Cold Saponification Method for Determination of 5 Lutein Stereoisomers in Dairy Products by High Performance Liquid Chromatography
	    		
		   		
		   			
		   		
	    	
    	 
    	10.11895/j.issn.0253-3820.140931
   		
        
        	
        		- VernacularTitle:冷皂化-高效液相色谱法测定乳制品中叶黄素的5种顺反式异构体
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Wanqin CHEN
			        		
			        		;
		        		
		        		
		        		
			        		Zhu LIU
			        		
			        		;
		        		
		        		
		        		
			        		Guogang ZHENG
			        		
			        		;
		        		
		        		
		        		
			        		Peiqing MAO
			        		
			        		;
		        		
		        		
		        		
			        		Bingqi ZHU
			        		
			        		;
		        		
		        		
		        		
			        		Yuqi DING
			        		
			        		;
		        		
		        		
		        		
			        		Saijing ZHOU
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		High performance liquid chromatography;
			        		
			        		
			        		
				        		Lutein;
			        		
			        		
			        		
				        		Stereoisomers;
			        		
			        		
			        		
				        		Dairy products
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Chinese Journal of Analytical Chemistry
	            		
	            		 2015;(3):404-408
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	A cold saponification method for determination of 5 lutein stereoisomers in dairy products by high performance liquid chromatography ( HPLC ) was developed. Samples were cold-saponified at room temperature and extracted by n-hexane/petroleum/dichloromethane ( 2: 2: 1 , V/V/V ) . Then 5 lutein stereoisomers were separated on a YMC C30 column with gradient elution using methanol/methyl tert-butyl ether as the mobile phase, and data were acquired by a photodiode array detector at wavelength of 445 nm. The calibration curve was linear in the range of 0 . 127-5 . 082 mg/L with correlation coefficient of 0 . 9999 , and the recoveries were from 96 . 7% to 102 . 2% with the RSDs in the range of 4 . 1%-5 . 4% ( n=6 ) . The limit of detection was 0 . 010 μg/g ( S/N=3 ) , and the limit of quantification was 0 . 030 μg/g ( S/N=10 ) . By presenting results of good accuracy, precision and sensitivity, this method validates its suitability for routine analysis of 5 lutein stereoisomers in dairy products.