Detection of SNR and geometric accuracy of MRI system based on ACR standards
	    		
		   		
		   			
		   		
	    	
    	 
    	10.7687/j.issn1003-8868.2017.09.074
   		
        
        	
        		- VernacularTitle:基于ACR标准的MRI设备图像信噪比及几何精度检测研究
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		yuan Li FU
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		gang Yong LIANG
			        		
			        		;
		        		
		        		
		        		
			        		Ping NI
			        		
			        		;
		        		
		        		
		        		
			        		kui Di LIN
			        		
			        		;
		        		
		        		
		        		
			        		Hui XIONG
			        		
			        		;
		        		
		        		
		        		
			        		chuan Bing LIU
			        		
			        		;
		        		
		        		
		        		
			        		xin Jian CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Qun ZHONG
			        		
			        		;
		        		
		        		
		        		
			        		Hui XIAO
			        		
			        		;
		        		
		        		
		        		
			        		Zi-qian SHANG-WEN
			        		
			        		;
		        		
		        		
		        		
			        		CHEN XU
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 福州总医院医学影像中心
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		MRI;
			        		
			        		
			        		
				        		SNR;
			        		
			        		
			        		
				        		geometric accuracy;
			        		
			        		
			        		
				        		ACR phantom;
			        		
			        		
			        		
				        		quality control
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Medical Equipment Journal
	            		
	            		 2017;38(9):74-77
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To detect the SNR and geometric accuracy of MRI based on American College of Radiology (ACR)standards.Methods The SNR and geometric accuracy of Siemens Skyra 3.0T,Siemens Trio 3.0T and GE Excite HD 1.5T MRI were measured with ACR phantom,and the detection results were calculated according to the standards.Results The SNR values of Siemens Skyra 3.0T,Siemens Trio 3.0T and GE Excite HD 1.5T MRI were 589.98,438.50 and 277.12 respectively.Siemens Skyra 3.0T MRI had the values of geometric accuracy being-1.93% at X direction,-3.20% at Y direction and 0.68% at Z direction,Siemens Trio 3.0T MRI had the value of geometric accuracy being-0.87% at X direction,-2.33% at Y direction and 1.49% at Z direction,GE Excite HD 1.5T MRI had the values of geometric accuracy being 0.20% at X direction,-1.53% at Y direction and 1.69% at Z direction.Conclusion The detection of the SNR and geometric accuracy of MRI can effectively guarantee the image quality.