Relationship among oxygen flow or the amount of wetting liquid and oxygen noise
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1671-8283.2013.08.021
   		
        
        	
        		- VernacularTitle:吸氧流量及湿化液量与吸氧噪音的关系
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Jiefang MO
			        		
			        		;
		        		
		        		
		        		
			        		Yongqun HUANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		oxygen uptake;
			        		
			        		
			        		
				        		oxygen flow;
			        		
			        		
			        		
				        		noise
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Modern Clinical Nursing
	            		
	            		 2013;(8):63-64,65
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To investigate the relationship among the oxygen flow or the amount of wetting liquid in humidification bottle and oxygen noise.Method Distilled water were added to one third or half of two sets of oxygen humidification bottle device as the humidification fluid,the noise value generated in the flow rate of 2 L/min,5L/min by two set of inspired oxygen apparatus were recorded.Results Different noise value with same capacity humidifying fluid but different oxygen flow was statistically significant(all P<0.01),the more flow of oxygen the greater the noise; Different noise value generated by the same oxygen flow but different capacity humidifying liquid oxygen was statistically significant(all P<0.01),more humidification fluid,greater the noise. Conclusion Noise generated with the increasing oxygen flow rate and the amount of wetting liquid.