Determination of Stilbene Glucosides in Yening Granules by RP-HPLC with Pre-column Derivation
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:柱前衍生化RP-HPLC法测定夜宁颗粒中二苯乙烯苷的含量
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Qin ZHU
			        		
			        		;
		        		
		        		
		        		
			        		Qiulian LIU
			        		
			        		;
		        		
		        		
		        		
			        		Xiangjie HUANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Pre-column derivation;
			        		
			        		
			        		
				        		Yening granules;
			        		
			        		
			        		
				        		Stilbene glycosides;
			        		
			        		
			        		
				        		RP-HPLC;
			        		
			        		
			        		
				        		DNS-Cl
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Pharmacist
	            		
	            		 2015;(11):2008-2010
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To establish a RP-HPLC method with pre-column derivation for determining the content of stilbene glyco-sides in Yening granules. Methods:DNS-Cl was used as the derivation reagent, a Wondasil-C8 column (250 mm × 4. 6 mm, 5 μm) was used as the stationary phase and 30mM ammonium phosphatere gulation (pH 3. 5)-acetonitrile (78∶22) was applied as the mobile phase, a fluorescence detector (λex=370 nm,λem=560 nm) was used, the flow rate was 1. 0 ml·ml-1 , the column temperature was 30℃ and the injection volume was 20 μl. Results: There was no interference from the other substances in the preparation. The linear range of the two stilbene glycosides was 0.071-27.280 mg·ml-1(r=0.999 9). The average recovery was 99.04%(RSD=0. 1%, n= 9). Conclusion:The method is simple, quick, sensitive and accurate, which can be used for the quality control of Ye-ning granules.