Research on detection analysis of CT quality control
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3969/J.ISSN.1672-8270.2016.01.037
   		
        
        	
        		- VernacularTitle:CT质量控制检测分析
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Xiaofeng DUAN
			        		
			        		;
		        		
		        		
		        		
			        		Yongshou ZHANG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		CT scanner;
			        		
			        		
			        		
				        		Quality assurance;
			        		
			        		
			        		
				        		Detection method;
			        		
			        		
			        		
				        		Quality control
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			China Medical Equipment
	            		
	            		 2016;13(1):119-121
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective:To used the internationally accepted methods of CT quality assurance for testing their quality, to maintain the normal operation of CT scanner and guarantee the quality of photographic image.Methods: Toshiba Activion 16 row CT, set the scanning conditions: tube voltage 120 kv, 310 mas exposure, scan time is 0.33 s, scanning layer thickness: 10 mm, 5 mm, 1.5 mm, view(FOV) 150 mm×150 mm, 512 x 512 reconstruction matrix. We measured the following performance parameters for CT spatial resolution, low contrast resolution,uniformity and noise with Catphan500 phantom made by the phantom laboratory of U.S.A.Results: The spatial resolution, low contrast resolution, uniformity and noise of the CT are all in line with the requirements. Obtained by detecting the X-ray dose, pixel, the thickness of the above parameters are affected, but also the interaction between the above parameters simultaneously.Conclusion: By detecting the spatial resolution, low contrast resolution of medical image, we can effectively guarantee system performance, maintaining optimal image quality.