PHOSPHORUS INTERFERENCE AND EFFECT OF DIFFERENT INTERFERENCE SUPPRESSING AGENTS ON THE DETERMINATION OF CALCIUM IN FOOD BY ATOMIC ABSORPTION SPECTROMETRY
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:原子吸收法测定食品中钙时磷的干扰及干扰抑制剂效果观察
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Jianxing ZHAO
			        		
			        		;
		        		
		        		
		        		
			        		Yuqun LIU
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		calcium food atomic absorption interference suppressing agent
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Acta Nutrimenta Sinica
	            		
	            		 1956;0(04):-
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Interference of P with determination of Ca in food and the effect of interference suppressing agents of La, Sr and 8-hydroxyquinoline were studied. The results showed that 1) interference of P with Ca related to P/Ca ratio, the relative absorbance of Ca were getting increase along with, the decrease of P/Ca ratio if P/Ca ratio about 0.6; 2) La was more effective than the other interference suppressing agents. The relative absorbance of calcium reached to about 90% when 0.5% or 1% La was used in the P/Ca ratio of 30; 3) Sr had two actions. On one hand Sr had some suppressing action on P interference, and on the other hand Sr depressed the Ca absorption signal obviously; and 4) 8-hydroxyqui-noline had some but not very effective interference suppressing action. The content of Ca determined in food SRM were lower than certified value using 0.1% 8-hydroxyquinoline as interference suppressing agent.