The improvement of the clinical performance for 1235 autoDelfia
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:AutoDelfia 1235时间分辨荧光免疫分析仪临床应用的改进
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Weishi QIN
			        		
			        		;
		        		
		        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Plate;
			        		
			        		
			        		
				        		Improvement;
			        		
			        		
			        		
				        		Sample;
			        		
			        		
			        		
				        		Test
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Immunology
	            		
	            		 1985;0(01):-
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To improve the dual  plate testing,using mono  plate reagent kits of thyroid humor items by re  programming for 1235 AutoDelfa and potimize test time of dual  plate by presetting the testing priority.Methods:The possibility of dual  plate testing using mono  plate reagent kits was evaluated,by comparative study of the volume of each reagent bottle located in the kits and the volume of each reagent used in clinical experiment of dual  plate testing for 5 thyroid hormone items.All the new control files and assay files of thyroid hormone items for dual  plate were created by reprogramming.The experiment time for dual  plate testing was decreased by presetting the testing priority.Results:The maximum samples per test were increased from 96 to 192 for thyroid hormone test.The time for dual  plate test was decreased from 8 hours to 6 hours.Conclusion:The clinical performance of the developed way was successful.The clinical performance of 1235 AutoDelfia could be developed by the application of this study.