Design and Verification of a COPD Screening Equipment Based on Dual-differential Pressure Throttling Technique.
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1671-7104.2019.04.008
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Mu WANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Zijun HE
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Wei FANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yanyan CHEN
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. Institute of Intelligent Machines, Chinese Academy of Sciences, Hefei,
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		COPD;
			        		
			        		
			        		
				        		FEV1;
			        		
			        		
			        		
				        		FVC;
			        		
			        		
			        		
				        		dual-differential pressure throttling technique
			        		
			        		
	        			
        			
        		
 
        	
            
            	- MeSH:
            	
	        			
	        				
	        				
				        		
					        		Forced Expiratory Volume;
				        		
			        		
				        		
					        		Humans;
				        		
			        		
				        		
					        		Pulmonary Disease, Chronic Obstructive;
				        		
			        		
				        		
					        		diagnosis;
				        		
			        		
				        		
					        		Reproducibility of Results;
				        		
			        		
				        		
					        		Respiratory Function Tests;
				        		
			        		
				        		
					        		economics;
				        		
			        		
				        		
					        		instrumentation;
				        		
			        		
				        		
					        		Tidal Volume;
				        		
			        		
				        		
					        		Vital Capacity
				        		
			        		
	        			
	        			
            	
            	
 
            
            
            	- From:
	            		
	            			Chinese Journal of Medical Instrumentation
	            		
	            		 2019;43(4):263-265
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	This study was to design a chronic obstructive pulmonary disease (COPD) screening equipment, based on the dual-differential pressure throttling technique. The technique combined a wide range, but low-resolution ratio sensor and a narrow range, but high-resolution ratio one. It can accurately detect the indexes of forced vital capacity (FVC), forced expiratory volume in one second (FEV), one second rate(FEV/FVC (%)), and achieve them in a low-cost way. The new designed machine will be compared with a British machine, named ML-3500. The correlations of FVC and FEV between new machine and ML-3500 were 0.998 and 0.999, respectively. The P values of paired test of these two indexes were over 0.05. Bland-Altman analysis of FVC, FEV and FEV/FVC (%) showed that more than 90% of the scatter points of the three parameters fell within the consistency interval. This machine can be used to accurately screen COPD and its low-cost would be advantage to promote in large population.