Uncertainty Evaluation of the Determination of Lamotrigine in Human Plasma by LC-MS/MS
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:LC-MS/MS 法测定人血浆中拉莫三嗪浓度的不确定度评定
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Li YANG
			        		
			        		;
		        		
		        		
		        		
			        		Leting ZHU
			        		
			        		;
		        		
		        		
		        		
			        		Weixing FENG
			        		
			        		;
		        		
		        		
		        		
			        		Zhigang ZHAO
			        		
			        		;
		        		
		        		
		        		
			        		Shenghui MEI
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		LC-MS/ MS;
			        		
			        		
			        		
				        		Lamotrigine;
			        		
			        		
			        		
				        		Human plasma;
			        		
			        		
			        		
				        		Uncertainty
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Pharmacist
	            		
	            		 2016;19(5):854-858
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To evaluate the uncertainty in the determination of lamotrigine(LTG)in human plasma by LC-MS/ MS. Methods:The uncertainty sources in the determination of LTG were analyzed and the uncertainty was evaluated and combined. Re-sults:The expanded uncertainty of LTG at low(0. 050 4 μg· ml - 1 )and high(1. 27 μg· ml - 1 )concentrations was 0. 005 18 μg· ml - 1 and 0. 066 4 μg· ml - 1 ,respectively(P = 95% ,k = 2). Conclusion:The uncertainty in the determination of LTG in human plasma by LC-MS/ MS is mainly caused by the protein precipitation recovery,matrix effect and sample preparation at low concentration, and by the matrix effect,sample preparation and repeatability at high concentration.