Dosimetric comparison of helical tomotherapy and volume-modulated arc therapy for upper thoracic esophageal carcinoma
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3760/cma.j.issn.0254-5098.2016.01.011
   		
        
        	
        		- VernacularTitle:胸上段食管癌螺旋断层放疗和容积旋转调强放疗的剂量学比较
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiangkun DAI
			        		
			        		;
		        		
		        		
		        		
			        		Boning CAI
			        		
			        		;
		        		
		        		
		        		
			        		Ruigang GE
			        		
			        		;
		        		
		        		
		        		
			        		Xiaoshen WANG
			        		
			        		;
		        		
		        		
		        		
			        		Yunlai WANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Esophageal neoplasms;
			        		
			        		
			        		
				        		Helical tomotherapy;
			        		
			        		
			        		
				        		Volume-modulated arc therapy;
			        		
			        		
			        		
				        		Dosimetry
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Radiological Medicine and Protection
	            		
	            		 2016;36(1):58-62
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To compare the dosimetric differences between helical tomotherapy (HT) and volume-modulated arc therapy (VMAT) in the treatment of upper thoracic esophageal carcinoma (UTEC).Methods A total of 10 patients with UTEC were randomly selected.HT plan and double-arc VMAT plan were designed and optimized for each patient.The prescription dose was 50 Gy/30 fractions for gross target volume (GTV), 66 Gy/30 fractions for planned target volume (PTV).The dose distribution and conformal index (CI), homogeneity index (HI) of target volume, the D1%, D5%, Dg5%, D99%, and dose of organ at risk (OAR) were analyzed by using the dose volume histogram (DVH).The monitor units and delivery time were also evaluated.Results For GTV and PTV, the D99% of HT plans were slightly higher than those of VMAT plans (t =4.476, 3.756, P < 0.05) , but no significant differences in D1% , D5% , D95% , HI and CI (P > 0.05) were found.The V10, V15, V20 and mean lung dose (MLD) to the total-lung of HT plans were all significantly lower than those of VMAT plans (t =-3.369,-4.824, -4.869,-3.657, P < 0.05).There were no significant differences for V5, V30 and Dmax of cord (P > 0.05).The monitor units and delivery time of VMAT plans were significantly lower than those of HT plans (t =13.970, 7.982, P < 0.05).Conclusions Both HT and VMAT are appropriate for esophageal cancer radiotherapy.HT significantly reduces the radiation dose of the total-lung, while VMAT has obvious advantages in efficiency.