A method to correct the characterizations of MDCT
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/J.ISSN.1672-8270.2015.09.002
   		
        
        	
        		- VernacularTitle:多排螺旋CT辐射剂量表征量校正方法初探
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Jingwen ZHUANG
			        		
			        		;
		        		
		        		
		        		
			        		Mei BAI
			        		
			        		;
		        		
		        		
		        		
			        		Junzheng ZHENG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Multislice spiral CT;
			        		
			        		
			        		
				        		Computed tomography dose index;
			        		
			        		
			        		
				        		Correct method
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Medical Equipment
	            		
	            		 2015;(9):5-7,8
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To correct the characterizations of MDCT radiation dose by exploring the relationship between CTDIw,∞ and CTDIw.Methods: CTDI100 and CTDI∞ were measured under the conditions of different collimations, pitches and tube voltages of Siemens Definition Flash CT, and CTDIw and CTDIw,∞ were calculated.Results: There were significant differences between CTDIw and CTDIw,∞ which were measured at 0.05 level. And there were no significant differences between CTDIw,∞ after corrected and CTDIw,∞ which were measured at 0.05 level.Conclusion:The characterizations of MDCT which were commonly used were not accurate enough. The result after correction were very closed to the real CTDIw,∞. This showed that the method to correct CTDIw of Siemens Definition Flash CT was mostly accurate. And methods to correct CTDIw of other MDCT needed to be further studied.