Contents Determination of Luteoloside,Quercetin and Hyperoside in the Extract of Lonicera japonica by HPCE
	    		
		   		
		   			
		   		
	    	
    	 
    	10.6039/j.issn.1001-0408.2017.18.27
   		
        
        	
        		- VernacularTitle:HPCE法测定金银花药材中木犀草苷、槲皮素和金丝桃苷的含量
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xue ZENG
			        		
			        		;
		        		
		        		
		        		
			        		Yuanjuan YANG
			        		
			        		;
		        		
		        		
		        		
			        		Zhu CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Qian TANG
			        		
			        		;
		        		
		        		
		        		
			        		Lei SHI
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		HPCE;
			        		
			        		
			        		
				        		Luteoloside;
			        		
			        		
			        		
				        		Quercetin;
			        		
			        		
			        		
				        		Hyperoside;
			        		
			        		
			        		
				        		Lonicera japonica
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Pharmacy
	            		
	            		 2017;28(18):2543-2545
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	OBJECTIVE:To establish a method for contents determination of luteoloside,quercetin and hyperoside in Lonicera japonica. METHODS:HPCE was performed silica capillary column with detection wavelength of 360 nm and separation voltage of 20 kV,electrokinetic sampling,sampling voltage of 15 kV,sampling time of 5 s,operation temperature of 25 ℃.The buffer was consisted of 60 mmol/L sodium tetraborate-50 mmol/L natrium carbonicum-50 mmol/L hydroxypropyl-β-cyclodextrin(pH 9.2). RE-SULTS:The linear ranges of luteoloside,quercetin and hyperoside were 0.06-0.56mg/mL (r=0.9881),0.08-0.56 mg/mL (r=0.9892),0.06-0.49 mg/mL(r=0.9796),respectively. RSDs of precision,stability and reproducibility tests were all lower than 2.0%. The recoveries were 96.12%-99.77%(RSD=1.29%,n=6),95.90%-98.35%(RSD=0.89%,n=6),94.07%-97.45%(RSD=1.33%,n=6),respectively. CONCLUSIONS:The method is simple,accurate,stable and reproducible,and can be used for simultaneous determination of luteoloside,quercetin and hyperoside in L. japonica.