Testing and analysing the characteristics of new double three-field iono-chamber
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:新型复式三野电离室的性能测试与分析
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Shimin LIU
			        		
			        		;
		        		
		        		
		        		
			        		Yunxi CAO
			        		
			        		;
		        		
		        		
		        		
			        		Pengcheng WANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Radiography;
			        		
			        		
			        		
				        		Diagnostic equipment;
			        		
			        		
			        		
				        		Equipment design
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Radiology
	            		
	            		 2001;0(07):-
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective  To test and analyse the characteristics of double three  field iono  chamber developed by ourselves. Methods  We tested the time  characteristic, dose rate  characteristic and X  ray quality  characteristic by using direct test method, and current  characteristic, voltage  characteristic and thickness  characteristic of double three  field iono  chamber by using photograph  test  method with NERO  6000B test  meter, RF 500 mA X  ray equipment and so on. The results were analysed. Results  The time  characteristic and dose rate  characteristic had satisfied linear  scales(the return coefficients of linety  scales were   R  =0.999 61 and   R  =  0.998 88   respectively). The X  ray quality  characteristic was steady (the maximum output variation was ?    max  ≤?5%). The good characteristics of current, voltage and thickness were found(the maximum variations of film density were ?D    max  ≤?0.01, ?D    max  ≤?0.15 and ?D    max  ≤?0.06 respectively). The excellent repeat was showed (repeating error of exposure time and dose were ?T    max  ≤?0.004 s and ?X    max  ≤0.01?10    -5   C/kg). Conclusion  The performace of the new Iono  chamber is excellent and steady, so it has practical value.