Compensation research on energy response for the semiconductor personal dosimeter
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/J.ISSN.1672-8270.2015.03.002
   		
        
        	
        		- VernacularTitle:半导体型个人剂量计能量补偿的研究
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiangming OU
			        		
			        		;
		        		
		        		
		        		
			        		Shian ZHAO
			        		
			        		;
		        		
		        		
		        		
			        		Yanqiu DING
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Semiconductor;
			        		
			        		
			        		
				        		Personal dosimeter;
			        		
			        		
			        		
				        		Pb/Sn additional filters;
			        		
			        		
			        		
				        		Energy compensation
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Medical Equipment
	            		
	            		 2015;(3):4-6,7
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To complete experiments of the energy response compensation for three semiconductor personal dosimeters in order to meet the requirements of IEC and National Standard.Methods: In Beijing SSDL, the semiconductor detectors were covered by the different thicknesses of Pb and Sn additional filters with several diameter holes to compensate the dosimeters’ energy response.Results: At the 50 keV-Cs-137 energy range, the measuring error of dosimeters is met within 30% of energy response.Conclusion:By the compensation of combined Pb and Sn filters with a hole, the semiconductor personal dosimeters can be used for radiation monitoring.