Performance verification of VITROS 5600 automatic biochemical analyzer
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/j.issn.1673-4130.2017.01.022
   		
        
        	
        		- VernacularTitle:强生VITROS 5600全自动生化免疫分析仪性能验证报告
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Miaochen YU
			        		
			        		;
		        		
		        		
		        		
			        		Qinmin SUN
			        		
			        		;
		        		
		        		
		        		
			        		Fengquan JIANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		dry chemical method;
			        		
			        		
			        		
				        		performance verification;
			        		
			        		
			        		
				        		biochemical analyzer
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			International Journal of Laboratory Medicine
	            		
	            		 2017;38(1):56-60
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To verify the performance of VITROS 5600 automatic biochemical analyzer.Methods According to laboratory accreditation criteria and performance verification documents of American CLIA'88,the precision,accuracy,liner range,clinical reportable range and biological reference interval were verified for the test of 17 conventional biochemical indicators performed on VITROS 5600 automatic biochemical analyzer.Results The precision、accuracy、liner range、clinical reportable range and biological reference interval of VITROS 5600 automatic biochemical analyzer were all acceptable.Conclusion VITROS 5600 automatic biochemical analyzer could fully meet the requirements in clinical application.