Study on the quality control of chitosan hemostatic sponge
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:壳糖止血海绵的质量研究
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Bingsheng SONG
			        		
			        		;
		        		
		        		
		        		
			        		Hanbao LI
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Chitosan;
			        		
			        		
			        		
				        		Hemostatic sponge;
			        		
			        		
			        		
				        		Quality control
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Journal of Medical Postgraduates
	            		
	            		 2003;0(09):-
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective: To establish a method for quality control of chitosan hemostatic Sponge. Methods: To study the indentification of chitosan , Lidocaine Hydrochloride ,Norfloxacin and inspect the suction water of sponge. Extracting Lidocaine Hydrochloride and Norfloxacin with acid and alkali then determining the content of them by spectrophotometic method. Results: The indetification experiment showed a good result . The suction water of sponge outstripped more than 20 times. The average content of Lidocaine Hydrochloride was 9.55mg/g ,Norfloxacin was 8.46mg/g. The average recovery of Lidocaine Hydrochloride and Norfloxacin was 97.9% and 96.5%. Conclusion: This method can be used to establish the quality control of chitosan hemostatic sponge.