Effect of Reproducibility of Nano-liquid Chromatography-Mass Spectrometry on Analysis of Urinary Peptidomics
	    		
		   		
		   			
		   		
	    	
    	 
    	10.11895/j.issn.0253-3820.171080
   		
        
        	
        		- VernacularTitle:纳升液相色谱-质谱分析方法的重现性对尿液多肽组分析结果的影响
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Yong WANG
			        		
			        		
			        		
			        			1
			        			
			        		
			        		
			        		
			        		
			        		;
		        		
		        		
		        		
			        		Li WU
			        		
			        		;
		        		
		        		
		        		
			        		Ling Jin XU
			        		
			        		;
		        		
		        		
		        		
			        		Ming Shui LI
			        		
			        		;
		        		
		        		
		        		
			        		Ning LIU
			        		
			        		;
		        		
		        		
		        		
			        		Liang JIANG
			        		
			        		
		        		
		        		
		        		
		        		
		        			
			        		
			        		Author Information
			        		
		        		
		        		
			        		
			        		
			        			1. 深圳大学生命科学学院
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Peptidomics;
			        		
			        		
			        		
				        		Urine;
			        		
			        		
			        		
				        		Reproducibility;
			        		
			        		
			        		
				        		Random error
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Analytical Chemistry
	            		
	            		 2017;45(10):1475-1481
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	As an extension of proteomics, peptidomics has been widely used in medical and biological researches. However, the effect of reproducibility of identification method on peptidomics is not yet clear. In this work, the urine sample of a healthy people was analyzed for seven times in parallel by nano-liquid chromatography-high-resolution tandem mass spectrometry. To illustrate the variability and stability among these experiments, the number of spectra, the utilization of total spectra, the number of identified peptides, the number of proteins, and the ionic strength and retention time of peptides were counted and compared. The average number of peptides was 208 and the standard deviation was 38. 7. After all of data were combined, 426 peptides belonging to 114 proteins were obtained, while only 109 peptides coming from 35 proteins were identified in each experiment, indicating that there were both an randomness and a relative stability for LC-MS analysis. Increasing the number of parallel experiments would expand the data set of peptidomics, but the rate of increase would decrease over 3 or more measurements. Compared with peptides, the results of peptidomics were more stable at protein level, indicating that proteins were more robustly peptidomics biomarker than the peptides.