Design and biomechanical study of a novel double-leaf proximal humeral locking plate
	    		
		   		
		   			
		   		
	    	
    	 
    	10.16156/j.1004-7220.2017.05.003
   		
        
        	
        		- VernacularTitle:新型双叶型肱骨近端锁定钢板的研制与生物力学研究
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Lang-Qing ZENG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Lu-Lu ZENG
			        		
			        		;
		        		
		        		
		        		
			        		Yun-Feng CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Yan-Jie LIU
			        		
			        		;
		        		
		        		
		        		
			        		Hui YANG
			        		
			        		;
		        		
		        		
		        		
			        		Zhi-Rong LIN
			        		
			        		;
		        		
		        		
		        		
			        		Wen ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Hai-Feng WEI
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 暨南大学附属珠海医院珠海市人民医院创伤骨科
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Proximal humeral fractures;
			        		
			        		
			        		
				        		Complex fractures;
			        		
			        		
			        		
				        		Locking plates;
			        		
			        		
			        		
				        		Internal fixation;
			        		
			        		
			        		
				        		Biomechanics
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Journal of Medical Biomechanics
	            		
	            		 2017;32(5):407-414
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To design a novel double-leaf proximal humeral locking plate for fixing greater and lesser tuberosities in complex proximal humeral fractures,and evaluate its fixing stability by biomechanical tests.Methods Twelve fresh-frozen humerus specimens with intact rotator cuff were randomly divided into two groups (Group A and Group B) to establish the same greater and lesser tuberosities fracture models.Specimens in Group A were fixed with the double-leaf proximal humeral locking plate,while specimens in Group B were fixed with the proximal humeral internal locking system (PHILOS) and tension band suture,and a 3.5-mm cannulated screw was added to stabilize the lesser tuberosity.The tensile test on subscapularis,infraspinatus and teres,supraspi natus as well as the load-to-failure test on greater and lesser tuberosities were performed on specimens in two groups.Results For subscapularis tensile tests,displacements under 150 N tensile stretch and after fatigue test in Group A were both significantly smaller than those in Group B (P < 0.05).For infraspinatus and teres tensile tests,there were no statistical differences between Group A and B in displacements under 150 N tensile stretch and after fatigue test (P > 0.05).For supraspinatus tensile tests,there were no statistical differences between Group A and B in displacements under 90 N tensile stretch and after fatigue test (P > 0.05).For load-to-failure tests on lesser tuberosity,the failure load in Group A was significantly greater than that in Group B (P < 0.05),and the failure displacement in Group A was significantly smaller than that in Group B (P < 0.05).For load-to-failure tests on greater tuberosity,there were no statistical differences between Group A and B in both the failure load and failure displacement (P > 0.05).Conclusions Compared with the ordinary tension band suture plus cannulated screw for fixing lesser tuberosity,the novel double-leaf proximal humeral locking plate shows more obvious biomechanical stability,with the advantage of simultaneously fixing greater and lesser tuberosities.The research findings provide a new choice for the clinical treatment of complex proximal humeral fractures.