A Low Noise Amplifier System for Nanopore-based Single Molecule Analysis
	    		
		   		
		   			
		   		
	    	
    	 
    	10.11895/j.issn.0253-3820.150159
   		
        
        	
        		- VernacularTitle:纳米通道单分子检测低噪音电流放大器系统的研究
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Bingyong YAN
			        		
			        		;
		        		
		        		
		        		
			        		Zhen GU
			        		
			        		;
		        		
		        		
		        		
			        		Rui GAO
			        		
			        		;
		        		
		        		
		        		
			        		Chan CAO
			        		
			        		;
		        		
		        		
		        		
			        		Yilun YING
			        		
			        		;
		        		
		        		
		        		
			        		Wei MA
			        		
			        		;
		        		
		        		
		        		
			        		Yitao LONG
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Nanopore;
			        		
			        		
			        		
				        		Single molecule detection;
			        		
			        		
			        		
				        		Current amplifier;
			        		
			        		
			        		
				        		Denoising
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Analytical Chemistry
	            		
	            		 2015;(7):971-976
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	A novel amplifier system was proposed for low-noise recording of pico-ampere current in nanopore experiment (<100 pA). As an example, the amplifier system was applied in α-hemolysin based nanopore detection of DNA-PEG-DNA conjugate to record the signals of translocation and bumping events in buffer solution (1 mol/L KCl, 10 mmol/L Tris--HCl, 1 mmol/L EDTA and pH=8. 0). The amplified current signal was filtered by a 3 kHz Bessel filter and sampled by a 100 kHz analog-digital convertor. As a result, the presented amplifier system could lower the noise in recording the current. The current blockages (<10 pA) of single molecules with low amplitude were recovered due to the high signal-to-noise ratio.