Application of self-made fine-tuning device in distal locking process of femoral interlocking intramedullary nail.
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1003-0034.2019.02.013
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Yun-Mei LI
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Bao-Rui XING
			        		
			        		
			        		
			        			2
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Dong-Hui GUO
			        		
			        		
			        		
			        			2
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Guang-Pu HAN
			        		
			        		
			        		
			        			2
			        			
			        		
			        		
			        		
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Department of Orthopaedics, Cangzhou Hospital of Integrated TCM-WM, Cangzhou 061001, Hebei, China; 172923438@qq.com.
			        		
			        			2. Department of Orthopaedics, Cangzhou Hospital of Integrated TCM-WM, Cangzhou 061001, Hebei, China.
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Femur;
			        		
			        		
			        		
				        		Fracture fixation, intramedullary;
			        		
			        		
			        		
				        		Surgical instruments
			        		
			        		
	        			
        			
        		
 
        	
            
            	- MeSH:
            	
	        			
	        				
	        				
				        		
					        		Adult;
				        		
			        		
				        		
					        		Bone Nails;
				        		
			        		
				        		
					        		Female;
				        		
			        		
				        		
					        		Femoral Fractures;
				        		
			        		
				        		
					        		Fracture Fixation, Intramedullary;
				        		
			        		
				        		
					        		Humans;
				        		
			        		
				        		
					        		Male;
				        		
			        		
				        		
					        		Middle Aged;
				        		
			        		
				        		
					        		Operative Time;
				        		
			        		
				        		
					        		Postoperative Period;
				        		
			        		
				        		
					        		Treatment Outcome
				        		
			        		
	        			
	        			
            	
            	
 
            
            
            	- From:
	            		
	            			China Journal of Orthopaedics and Traumatology
	            		
	            		 2019;32(2):161-165
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	OBJECTIVE:To evaluate the curative efficacy of self-made fine-tuning setting in the process of femoral distal implantation of intramedullary nail.
				        	
				        
				        	METHODS:From October 2015 to October 2017, 66 cases of femoral shaft fracture were treated with anterograde interlocking intramedullary nail including 45 males and 21 females with a mean age of(37.21±11.18) years old. Among them, 36 cases were treated with the manufacture's aiming device and self-made fine-tuning setting (research group), other 30 cases were treated with the manufacture's aiming device(control group). The mean operation time, the times of C-arm scan in surgery, the post-operation complications and the fracture union were observed and compared in two groups.
				        	
				        
				        	RESULTS:Sixty-two cases acquired 8 to 15 months with a mean time of 12.4 months follow-up visit. The post-operation complications and the fracture union between the two groups had no significant difference(>0.05), the mean operation time and the times of C-arm scan in surgery had statistically significant difference(<0.05).
				        	
				        
				        	CONCLUSIONS:Self-made fine-tuning setting in the process of femoral interblocking intramedullary nail could shorten operation time and reduce the the times of C-arm scan.