Evaluation of quantitative detection limit and functional sensitivity of prealbumin and their clinical applications analysis
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1673-4130.2014.14.039
   		
        
        	
        		- VernacularTitle:前清蛋白定量检测限和功能灵敏度的评价及临床应用
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Zemin WAN
			        		
			        		;
		        		
		        		
		        		
			        		Xiujuan ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Lei LI
			        		
			        		;
		        		
		        		
		        		
			        		Yunxiu WANG
			        		
			        		;
		        		
		        		
		        		
			        		Xiaobin WU
			        		
			        		;
		        		
		        		
		        		
			        		Weiye CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Cheng ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Jianbing WANG
			        		
			        		;
		        		
		        		
		        		
			        		Peifeng KE
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		prealbumin;
			        		
			        		
			        		
				        		immune nephelometry;
			        		
			        		
			        		
				        		limit of blank;
			        		
			        		
			        		
				        		limit of detection;
			        		
			        		
			        		
				        		functional sensitivity
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			International Journal of Laboratory Medicine
	            		
	            		 2014;(14):1908-1909,1913
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To evaluate the limit of blank (LoB),limit of detection (LoD),limit of quantitation(LoQ)and functional sensitivity (FS)of prealbumin (PA)detected by the Roche Modular P automatic biochemical analyzer.Methods According to the EP17A file of the American Clinical and Laboratory Standards Institute (CLSI),saline as the blank sample and a series of low con-centration samples were detected by the Roche Modular P automatic biochemical analyzer for determining LoB,LoD and LoQ.And FS was determined based on the domestic universal method .Results LoB of PA was 16.35 mg/L,LoD was 18.23 mg/L,LoQ was temporarily unable to evaluate and FS was 25.00 mg/L.The report scope and the report mode in clinic were affirmed by combi-ning with the low value of the reportable scope.Conclusion LoD of PA detected by the Roche Modular P automatic biochemical an-alyzer is established,which provides more valuable information for clinical diagnosis and treatment.Conducting the comparison of different evaluation methods determines the advantages and limitation of the practical application of different methods.