Highly sensitive assay for the determination of therapeutic peptide desmopressin in human plasma by UPLC–MS/MS
	    		
		   		
	    	
    	
    	
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Gudlawar Kumar Shiva
			        		
			        		;
		        		
		        		
		        		
			        		Pilli Rao Nageswara
			        		
			        		;
		        		
		        		
		        		
			        		Siddiraju Sridhar
			        		
			        		;
		        		
		        		
		        		
			        		Dwivedi Jaya
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Desmopressin;
			        		
			        		
			        		
				        		Solid-phase extraction (SPE);
			        		
			        		
			        		
				        		Ultra performance liquid chromatography–tandem mass spectrometry;
			        		
			        		
			        		
				        		Method validation;
			        		
			        		
			        		
				        		Pharmacokinetics
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Journal of Pharmaceutical Analysis
	            		
	            		 2017;7(3):196-202
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	An analytical method based on ultra-performance liquid chromatography with positive ion electrospray ionization (ESI) coupled with tandem mass spectrometry (UPLC–MS/MS) was developed and validated for the determination of therapeutic peptide desmopressin in human plasma. A desmopressin stable labeled isotope (desmopressin d8) was used as an internal standard. Analyte and the internal standard were extracted from 200 μL of human plasma via solid-phase extraction technique using Oasis WCX cartridges. The chromatographic separation was achieved on an Aquity UPLC HSS T3 column by using a gradient mixture of methanol and 1 mM ammonium formate buffer as the mobile phase. The calibration curve obtained was linear (r2≥0.99) over the concentration range of 1.01–200 pg/mL. Method validation was performed as per FDA guidelines and the results met the acceptance criteria. The results of the intra- and inter-day precision and accuracy studies were well within the acceptable limits. The proposed method was successfully applied to pharmacokinetic studies in humans.