Technique in linear accelerator total body irradiation
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1005-202X.2001.01.009
   		
        
        	
        		- VernacularTitle:直线加速器全身照射技术
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Jiutang ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Zhihong WU
			        		
			        		;
		        		
		        		
		        		
			        		Xuwei LU
			        		
			        		;
		        		
		        		
		        		
			        		Jinlian HE
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Medical Physics
	            		
	            		 2001;18(1):18-19,26
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	This article describes the physical, technical, and dosimetric aspects of total body irradiation (TBI) that was carried out by using 6MV X-Ray from Varian 2300 C/D Linear Accelerator at a distance of 450 cm from target to the treatment table and at a gantry angle of 270°.The dose to lung tissue was limit by setting the individual lead compensators customized before, and using DPD-510 to monitor the absorbed dose of the reference point the absorbed dose in depth of half of body will be (Din+Dout)/2 after taking treatment in both AP position and PA position.