        		
		        		
		        		
			        		Shinae YU
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Sae Am SONG
			        		
			        		;
		        		
		        		
		        		
			        		Kyung Ran JUN
			        		
			        		;
		        		
		        		
		        		
			        		Jeong Nyeo LEE
			        		
			        		
		        		
		        		
		        		
			        		
			        		Author Information
			        		
 - Publication Type:Original Article
 - Keywords: Albuminuria; Proteinuria; Albumin-to-creatinine ratio; Protein-to-creatinine ratio; Chronic kidney disease
 - MeSH: Albuminuria; Chronic Disease; Humans; Hypertension; Kidney Diseases; Mass Screening; Proteinuria; Renal Insufficiency, Chronic; Sensitivity and Specificity
 - From:Laboratory Medicine Online 2020;10(1):52-57
 - CountryRepublic of Korea
 - Language:English
 - Abstract: 30 mg/g), the concordance rate, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) of UACR, analyzed using MEDITAPE UC-11A, were 80.5, 97.5, 67.0, 70.3, and 97.1%, respectively. Using UPCR, analyzed via quantitative assay, as a reference to estimate proteinuria (UPCR >0.15 g/g), the concordance rate, sensitivity, specificity, PPV, and NPV of UPCR, analyzed using MEDITAPE UC-11A, were 86.7, 94.4, 81.5, 77.6, and 95.6%, respectively.CONCLUSIONS: UACR and UPCR, analyzed using MEDITAPE UC-11A, exhibited relatively high sensitivity and NPV, which is beneficial for laboratory screening for both albuminuria and proteinuria.]]>
 
            