High-sensitivity simultaneous liquid chromatography-tandem mass spectrometry assay of ethinyl estradiol and levonorgestrel in human plasma
	    		
		   		
		   			
		   		
	    	
    	 
    	10.1016/j.jpha.2015.02.002
   		
        
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Abhishek GANDHI
			        		
			        		;
		        		
		        		
		        		
			        		Swati GUTTIKAR
			        		
			        		;
		        		
		        		
		        		
			        		Priti TRIVEDI
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Ethinyl estradiol;
			        		
			        		
			        		
				        		Levonorgestrel;
			        		
			        		
			        		
				        		LC-MS/MS;
			        		
			        		
			        		
				        		Human plasma;
			        		
			        		
			        		
				        		Derivatization
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Journal of Pharmaceutical Analysis
	            		
	            		 2015;5(5):316-326
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	A sensitive and simultaneous liquid chromatography-tandem mass spectrometry method was developed and validated for quantification of ethinyl estradiol and levonorgestrel. The analytes were extracted with methyl-tert-butyl ether: n-hexane (50:50, v/v) solvent mixture, followed by dansyl derivatization. The chromatographic separation was performed on a Kinetex C18 (50 mm × 4.6 mm, 2.6μm) column with a mobile phase of 0.1% (v/v) formic acid in water and acetonitrile in gradient composition. The mass transitions were monitored in electrospray positive ionization mode. The assay exhibited a linear range of 0.100-20.0 ng/mL for levonorgestrel and 4.00-500 pg/mL for ethinyl estradiol in human plasma. A run time of 9.0 min for each sample made it possible to analyze a throughput of more than 100 samples per day. The validated method has been successfully used to analyze human plasma samples for application in pharmacokinetic and bioequivalence studies.