The simulation and optimization of front window airflow in biological safety cabinets
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/J.ISSN.1672-8270.2016.06.006
   		
        
        	
        		- VernacularTitle:生物安全柜前窗气流速度模拟及优化
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiujuan FAN
			        		
			        		;
		        		
		        		
		        		
			        		Linhua LV
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Biological safety cabinets;
			        		
			        		
			        		
				        		Computational fluid dynamics;
			        		
			        		
			        		
				        		Front window airflow;
			        		
			        		
			        		
				        		Biosafety
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Medical Equipment
	            		
	            		 2016;13(6):17-18,19
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:In order to explore the maximum and the optimal velocity of front window air in biological safety cabinets and provide a scientific evidence for relevant test. Methods: As classⅡ biological safety cabinet an testing sample, computational fluid dynamics method is adopted to establish numerical model, then made the simulation of front window airflow. Results: The velocity relationship between front window airflow and the internal air has been confirmed. Conclusion:The maximum velocity of the front window flow is 1.4 m/s and the optimal velocity is 1.0 m/s. The tested data and conclusion can give some suggestions and reference for biological safety cabinets detection.