The application of IFCC reference methods in methodological evaluation of glycated hemoglobin measurement systems
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.1009-9158.2018.11.007
   		
        
        	
        		- VernacularTitle:IFCC参考方法在五种糖化血红蛋白检测系统评价中的应用
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Tianjiao ZHANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Yungang PU
			        		
			        		;
		        		
		        		
		        		
			        		Huijuan ZHOU
			        		
			        		;
		        		
		        		
		        		
			        		Rong MA
			        		
			        		;
		        		
		        		
		        		
			        		Jiangtao ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Daguang WANG
			        		
			        		;
		        		
		        		
		        		
			        		Chuanbao LI
			        		
			        		;
		        		
		        		
		        		
			        		Miao LI
			        		
			        		;
		        		
		        		
		        		
			        		Yufei WANG
			        		
			        		;
		        		
		        		
		        		
			        		Chuanbao ZHANG
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 100730,北京医院国家老年医学研究中心 卫生部临床检验中心 北京市临床检验工程技术研究中心
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Hemoglobin A,glycosylated;
			        		
			        		
			        		
				        		Reference values;
			        		
			        		
			        		
				        		Reference standards;
			        		
			        		
			        		
				        		Chromatography,high pressure liquid;
			        		
			        		
			        		
				        		Chromatography,ion exchange;
			        		
			        		
			        		
				        		Electrophoresis,capillary;
			        		
			        		
			        		
				        		Evaluation studies
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Laboratory Medicine
	            		
	            		 2018;41(11):821-826
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To perform a methodological evaluation study of 4 HPLC based systems and a capillary electrophoresis based system , with the International Federation of Clinical Chemistry and Laboratory Medicine ( IFCC) glycated hemoglobin reference method as a comparative method .Methods 40 hemolysis samples of variety concentrations were prepared .The samples were measured by IFCC reference method and 5 glycated hemoglobin testing systems , respectively and trueness verification was performed according to the Clinical &Laboratory Standards Institute (CLSI) guideline EP9-A3.Whole blood samples were used to test the systems'precision, linearity and analytical interferences .Results The average CV of IFCC reference method results of 40 hemolysis samples was 1.4%( range from 0.2%to 2.5%).No outlier was found in the results of the 5 testing system.The slopes ranged from 0.9902 to 1.0267, and intercepts from -0.1526 mmol/mol to +0.1512 mmol/mol, squared correlation coefficient from 0.9962-0.9971, biases at two medical decision level were less than 0.3%HbA1c.Within-laboratory precisions were less than 2% (NGSP unit).Bias between all test results and predicted results were less than 5%.High concentration of glucose showed certain interference to glycated Hemoglobin tests but had little influence in clinical practice.Conclusions The results of the 5 testing system are comparable to the IFCC reference method , the results of precision and linearity evaluation meet the requirements of related guidelines .