Eatimation of effective doses from a dental cone beam computed tomographic scanner
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.0254-5098.2017.03.012
   		
        
        	
        		- VernacularTitle:对一种口腔锥形束CT的有效剂量的估算
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Wei LUO
			        		
			        		;
		        		
		        		
		        		
			        		Yue ZHAO
			        		
			        		;
		        		
		        		
		        		
			        		Jianchao WANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Cone beam computed tomography(CBCT);
			        		
			        		
			        		
				        		Dose-area product(DAP);
			        		
			        		
			        		
				        		Effective dose;
			        		
			        		
			        		
				        		Conversion coefficient
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Radiological Medicine and Protection
	            		
	            		 2017;37(3):222-225
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To calculate effective dose based on the dose-area Product (DAP) measured in dental cone beam computed tomograhic (CBCT) scanning and using a conversion coefficient.Methods The method of the DAP was used to calculate the effective dose from KODAK 9500 dental CBCT scanner with fields of view (FOV) of 16 cm×15 cm and 12 cmu×8 cm.The DAP-values (mGy · cm2) were measured using a transmission ionization chamber connected to an electrometer (Doseguard 100,RTI Electronics AB,Sweden) and the chamber was placed directly on the exit of the X-ray tube.All measurements were repeated three times and averaged on each FOV data.The conversion coefficient of 0.080 μSv/mGy · cm2 was used to estimate the effective dose on the basis of DAP values,which was quoted from a literature.Results DAP values were found between the two FAOs to vary from 512 to 1 083 mGy · cm2,and the effective doses calculated with 0.080 μSv/mGy · cm2were 41 to 87 μSv.The effective doses given in this paper was lower than those reported in other literatures.Conclusions DAP measurement was found to be well defined and easy-to-use method of determining effective dose for some CBCT Units.However,determination of specific conversion coefficient in the CBCT must be further developed.