Contents Determination of Microelements in Chrysanthemum indicum from Different Production Fields
	    		
		   		
		   			
		   		
	    	
    	 
    	10.6039/j.issn.1001-0408.2016.06.34
   		
        
        	
        		- VernacularTitle:不同产地野菊花中微量元素含量的测定
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Shiren LUO
			        		
			        		;
		        		
		        		
		        		
			        		Liguo WANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Chrysanthemum indicum;
			        		
			        		
			        		
				        		Microelement;
			        		
			        		
			        		
				        		Atomic absorption spectrometry;
			        		
			        		
			        		
				        		Graphite furnace atomic absorption spec-trometry;
			        		
			        		
			        		
				        		Atomic cooling absorption spectrometry;
			        		
			        		
			        		
				        		Flame atomic absorption spectrometry
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Pharmacy
	            		
	            		 2016;(6):827-829,830
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	OBJECTIVE:To establish a method for the contents determination of microelements in Chrysanthemum indicum from different production fields. METHODS:Graphite furnace atomic absorption (GFAAS) was used to determine the contents of Pb,Cd and As;HGAAS was used to determination the content of Hg;and FAAS was used to determine the contents of Cu,Fe, Mn,Ca,Mg and Zn. RESULTS:The linear range was 0-50 μg/L for Pb(r=0.999 9),0-10 μg/L for Cd(r=0.999 2),0-50 μg/L for As(r=0.999 0),0-20 μg/L for Hg(r=0.999 5),0-5 μg/L for Cu(r=0.999 3),0-15 μg/L for Fe(r=0.999 8),0-2 μg/L for Mn (r=0.999 9),0-50 μg/L for Ca(r=1.000 0),0-20 μg/L for Mg(r=0.999 9)and 0-2 μg/L for Zn(r=0.999 8);RSDs of precision, stability and reroducibility tests were lower than 3%;recoveries were 94.25%~97.43%(RSD=1.07%)、94.97%~99.46%(RSD=1.68%)、96.25%~99.46%(RSD=1.09%,n=6)、96.61%~99.91%(RSD=1.26%,n=6)、94.11%~98.41%(RSD=1.68%,n=6)、93.11%~99.59%(RSD=2.73%,n=6)、93.11%~99.48%(RSD=2.63%,n=6)、93.01%~99.85%(RSD=2.49%,n=6)、95.13%~99.75%(RSD=1.58%,n=6)、94.08%~97.37%(RSD=1.18%,n=6),respectively. CONCLUSIONS:The method is simple,stable and reproducible,and can be used for the contents determination of microelements in C. indicum from different pro-duction fields.