Reproducibility of normal hepatic proton MRS at 3.0T
	    		
		   		
		   			
		   		
	    	
    	 
    	10.3321/j.issn:1003-3289.2009.10.025
   		
        
        	
        		- VernacularTitle:正常肝脏3.0T质子MRS检查的可重复性
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Li XU
			        		
			        		;
		        		
		        		
		        		
			        		Changhong LIANG
			        		
			        		;
		        		
		        		
		        		
			        		Yuanqiu XIAO
			        		
			        		;
		        		
		        		
		        		
			        		Zhonglin ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Shuixing ZHANG
			        		
			        		;
		        		
		        		
		        		
			        		Shufei XIE
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Liver;
			        		
			        		
			        		
				        		Magnetic resonance spectroscopy;
			        		
			        		
			        		
				        		Reproducibility
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Medical Imaging Technology
	            		
	            		 2009;25(10):1821-1823
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To explore the reproducibility of normal hepatic MRS at 3.0T. Methods The hepatic proton MRS was performed with GE Signa Excite HD 3.0T system and eight-channel torso phased-array coils using PRESS sequence. Thirty-one healthy individuals were enrolled in this study. Liver spectra were collected with TR of 1500 ms, TE of 30 ms, ROI of 2 cm×2 cm×2 cm, NSA of 64 times. The outcomes were statistically analyzed with Wilcoxon signed ranks test and Spearman correlation test.Results There was no significant difference of signal to noise ratio (Z=-0.535,P=0.593), baseline stability (Z=-0.333, P=0.739), linewidth of automatic shimming (Z=-0.305, P=0.761), water suppression (Z=-1.232, P=0.218), height of lipid peak (Z=-0.558,P=0.557), area under the lipid peak (Z=-1.195,P=0.232), height of water peak (Z=-0.647,P=0.518) and area under the warter peak (Z=-0.118, P=0.906) between first examination and second examination. Correlation coefficient of the former and the later measurements of lipid area and water area were 0.784 (P<0.001) and 0.799 (P<0.001), respectively.Conclusion The reproducibility is good for in vivo liver proton MRS at 3.0T.