Numerical Analysis of the Influence of Cement Thickness on Stress Distribution in All-ceramic Crowns
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:粘接层厚度对全瓷冠修复效果影响的数值分析
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Guoliang CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Wentao JIANG
			        		
			        		;
		        		
		        		
		        		
			        		Huarong WANG
			        		
			        		;
		        		
		        		
		        		
			        		Yubo FAN
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		all-ceramic crowns;
			        		
			        		
			        		
				        		cement thickness;
			        		
			        		
			        		
				        		finite element analysis
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:Space Medicine & Medical Engineering
	            		
	            		 2006;0(01):-
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To provide theoretical basis for clinical treatment of all-ceramic crown prosthodontics,the influence of cement thickness on the stress distribution in all-ceramic crowns under load was studied numerically.Methods The finite element analysis(FEA) model of a standard mandibular first molar was established using spiral CT and CAD software,and then it was applied to analyze the stress distribution of all-ceramic crowns under vertical loading using FEA software.Results Stress concentration was found in mesial and distal cement;the stress distribution at the cement upper and lower interface was found to be similar,and the stress magnitude discrepancy between the two interfaces was small;the stress of the model with 0.15 mm cement in the cervix was lower about 30% than the one of the 0.05 mm model.Conclusions The cement thickness plays an important role in the stress distribution of all-ceramic restoration system.The model with 0.15 mm cement shows a better resistance to destruction among the three models investigated.