		        	Abstract:
			       	
			       		
				        
				        	Objective To provide a proper concentration of Internal Standard Solution by comparing five different degrees to the simultaneous analysis of 10 metals including V,Cr,Mn,Fe,Ni,Cu,As,Cd,Pb and Hg in Cardeniajasminoides.Method An Inductively Coupled Plasma-Mass Spectrometry method with Microwave-Assisted Digestion was developed with10,50,200,500,1 000 μg/L of Ge,In,Bi as internal standard.Result The degree of 500 μg/L concentration of internal stand solution can both correct the matrix effect and alleviate the contamination as the 1 000 μg/L.Conclusion With excellent detection limits,high sensitivity and wide linear range,the 500 μg/L internal standard may serve as reference in the determination of elements in other traditional Chinese medicine (TCM).